Siemens
СРЕДСТВА ПРОМЫШЛЕННОЙ АВТОМАТИЗАЦИИ
официальный партнер Сименс
Каталог СА01 2018
(4872) 700-366
skenergo@mail.ru

Live injection

The SITRANS CV has a two-stage injection system. Using a micro injection valve, a defined quantity of sample is first brought up to the carrier gas pressure. This eliminates the pressure-dependent error in the dosing quantity present with conventional systems. In the second stage, the sample is transferred to the column by a valveless micro injection system (live dosing). The result is an "active" injection.

The injection volume can be varied time-controlled, and exactly matched to the column requirements.

Valveless live column switching

Because of the high dead volume of conventional valves, only the valveless version can be considered for a miniaturized system. In this case, the generation of differences in flow using several electronic pressure regulators at appropriate positions of the column setup causes a change in the flow directions. (The system operates according to the Wheatstone principle, but pneumatically.) The functions "Cut" and "Backflushing" can then be implemented free of dead volume.

The column system

The separation system consists of three separation columns connected in series. Micro TCDs or micro live circuits are installed in sequence ("inline") upstream and downstream of the individual columns. Three electronic pressure regulators supply the columns with carrier gas and carry out the switching functions (injection, backflushing and cut).

By using narrow-bore capillary columns, the separation at high resolution is carried out within a much shorter time, approx. factor 2 to 3 compared to standard capillary columns.

Electronic pressure regulators

A high pressure stability together with rapid changing rates in the hPa range are required for precise and fast switching. This is achieved in the electronic pressure regulators by means of a piezo actuator.

Detector

The micro TCDs (silicon wafer technology) work on the principle of continuous measurement of the different thermal conductivities of the carrier gas and the components to be measured.

The measurement can be carried out without falsification by avoiding catalytic effects on the heating wires and maintaining a constant flow velocity. This permits consistent in-line detection, i.e. without qualitative or quantitative losses of substances.

Modules

The standardized application modules generally feature live injection and live switching functions, detectors and separation columns.

 

Detector

Column 1

Detector

Column 2

Detector

Circuit

Column 3

Detector

C09

Injection

 

Sil5

Non-polar aromatic and aliphatic hydrocarbons

TCD

Sil5

Non-polar aromatic and aliphatic hydrocarbons

TCD

Live

Porabond Q

All components except molecular filter components

TCD

C01

Injection

TCD

Sil5

C3, C4, C5, C6+

TCD

PoraPLOT/Porabond Q

CO2, C2, H2O

TCD

Live

Molecular filter

H2, (Ar+O2), N2, C1, CO

TCD

C13

Injection

TCD

RTX-1

C3, i‑C4, n‑C4, neo‑C5, i‑C5, n‑C5

Sum C6+ as sum peak in the backflush

TCD

HayeSepN

N2, CH4, CO2, C2

TCD

Live

  


Application

The SITRANS CV is a storage product. Precalibration is carried out at the factory, using helium and argon (as the carrier gas) and a calibration gas. The measured components and switching functions (live injection, backflushing, cut) are saved in the GC. The calibration process itself should be performed during commissioning on-site.

Measurements can be made within the following working ranges:

Component

Checked working range (%)

Possible working range (%)

Methane

57 ... 100

50 ... 100

Nitrogen1)

0 ... 22

0 ... 25

Carbon dioxide

0 ... 12

0 ... 20

Ethane

0 ... 14

0 ... 20

Propane

0 ... 5

0 ... 15

i-butane

0 ... 0.9

0 ... 10

n-butane

0 ... 1.8

0 ... 10

Neopentane

0 ... 0.1

0 ... 1

i-pentane

0 ... 0.12

0 ... 1

n-pentane

0 ... 0.12

0 ... 1

Hexane+2)

0 ... 0.08

0 ... 3

Hexane

 

0 ... 1

Heptane+3)

 

0 ... 1

Octane

 

0 ... 1

Nonane+4)

 

0 ... 1

Helium

Concentration can be entered as a fixed value in the component list

H2S

< 500 ppm

No measured component

High/low calorific value

Calculated

Calculated

Density and relative density

Calculated

Calculated

Wobbe index

Calculated

Calculated

Compressibility factor

Calculated

Calculated

Normalization factor

Calculated

Calculated



Table 1: Measured components and performance parameters for Pos. 8_0 (master setup, standard calorific value analysis in accordance with ISO 6976-1995)

1) Any oxygen or carbon monoxide present in the sample will be detected along with the nitrogen and, therefore, taken into account when the nitrogen concentration is determined.

