| | | | | | | | | |
?
Live injection
The SITRANS CV has a two-stage injection system. Using a micro injection valve, a defined quantity of sample is first brought up to the carrier gas pressure. This eliminates the pressure-dependent error in the dosing quantity present with conventional systems. In the second stage, the sample is transferred to the column by a valveless micro injection system (live dosing). The result is an "active" injection.
The injection volume can be varied time-controlled, and exactly matched to the column requirements.
Valveless live column switching
Because of the high dead volume of conventional valves, only the valveless version can be considered for a miniaturized system. In this case, the generation of differences in flow using several electronic pressure regulators at appropriate positions of the column setup causes a change in the flow directions. (The system operates according to the Wheatstone principle, but pneumatically.) The functions "Cut" and "Backflushing" can then be implemented free of dead volume.
The column system
The separation system consists of up to three separation columns connected in series. Micro TCDs or micro live circuits are installed in sequence ("inline") upstream and downstream of the individual columns. Three electronic pressure regulators supply the columns with carrier gas and carry out the switching functions (injection, backflushing and cut).
By using narrow-bore capillary columns, the separation at high resolution is carried out within a much shorter time, approx. factor 2 to 3 compared to standard capillary columns.
Electronic pressure regulators
A high pressure stability together with rapid changing rates in the hPa range are required for precise and fast switching. This is achieved in the electronic pressure regulators by means of a piezo actuator.
Detector
The micro TCDs (silicon wafer technology) work on the principle of continuous measurement of the different thermal conductivities of the carrier gas and the components to be measured.
The measurement can be carried out without falsification by avoiding catalytic effects on the heating wires and maintaining a constant flow velocity. This permits consistent in-line detection, i.e. without qualitative or quantitative losses of substances.
Modules
The standardized application modules generally feature live injection and live switching functions, detectors and separation columns.
|
Detector
|
Column 1
|
Detector
|
Column 2
|
Detector
|
Circuit
|
Column 3
|
Detector
|
C09
Injection
|
|
Sil5
Non-polar aromatic and aliphatic hydrocarbons
|
TCD
|
Sil5
Non-polar aromatic and aliphatic hydrocarbons
|
TCD
|
Live
|
Porabond Q
All components except molecular filter components
|
TCD
|
C01
Injection
|
TCD
|
Sil5
C3, C4, C5, C6+
|
TCD
|
PoraPLOT/Porabond Q
CO2, C2, H2O
|
TCD
|
Live
|
Molecular filter
H2, (Ar+O2), N2, C1, CO
|
TCD
|
C13
Injection
|
TCD
|
RTX-1
C3, i-C4, n-C4, neo-C5, i-C5, n-C5
Sum C6+ as sum peak in the backflush
|
TCD
|
HayeSepN
N2, CH4, CO2, C2
|
TCD
|
Live
|
|
|
Application
The SITRANS CV is a storage product. Precalibration is carried out at the factory, using helium and argon (as the carrier gas) and a calibration gas. The measured components and switching functions (live injection, backflushing, cut) are saved in the GC. The calibration process itself should be performed during commissioning on-site.
Measurements can be made within the following working ranges:
Component
|
Checked working range (%)
|
Possible working range (%)
|
Methane
|
57 ... 100
|
50 ... 100
|
Nitrogen1)
|
0 ... 22
|
0 ... 25
|
Carbon dioxide
|
0 ... 12
|
0 ... 20
|
Ethane
|
0 ... 14
|
0 ... 20
|
Propane
|
0 ... 5
|
0 ... 15
|
i-butane
|
0 ... 0.9
|
0 ... 10
|
n-butane
|
0 ... 1.8
|
0 ... 10
|
Neopentane
|
0 ... 0.1
|
0 ... 1
|
i-pentane
|
0 ... 0.12
|
0 ... 1
|
n-pentane
|
0 ... 0.12
|
0 ... 1
|
Hexane+2)
|
0 ... 0.08
|
0 ... 3
|
Hexane
|
|
0 ... 1
|
Heptane+3)
|
|
0 ... 1
|
Octane
|
|
0 ... 1
|
Nonane+4)
|
|
0 ... 1
|
Helium
|
Concentration can be entered as a fixed value in the component list
|
H2S
|
< 500 ppm
|
No measured component
|
High/low calorific value
|
Calculated
|
Calculated
|
Density and relative density
|
Calculated
|
Calculated
|
Wobbe index
|
Calculated
|
Calculated
|
Compressibility factor
|
Calculated
|
Calculated
|
Normalization factor
|
Calculated
|
Calculated
|
Table 1: Measured components and performance parameters for Pos. 8_0 (master setup, standard calorific value analysis in accordance with ISO 6976-1995)
1) Any oxygen or carbon monoxide present in the sample will be detected along with the nitrogen and, therefore, taken into account when the nitrogen concentration is determined.
