Climatic conditions
|
Permissible ambient temperature
|
-20 ... +55°C (depending on oven temperature)
|
Permissible storage/transport temperature
|
-30 ... +70?
|
Permissible relative humidity
|
Max. 90%
|
Protection against dust and moisture
|
|
- According to EN 60529/IEC 60529
|
IP 65
|
|
NEMA 4X
|
Power supply
|
Auxiliary power
|
24 V DC (18.5 ... 30.2 V)
|
External fuse
|
T2.5 A
|
Power consumption, typical
|
18 W
|
Power consumption, maximum
|
60 W
|
Dimensions and weights
|
Width x depth x height
|
360 x 300 x 220 mm (approx. 14" x 12" x 9")
|
Weight
|
15 kg (35 lb.)
|
Mounting
|
Installation on
|
Post, pipe or wall
|
Distance from wall or next chromatograph
|
300 mm (12")
|
Distance from ceiling or floor
|
200 mm (8")
|
Electromagnetic compatibility
|
Noise suppression
|
According to CISPR 11 / EN 55011 / DIN VDE 0875 Limit class B
|
EMC immunity
|
According to IEC 60801/DIN VDE 0843
|
Conducted interferences on AC supply lines
|
|
- According to Part 4 (burst)
|
2 kV
|
- According to Part 5 (ms pulses), line against line
|
1 kV
|
- According to Part 5 (ms pulses), line against ground
|
2 kV
|
Conducted interferences on signal lines
|
|
- According to Part 4 (burst)
|
1 kV
|
Immunity to static discharge
|
|
- According to Part 2 (ESD)
|
8 kV
|
Immunity to fields
|
|
- According to Part 3 and Part 6
|
10 V/m
|
Safety
|
Electrical safety
|
IEC 61010 / DIN VDE 0411
|
Explosion protection
|
ATEX and IEC Ex: II 2 G Ex d IIC T4 Gb
Class I, Zone 1, Group IIB + H2 T4
Class I, Div 1, Groups B, C, D T4
Factory Sealed
|
Oven
|
Number/type
|
1 / isothermal
|
Purging with N2
|
Possible
|
Dimensions (D x H)
|
160 x 10 mm
|
Max. heating power
|
35 VA
|
Temperature range
|
60 … 165 °C
|
Temperature stability
|
± 0.1 K (60 ... 165 °C)
|
Temperature accuracy
|
± 3 K (60 ... 165 °C)
|
Retention time variations per 10 °C change in ambient temperature
|
Approx. 0.3%
|
Warm-up period from 30 … 100 °C
|
10 minutes
|
Columns and gases
|
Separating column switching
|
Multidimensional chromatography with backflushing and cut in live system
|
Multifunctional diaphragm valve
|
For injection and backflushing
|
Gas connections
|
Swagelok 1/8"
|
Pressure regulators
|
Max. 4 single-channel electronic pressure regulators
|
Solenoid valves for control of diaphragm valve
|
2 NC contacts, 2 NO contacts
|
Carrier gas
|
He, Ar
Notice:
The carrier gas defined for the delivered state must be used.
Changing the carrier gas could destroy the thermal conductivity detectors.
|
- Gas purity (minimum requirement)
|
? 99.999 % (5.0)
|
|
< 0.1 ?m
|
|
Degree of separation 99.99 % for 0.1 ?m particles
|
|
< 35 ml/min
|
|
500 …700 kPa (g)
600 kPa (g) recommended
Important: A continuous carrier gas supply is required for error-free operation (frequent carrier gas failure has a negative effect on the service life of the detectors and the device-internal pressure regulator). In addition, an external two-layer pressure regulator for the carrier gas pressure is strongly recommended.
|
Instrument air
|
Not required
|
Sample and injection
|
Sample streams
|
3
|
Calibration sample streams
|
1
|
Phase
|
Gaseous
|
Permissible sample pressure
|
10 … 60 kPa above atmospheric pressure NOTICE: Sample must not contain ethine (acetylene).
|
Sample flow
|
20 … 100 ml/min
|
Max. sample temperature
|
120°C
|
Solid components
|
< 0.1 ?m
|
Required filtration
|
Degree of separation 99.99 % for 0.1 ?m particles
|
Material with which the sample comes into contact
|
Stainless steel, fused silica, polyimide
|
Injection
|
"Valveless" live injection
|
|
With multifunctional diaphragm valve
|
- Injection volume adjustable using switching times
|
From 2 ... 50 ?l
|
Detectors, calibration and performance data
|
Detector type
|
TCD, max. 8 sensors
|
Cell volume
|
0.02 µl
|
Calibration
|
Manual or automatic, single level
|
Repeatability for calorific value and density
|
? 0.01% (for natural gas)
|
Accuracy for calorific value and density
|
? 0.1% (for natural gas)
|
Linear range
|
Typically ? 104
|
Cycle time
|
Application-dependent
|
Ambient temperature influence
|
Negligible
|
Mean Time to Repair/MTBF
|
< 1 hour / 3 years (without consumables)
|
Electronics: Communication and analytical controller (CAC)
|
Microprocessor
|
Intel 586 architecture
|
Flash EPROM
|
128 MB
|
Dynamic RAM
|
64 MB
|
Operating system
|
Windows CE 5.0
|
Software
|
Preinstalled. Modifications or upgrades for operation PC downloadable via network or locally
|
Electronics: Realtime signal processor (RSP)
|
Microprocessor
|
Motorola 68376, 20 MHz
|
Flash EPROM
|
1 MB
|
Static RAM
|
1 MB
|
Operating system
|
Forth
|
Software
|
Preinstalled. Modifications or upgrades downloadable via internal service interface
|
Interfaces
|
Communication
|
1 x Ethernet 10BaseT/TCP/IP
|
Control system coupling
|
1 x Modbus RS 485/RS 232 RTU/ASCII
|
Inputs/outputs: Basic equipment
|
Digital outputs (relay contact 0.4 A/24 V DC)
|
4, 3 x samples, 1 x calibration
|
Digital inputs (24 V to optocoupler)
|
4, for 1 = sample flow; 2 = time synchronization; 3 = revision (results have no effect on average values); 4 = calibration
|
Status indicator
|
LEDs for
|
- Supply voltage
- Software Heartbeat
- Ready
- Maintenance request alert
- Fault
- Sample flow
|
LCD for
|
- Sample stream: S1, S2, S3, S4
- Sample components: e.g. CO2, propane, etc.
- Measured value of sample as numeric value
|
Recommended operator panel
|
Personal computer
|
Desktop or laptop
|
Processor
|
At least Pentium III
|
Clock
|
? 800 MHz
|
Interfaces
|
1 x Ethernet
|
Operating system
|
Windows XP, Windows 7
|
Software
|
CV Control version 1.30.0.0 and higher
|