The transmitter related to this system is the SITRANS FUS060. Technical specifications to the FUS060 see page 4/216. 2-path sensor with flanges and inline transducers | Error in measurement | Error in measurement at reference conditions | V > 0.5 … 10 m/s, < ± 0.5 % of rate (v = flow speed) | Max. flow velocity | 10 m/s (32 ft/s) | Nominal size | DN 50, DN 65, DN 80, DN 100, DN 125, DN 150, DN 200, DN 250, DN 300 (2” … 12”) | Media temperature | Separate version: -10 … +160 °C (14 … 320 °F) | Ambient temperature (sensor) | Separate version: -20 … +60 °C (-4 … +140 °F) | | | Storage: -40 … +85 °C (-40 … +185 °F) | Enclosure | Standard version: IP67 (NEMA 4X/NEMA 6) | ATEX version: As standard, but with ATEX approval (see below) | Process connections | | PN designated EN 1092-1 type 11 (B) | - DN 50 … 300 (2” … 12”),
PN 40
| - DN 100 … 300 (4” … 12”),
PN 16
| - DN 200 ... 300 (8” … 12”),
PN 10
| Class designated EN 1759-1 | - DN 50 ... 300 (2” … 12”),
class 150
| - DN 50 ... 300 (2” … 12”),
class 300
| Transducer | Inline version welded into pipe | Materials | | Pipe | - DN 50 ... DN 80 (2” … 3”): Cast steel EN 1.1131-GS-15Mn5
| | | - DN 100 ... DN 300 (4” … 12”): Carbon steel EN 1.0345-P235GH
| Flange | - DN 50 ... DN 300 (2” … 12”): EN 1.0025-S235JRG2
| Class | ASTM A105 | Transducer | Stainless steel AISI 316 or similar | Certificates and approvals | | Conformity certificate | The devices are supplied as standard with a Siemens Certificate of Conformity on DVD. | Material certificate | Material certificate according to EN 10204-3.1 available | NDT examination report | Extended material certificate is optionally available on special request (PVR) | Calibration report | A standard calibration report is shipped with each flowmeter. | Extended accredited ISO/IEC 17025 calibration certificates | Optionally available | Approvals | No custody transfer approvals | Ex approval | System ATEX approval for SONO 3300 with remote transmitter FUS060-Ex (ATEX II 2 G Ex dem [ia/ib] IIC T6/T4/T3 Gb) | | | For Ex version the transducer cable length is restricted to 3 m (9.84 ft), in order to meet requirements. |
The sensors are approved according to EU directive 2014/68/EU regarding fluid group 1, classified in category III. Design according to EN 13480 (PED Directive). Coax cable between sensor SONO 3300 and transmitter FUS060Standard Coax cable (75 Ω) | Coax cable with SMB straight plug on one end for the FUS060 connector | |
|---|
Outside diameter | Ø 5.8 mm |  | Length | 3, 15, 30, 60, 90, 120 m (9.84, 49.21, 98.43, 196.85, 295.28, 393.70 ft) between sensor and transmitter | Material (outside jacket) | black PE | Ambient temperature | -10 … +70 °C (14 … 158 °F) | High temperature Coax cable (75 Ω) | Coax cable with SMB straight plug on one end for the FUS060 connector | Outside diameter | Ø 5.13 mm (first 0.3 m (9.84 ft) part to the transducer), Ø 5,8 mm (for remaining cable to the transmitter – with SMB plug at the end) and between these is a black hot melt junction Ø 16 mm (length 70 mm) | Length | 3, 15, 30, 60, 90, 120 m (9.84, 49.21, 98.43, 196.85, 295.28, 393.70 ft) between sensor and transmitter (max. 3 m (9.84 ft)) transducer cable length for Ex area mounted transmitters | Material (outside jacket) | Brown PTFE (0.3 m (9.84 ft) part) and black PE (for remaining cable) | Ambient temperature | -200 … +200 °C (-328 … +392 °F) (brown PTFE transducer part) and -10 … +70 °C (14 … 158 °F) (black PE for remaining transmitter cable part) |
|
|