Climatic conditions | |
Ambient temperature | +5° ... +40 °C (standard) |
| Min. ‑5 °C |
Relative humidity | 75% (annual average), non-condensing |
Sample gas conditions | Sample gas must not be flammable or explosive. |
Max. sample gas pressure at inlet to sample preparation system | 500 hPa (mbar) |
Max. moisture content in sample gas1)2) | - 17 vol % (cooler type: CSS), with PVDF heat exchanger
- 25 vol % (cooler type: EGK 2‑19), with glass heat exchanger
|
Sample gas temperature | Max. 200 °C at cabinet entry |
Sample gas flow | Approx. 60 l/h per analyzer |
Sampling probe | - Dust load: < 2 g/m3
- Mounting flange: DN 65, PN 6, form B
- Including temperature controller with Pt100
- With internal sampling tube, stainless steel, length: 1 m (can be shortened)
- With filter in probe, to 600 °C
|
Sample gas line, electrically heated | Max. 50 m |
Power supply | |
Supply 1 | 230 V AC, 50 … 60 Hz (-15%, +10%); on request |
Supply 2 | 400 V AC, 50 … 60 Hz (-15%, +10%) |
Power | Max. 4 000 VA; without heated sample gas line |
System design | |
Fusing of electronic consumers | 1-pole or 2-pole (selectable) |
Sample gas cooler | 2-stage |
Output signals | - 4 … 20 mA; corresponding to the analyzer information or via PROFIBUS DP
- Additional digital inputs and outputs via PLC (SIMATIC S7‑1200)
|
Color | RAL 7035 |
Weight | Approx. 160 kg |
Sheet-steel cabinet/frame | Indoor installation |
Explosion protection classification | Installation outside the Ex zone |
Degree of protection | IP54 |
Calibration | Semi-automatic for fully automatic; AUTOCAL on ULTRAMAT 23 freely adjustable up to max. 24-hour interval |
Dimensions | |
Sheet-steel cabinet (with base) for indoor installation | 2 100 x 800 x 800 mm (H x W x D) |
500 mm spacing on the right or left must be provided for the cable inlet and connection of the heated sample gas line. |