2) Hexane+ = group(iso/n-hexane to iso/n-nonane)
3) Heptane+ = group(iso/n-hexane) and group(iso/n-heptane to iso/n-nonane)
4) Nonane+ = group(iso/n-hexane), group(iso/n-heptane), group(iso/n-octane), group(iso/n-nonane)
Component
|
Possible working range (%)
|
Oxygen
|
0 ... 4
|
Table 2: Measuring range of the additional measured component oxygen of the extended calorific value analysis (see article no. 7KQ3105-1)
The remark in footnote 1 about the detection of oxygen and nitrogen is not valid in the case of an extended calorific value analysis. In this case, all components from Table 1 "Measured components and performance parameters for Pos. 8_0 (master setup, standard calorific value analysis in accordance with ISO 6976-1995)" plus oxygen are detected and quantified.
For the analysis of biomethane the following components and their working ranges are measured (Table 3).
Component
|
Possible working range (%)
|
Calibration gas for biomethane measurement (%).
|
Methane
|
> 80
|
89
|
Nitrogen
|
< 8
|
4
|
Ethane
|
< 6
|
2.5
|
Carbon dioxide
|
< 4
|
2.5
|
Propane
|
< 5
|
1.0
|
Butane
|
< 1.2
|
0.2
|
Oxygen
|
< 3
|
0.2
|
2-Methylpropane (isobutane)
|
< 0.7
|
0.2
|
Hydrogen
|
< 3
|
0.2
|
Table 3: Measured components, working ranges and calibration gas for the analysis of biomethane
For analysis of natural gas with backflush summation, the following components and working ranges are measured:
Component
|
Possible working range (%)
|
Methane
|
50 ... 100
|
Nitrogen
|
0 ... 25
|
Carbon dioxide
|
0 ... 20
|
Ethane
|
0 ... 20
|
Propane
|
0 ... 15
|
i-butane
|
0 ... 10
|
n-butane
|
0 ... 10
|
Neopentane*
|
|
i-pentane
|
0 ... 1
|
n-pentane
|
0 ... 1
|
Hexane+
|
0 ... 3
|
Helium
|
Concentration can be entered as a fixed value in the component list
|
H2S
|
No measured component
|
High/low calorific value
|
Calculated
|
Density and relative density
|
Calculated
|
Wobbe index
|
Calculated
|
Compressibility factor
|
Calculated
|
Normalization factor
|
Calculated
|
Table 4: Component and measuring ranges for the analysis, including backflush summation
* Because the neopentane concentration is very small in practice, this component is not calibrated and is measured with the relative response factor of isopentane. For this reason, a possible working range is not indicated.
Analyses within the checked working range as well as the quality parameters resulting from these (high and low compression and normalization factors, calorific value, density, relative density and Wobbe index) correspond to the requirements listed below.
Measurements within the scope of the possible working ranges (Table 1 "Measured components and performance parameters for Pos. 8_0 (Master setup, standard analysis of calorific value in accordance with ISO 6976-1995)", right column, and Table 2 "Measuring range of the additional measured component oxygen of the extended analysis of calorific value (see article no. 7KQ3105-1)") are possible. However, checking of the repeatability and correctness has not been carried out by the official German body "Physikalisch technischer Bundesanstalt (PTB)".
Concentration range (mol.%)
|
Repeatability according to ISO 6974-5 (2001); molar fraction (%), absolute
|
50 < xi < 100
|
0.03 ... 0.035
|
1 < xi < 50
|
0.011 ... 0.03
|
0.1 < xi < 1
|
0.006 ... 0.011
|
xi < 0.1
|
< 0.006
|
Table 5: The repeatability of the measured components complies with ISO 6974-5 (2001) – Annex B (article no. 7KQ3105-0, 7KQ3105-1)
The repeatability of the calorific value and standard density achieve a relative standard deviation of < 0.01%. SITRANS CV for the analysis of biomethane achieves a relative standard deviation of < 0.05%.