2) Hexane+ = group(iso/n-hexane to iso/n-nonane)

3) Heptane+ = group(iso/n-hexane) and group(iso/n-heptane to iso/n-nonane)

4) Nonane+ = group(iso/n-hexane), group(iso/n-heptane), group(iso/n-octane), group(iso/n-nonane)

Component

Possible working range (%)

Oxygen

0 ... 4



Table 2: Measuring range of the additional measured component oxygen of the extended calorific value analysis (see article no. 7KQ3105-1)

The remark in footnote 1 about the detection of oxygen and nitrogen is not valid in the case of an extended calorific value analysis. In this case, all components from Table 1 "Measured components and performance parameters for Pos. 8_0 (master setup, standard calorific value analysis in accordance with ISO 6976-1995)" plus oxygen are detected and quantified.

For the analysis of biomethane the following components and their working ranges are measured (Table 3).

Component

Possible working range (%)

Calibration gas for biomethane measurement (%).

Methane

> 80

89

Nitrogen

< 8

4

Ethane

< 6

2.5

Carbon dioxide

< 4

2.5

Propane

< 5

1.0

Butane

< 1.2

0.2

Oxygen

< 3

0.2

2-Methylpropane (isobutane)

< 0.7

0.2

Hydrogen

< 3

0.2



Table 3: Measured components, working ranges and calibration gas for the analysis of biomethane

For analysis of natural gas with backflush summation, the following components and working ranges are measured:

Component

Possible working range (%)

Methane

50 ... 100

Nitrogen

0 ... 25

Carbon dioxide

0 ... 20

Ethane

0 ... 20

Propane

0 ... 15

i-butane

0 ... 10

n-butane

0 ... 10

Neopentane*

 

i-pentane

0 ... 1

n-pentane

0 ... 1

Hexane+

0 ... 3

Helium

Concentration can be entered as a fixed value in the component list

H2S

No measured component

High/low calorific value

Calculated

Density and relative density

Calculated

Wobbe index

Calculated

Compressibility factor

Calculated

Normalization factor

Calculated



Table 4: Component and measuring ranges for the analysis, including backflush summation

* Because the neopentane concentration is very small in practice, this component is not calibrated and is measured with the relative response factor of isopentane. For this reason, a possible working range is not indicated.

Analyses within the checked working range as well as the quality parameters resulting from these (high and low compression and normalization factors, calorific value, density, relative density and Wobbe index) correspond to the requirements listed below.

Measurements within the scope of the possible working ranges (Table 1 "Measured components and performance parameters for Pos. 8_0 (Master setup, standard analysis of calorific value in accordance with ISO 6976-1995)", right column, and Table 2 "Measuring range of the additional measured component oxygen of the extended analysis of calorific value (see article no. 7KQ3105-1)") are possible. However, checking of the repeatability and correctness has not been carried out by the official German body "Physikalisch technischer Bundesanstalt (PTB)".

Concentration range (mol.%)

Repeatability according to ISO 6974‑5 (2001); molar fraction (%), absolute

50 < xi < 100

0.03 ... 0.035

1 < xi < 50

0.011 ... 0.03

0.1 < xi < 1

0.006 ... 0.011

xi < 0.1

< 0.006



Table 5: The repeatability of the measured components complies with ISO 6974-5 (2001) – Annex B (article no. 7KQ3105-0, 7KQ3105-1)

The repeatability of the calorific value and standard density achieve a relative standard deviation of < 0.01%. SITRANS CV for the analysis of biomethane achieves a relative standard deviation of < 0.05%.

