Climatic conditions | |
Permissible ambient temperature | -20 ... +55 °C (depending on oven temperature) |
Permissible storage/transport temperature | -30 ... +70 °C |
Permissible relative humidity | Max. 90% |
Protection against dust and moisture | |
- According to EN 60529/IEC 60529
| IP 65 |
| NEMA 4X |
Power supply | |
Auxiliary power | 24 V DC (18.5 ... 30.2 V) |
External fuse | T2.5 A |
Power consumption, typical | 18 W |
Power consumption, maximum | 60 W |
Dimensions and weights | |
Width x depth x height | 360 x 300 x 220 mm (approx. 14" x 12" x 9") |
Weight | 15 kg (35 lb.) |
Mounting | |
Installation on | Post, pipe or wall |
Distance from wall or next chromatograph | 300 mm (12") |
Distance from ceiling or floor | 200 mm (8") |
Electromagnetic compatibility | |
Noise suppression | According to CISPR 11 / EN 55011 / DIN VDE 0875 Limit class B |
EMC immunity | According to IEC 60801/DIN VDE 0843 |
Conducted interferences on AC supply lines | |
- According to Part 4 (burst)
| 2 kV |
- According to Part 5 (ms pulses), line against line
| 1 kV |
- According to Part 5 (ms pulses), line against ground
| 2 kV |
Conducted interferences on signal lines | |
- According to Part 4 (burst)
| 1 kV |
Immunity to static discharge | |
- According to Part 2 (ESD)
| 8 kV |
Immunity to fields | |
- According to Part 3 and Part 6
| 10 V/m |
Safety | |
Electrical safety | IEC 61010 / DIN VDE 0411 |
Explosion protection | ATEX and IEC Ex: II 2 G Ex d IIC T4 Gb Class I, Zone 1, Group IIB + H2 T4 Class I, Div 1, Groups B, C, D T4 Factory Sealed |
Oven | |
Number/type | 1 / isothermal |
Purging with N2 | Possible |
Dimensions (D x H) | 160 x 10 mm |
Max. heating power | 35 VA |
Temperature range | 60 … 165 °C |
Temperature stability | ± 0.1 K (60 ... 165 °C) |
Temperature accuracy | ± 3 K (60 ... 165 °C) |
Retention time variations per 10 °C change in ambient temperature | Approx. 0.3% |
Warm-up period from 30 … 100 °C | 10 minutes |
Columns and gases | |
Separating column switching | Multidimensional chromatography with backflushing and cut in live system |
Multifunctional diaphragm valve | For injection and backflushing |
Gas connections | Swagelok 1/8" |
Pressure regulators | Max. 4 single-channel electronic pressure regulators |
Solenoid valves for control of diaphragm valve | 2 NC contacts, 2 NO contacts |
Carrier gas | He, Ar Notice The carrier gas defined for the delivered state must be used. Changing the carrier gas could destroy the thermal conductivity detectors. |
- Gas purity (minimum requirement)
| ≥ 99.999 % (5.0) |
| < 0.1 Ојm |
| Degree of separation 99.99 % for 0.1 Ојm particles |
| < 35 ml/min |
| 500 …700 kPa (g) 600 kPa (g) recommended Important: A continuous carrier gas supply is required for error-free operation (frequent carrier gas failure has a negative effect on the service life of the detectors and the device-internal pressure regulator). In addition, an external two-layer pressure regulator for the carrier gas pressure is strongly recommended. |
Instrument air | Not required |
Sample and injection | |
Sample streams | 3 |
Calibration sample streams | 1 |
Phase | Gaseous |
Permissible sample pressure | 10 … 60 kPa above atmospheric pressure NOTICE: Sample must not contain ethine (acetylene). |
Sample flow | 20 … 100 ml/min |
Max. sample temperature | 120°C |
Solid components | < 0.1 Ојm |
Required filtration | Degree of separation 99.99 % for 0.1 Ојm particles |
Material with which the sample comes into contact | Stainless steel, fused silica, polyimide |
Injection | "Valveless" live injection |
| With multifunctional diaphragm valve |
- Injection volume adjustable using switching times
| From 2 ... 50 Ојl |
Detectors, calibration and performance data | |
Detector type | TCD, max. 8 sensors |
Cell volume | 0.02 µl |
Calibration | Manual or automatic, single level |
Repeatability for calorific value and density | ≤ 0.01% (for natural gas) |
Accuracy for calorific value and density | ≤ 0.1% (for natural gas) |
Linear range | Typically ≥ 104 |
Cycle time | Application-dependent |
Ambient temperature influence | Negligible |
Mean Time to Repair/MTBF | < 1 hour / 3 years (without consumables) |
Electronics: Communication and analytical controller (CAC) | |
Microprocessor | Intel 586 architecture |
Flash EPROM | 128 MB |
Dynamic RAM | 64 MB |
Operating system | Windows CE 5.0 |
Software | Preinstalled. Modifications or upgrades for operation PC downloadable via network or locally |
Electronics: Realtime signal processor (RSP) | |
Microprocessor | Motorola 68376, 20 MHz |
Flash EPROM | 1 MB |
Static RAM | 1 MB |
Operating system | Forth |
Software | Preinstalled. Modifications or upgrades downloadable via internal service interface |
Interfaces | |
Communication | 1 x Ethernet 10BaseT/TCP/IP |
Control system coupling | 1 x Modbus RS 485/RS 232 RTU/ASCII |
Inputs/outputs: Basic equipment | |
Digital outputs (relay contact 0.4 A/24 V DC) | 4, 3 x samples, 1 x calibration |
Digital inputs (24 V to optocoupler) | 4, for 1 = sample flow; 2 = time synchronization; 3 = revision (results have no effect on average values); 4 = calibration |
Status indicator | |
LEDs | For: - Supply voltage
- Software Heartbeat
- Ready
- Maintenance request alert
- Fault
- Sample flow
|
LCD | For: - Sample stream: S1, S2, S3, S4
- Sample components: e.g. CO2, propane etc.
- Measured value of sample as numeric value
|
Recommended operator panel | |
Personal computer | Desktop or laptop |
Processor | At least Pentium III |
Clock | ≥ 800 MHz |
Interfaces | 1 x Ethernet |
Operating system | Windows XP, Windows 7 |
Software | CV Control version 1.30.0.0 and higher |