The calibration gas is an extremely important factor for consideration in terms of the MPE (maximum permissible error), and has a significant effect on the accuracy of the overall measuring system. For this reason, SITRANS CV - based on a comparative measuring procedure - can never be more accurate than the calibration gas used. Other parameters besides the accuracy data on the calibration gas certificate are important for the accuracy of a system. Examples of these include the optimum gas composition, the ambient temperatures of the calibration gas cylinders during transportation and operation, potential condensation of, for instance, higher hydrocarbons in a calibration gas cylinder, and the functionality of the sample preparation system.
Under optimum conditions, the SITRANS CV achieves an MPE of < 0.1% for the calorific value and the standard density, whereby the system for measuring biomethane produces an MPE of < 0.5%.
SITRANS CV is designed for measuring with various configurations; the calibration gases required for this purpose are shown below. (Table 6, Measurement and calibration gas components):
SITRANS CV – Overview of possible configurations and the required calibration gases
|
Carrier gas
|
He
|
He
|
Ar
|
He
|
Analyzer module
|
C09
|
C01
|
C01
|
C13
|
|
Calorific value analysis C6+
|
Calorific value analysis C6+ with oxygen
|
Basic Bio-CH4
|
Extended calorific value analysis Bio-CH4
|
C6+ backflush
|
Calculation standard
|
Calculation standard is ISO 6976, GOST and AGA 8 can be selected
|
Article No.
|
7KQ 3105-0
|
7KQ 3105-1
|
7KQ 3105-2
|
7KQ 3105-2
|
Hydrogen
|
-
|
-
|
-
|
M CR
|
-
|
Oxygen
|
-
|
M CR
|
M CR
|
M CR
|
-
|
Nitrogen
|
M CR
|
M CR
|
M CR
|
M CR
|
M CR
|
Carbon dioxide
|
M CR
|
M CR
|
M CR
|
M CR
|
M CR
|
Methane
|
M CR
|
M CR
|
M CR
|
M CR
|
M CR
|
Ethane
|
M CR
|
M CR
|
-
|
M CR
|
M CR
|
Propane
|
M CR
|
M CR
|
-
|
M CR
|
M CR
|
Isobutane
|
M CR
|
M CR
|
-
|
M CR
|
M CR
|
Butane
|
M CR
|
M CR
|
-
|
M CR
|
M CR
|
Neopentane
|
M*1
|
M*1
|
-
|
-
|
M*1
|
Isopentane
|
M CR
|
M CR
|
-
|
-
|
M CR
|
Pentane
|
M CR
|
M CR
|
-
|
-
|
M CR
|
Hexane
|
CR
|
CR
|
-
|
-
|
-
|
Group C6+
|
M*1
|
M*1
|
-
|
-
|
M*2 CR
|
Group C6+ backflush
|
-
|
-
|
-
|
-
|
-
|
Heptane
|
Extended application 7KQ 3105- B02
|
|
|
|
Separate measurement of Group C6 and Group C7+
|
M*3 CR*3
|
M*3 CR*3
|
-
|
-
|
-
|
Separate Groups C6, C7, C8, C9
|
M*4 CR*4
|
M*4 CR*4
|
-
|
-
|
-
|
Caution!
|
Use of the SITRANS CV with a carrier gas different to that of the supplied solution can lead to faults and to the destruction of the analysis module. Depending on the composition of the calibration gas, external heating for the calibration gas cylinder may be necessary.
|
M
|
Measured
|
CR
|
Required as calibration component; composition see catalog PA 01 – SITRANS CV - Function
|
M*1
|
Neopentane is measured with the response factor of isopentane; for direct calibration of neopentane: see operating instructions
|
M*2
|
Group C6+ is measured with the relative response factor of n-hexane
|
M*3/CR*3
|
Groups C6 and C7+ are measured separately and calibrated with n-hexane and n-heptane, respectively
|
M*4/CR*4
|
Group C6, Group C7, Group C8, Group C9 are measured and calibrated separately
|
Table 6: Overview of device versions and available measurement configurations and the calibration gas compositions required for them
SITRANS CV with SIMATIC Extension Unit
|
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Арматура DENDOR
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Датчики и измерители
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Регуляторы и регистраторы
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Пневматическое оборудование
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Краны и Клапаны
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Измерительные приборы
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Системы беспроводного управления «умный дом»
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Бесконтактные выключатели Конечные выключатели Оптические датчики Энкодеры
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FS - Установка умягчения
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FK - Установка обезжелезивания
|