| 19" rack unit enclosure | Wall enclosure | Field control unit |
General information | | | |
Operating position | Horizontal | Vertical | Horizontal |
Conformity | CE mark in accordance with EN 50081-1 and EN 50082-2 | |
Design, enclosure | | | |
Weight without module | 8.6 kg | 23 kg | 27 kg |
Degree of protection | IP20 according to EN 60529 | IP65 in accordance with EN 60529, restricted breathing enclosure to EN 50021 | IP55 according to EN 60529 |
Electrical characteristics | | | |
Auxiliary power | 100 ... 240 V AC (nominal range of use 85 ... 264 V), 50 ... 60 Hz (nominal range of use 47 ... 63 Hz) |
Power consumption | Max. 280 VA |
EMC interference immunity (electromagnetic compatibility) | In accordance with the standard requirements of NAMUR NE21 (05/2006) and EN 61326-1 (2013) |
Electrical safety | In accordance with EN 61010-1, overvoltage category II |
Gas inlet conditions, purging gas pressure | | | |
Continuous (recommended) | - | 30 hPa above atmospheric pressure | - |
Continuous (maximum) | - | < 100 hPa above atmospheric pressure | - |
Transient (maximum) | - | 165 hPa above atmospheric pressure | - |
Electrical inputs and outputs | | | |
Analog outputs | - | - | 1 for each component 0/4 ... 20 mA, floating; load ≤ 100 Ω, RL ≤ 750 Ω |
Relay outputs | 8, with changeover contacts, can be freely parameterized, e.g. for measuring range identification; max. load: 24 V AC/DC/1.7 A (total load for all 8 relay outputs in continuous operation max. 160 W), floating, non-sparking | 5, with changeover contacts, can be freely configured, e.g. for measuring range identification; load rating: 24 V AC/DC/1.7 A, isolated, non-sparking |
Digital inputs | 8, designed for 24 V, floating, freely configurable, e.g. for measuring range switchover | 5, designed for 24 V, floating, can be freely configured, e.g. for measuring range switchover |
Ethernet interface Ethernet RJ45, 100-megabit | Rear | Underside | Underside |
Service interface Ethernet RJ45, 100-megabit | Front (behind door) | Inside on the processing unit | Inside on the processing unit |
Interface module 1.1 | 12 relay outputs, with changeover contacts, load rating: 24 V AC/DC/1.7 A (total load for all 12 relay outputs in continuous operation max. 244 W), floating, non-sparking 8 digital inputs, designed for 24 V, floating, freely configurable | - |
Interface module 2.1 | 1 analog output for each component 0/4 ... 20 mA, floating; load 100 Ω ≤ RL ≤ 750 Ω; 3 relay outputs per module, load rating: 24 V AC/DC/1.7 A (total load for all 6 relay outputs in continuous operation max. 122 W), floating, non-sparking | - |
Interface module 2.2 | 1 analog output for each component 0/4 ... 20 mA, floating; load 100 Ω ≤ RL ≤ 750 Ω; 4 analog inputs 0/4 … 20 mA, non-isolated, internal resistance ≤ 100 Ω 4 digital inputs, designed for 24 V, floating | 4 analog inputs 0/4 … 20 mA, non-floating, internal resistance ≤ 100 W |
Climatic conditions | | | |
Permissible operating altitude | 3 000 m above sea level | 2 000 m above sea level |
Permissible ambient temperature (with one module; application-dependent with two modules) | Depends on application, See technical specifications of the modules Ventilation slits must not be covered (recommended minimum clearance upward from the next device when installing 2 modules and at maximum ambient temperature: min. 1 HU) | Depends on application, See technical specifications of the modules | -30 ... + 70 °C during storage and transportation 5 ... 55 °C for regular operation with OXYMAT 7 5 ... 60 °C for operation with OXYMAT 7 and with limited measuring accuracy |
Permissible humidity | < 90% RH (RH: relative humidity), during storage, transportation and operation (must not fall below dew point) |