The calibration gas is an extremely important factor for consideration in terms of the MPE (maximum permissible error), and has a significant effect on the accuracy of the overall measuring system. For this reason, SITRANS CV - based on a comparative measuring procedure - can never be more accurate than the calibration gas used. Other parameters besides the accuracy data on the calibration gas certificate are important for the accuracy of a system. Examples of these include the optimum gas composition, the ambient temperatures of the calibration gas cylinders during transportation and operation, potential condensation of, for instance, higher hydrocarbons in a calibration gas cylinder, and the functionality of the sample preparation system.

Under optimum conditions, the SITRANS CV achieves an MPE of < 0.1% for the calorific value and the standard density, whereby the system for measuring biomethane produces an MPE of < 0.5%.

SITRANS CV is designed for measuring with various configurations; the calibration gases required for this purpose are shown below. (Table 6, Measurement and calibration gas components):

SITRANS CV – Overview of possible configurations and the required calibration gases

Carrier gas

He

He

Ar

He

Analyzer module

C09

C01

C01

C13

 

Calorific value analysis C6+

Calorific value analysis C6+ with oxygen

Basic Bio-CH4

Extended calorific value analysis Bio-CH4

C6+ backflush

Calculation standard

Calculation standard is ISO 6976, GOST and AGA 8 can be selected

Article No.

7KQ 3105-0

7KQ 3105-1

7KQ 3105-2

7KQ 3105-2

Hydrogen

-

-

-

M CR

-

Oxygen

-

M CR

M CR

M CR

-

Nitrogen

M CR

M CR

M CR

M CR

M CR

Carbon dioxide

M CR

M CR

M CR

M CR

M CR

Methane

M CR

M CR

M CR

M CR

M CR

Ethane

M CR

M CR

-

M CR

M CR

Propane

M CR

M CR

-

M CR

M CR

Isobutane

M CR

M CR

-

M CR

M CR

Butane

M CR

M CR

-

M CR

M CR

Neopentane

M*1

M*1

-

-

M*1

Isopentane

M CR

M CR

-

-

M CR

Pentane

M CR

M CR

-

-

M CR

Group C6+

M*2 CR

M*2 CR

-

-

-

Group C6+ backflush

-

-

-

-

M*2 CR

 

Extended application 7KQ 3105- B02

   

Separate measurement of
Group C6 and Group C7+

M*3 CR*3

M*3 CR*3

-

-

-

Separate groups C6, C7, C8, C9

M*4 CR*4

M*4 CR*4

-

-

-

Caution!

Use of the SITRANS CV with a carrier gas different to that of the supplied solution can lead to faults and to the destruction of the analysis module. Depending on the composition of the calibration gas, external heating for the calibration gas cylinder may be necessary.

M

Measured

CR

Needed as calibration component

M*1

Neopentane is measured with the response factor of isopentane; for direct calibration of neopentane: see operating instructions

M*2

Group C6+ is measured with the relative response factor of n-hexane

M*3/CR*3

Groups C6 and C7+ are measured separately and calibrated with n-hexane and n-heptane, respectively

M*4/CR*4

Group C6, Group C7, Group C8, Group C9 are measured and calibrated separately



Table 6: Overview of device versions and available measurement configurations and the calibration gas compositions required for them

SITRANS CV with SIMATIC Extension Unit

















skener.ru

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Арматура DENDOR

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Датчики и измерители

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Регуляторы и регистраторы

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Пневматическое оборудование

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Краны и Клапаны

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Измерительные приборы

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Системы беспроводного управления «умный дом»

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Бесконтактные выключатели Конечные выключатели Оптические датчики Энкодеры

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FS - Установка умягчения

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30


Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FK - Установка обезжелезивания

  © ООО "СК ЭНЕРГО" 2007-2022
  (4872) 700-366  skenergo@mail.ru
Яндекс.Метрика