General information | Based on DINВ ENВ 61207/IECВ 1207. All data based on digital gas mixture H2 in N2 |
Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Span | Application-dependent (see ordering data) |
Measuring ranges with suppressed zero point | Application-dependent (see ordering data) |
Operating position | Front wall, vertical |
Conformity | CE marking in accordance with EN 50081‑1/EN 50081‑2 and RoHS |
Design, enclosure | В |
Degree of protection | IP20 according to ENВ 60529 |
Weight | Approx. 13 kg |
Electrical characteristics | В |
EMC interference immunity (electromagnetic compatibility) | In accordance with standard requirements of NAMUR NE21 (08/98) and ENВ 61326 |
Electrical safety | In accordance with ENВ 61010-1; overvoltage category II |
Auxiliary power (see nameplate) | 100В VВ AC -10%В ...В 120В VВ AC +10%, 48В ...В 63В Hz or 200В VВ AC -10%В ...В 240В VВ AC +10%, 48В ...В 63В Hz |
Power consumption | Approx. 30 VA |
Fuse values | 100В to 120В V: 1.0 T/250 200В ... 240В V: 0.63 T/250 |
Gas inlet conditions | В |
Sample gas pressure | 800В ...В 1В 100В hPa (absolute) |
Sample gas flow | 30В ...В 90В l/h |
Sample gas temperature | Min. 0 to max. 50В В°C, but above the dew point |
Temperature of the measuring cell | 70 ВєC |
Time response | The time and measuring response refers to the measurement of H2 in N2 |
Warm-up period | <В 30В min at room temperature (the technical specification will be met after 2 hours) |
Delayed display (T90) | Approx. 35В s (including dead time) |
Damping (electrical time constant) | 0В ...В 100В s, configurable |
Dead time (the diffusion to the probes is the determining variable) | Approx. 34В s |
Measuring response | The time and measuring response refers to the measurement of H2 in N2 (based on the sample gas pressure 1В 000В hPa absolute, sample gas flow 0.5В l/min, and ambient temperature 25В В°C) |
Output signal fluctuation (3Пѓ value) | <В В±В 1% of the smallest possible span according to rating plate with electronic damping constant of 1В s |
Zero point drift | <В В±В 1% of the current span/week |
Measured-value drift | <В В±В 1% of the smallest possible span (according to rating plate)/week |
Repeatability | <В В±В 1% of the current span |
Detection limit | 1% of the smallest possible span according to rating plate |
Linearity error | <В В±В 1% of the current span |
Influencing variables | Based on sample gas pressure 1В 000В hPa absolute, 0.5 l/min sample gas flow and 25В В°C ambient temperature |
Ambient temperature | <В 2%/10 K referred to smallest possible span according to label |
Accompanying gases | Deviation from zero point (for influence of interfering gas, see section "Cross-interference") |
Sample gas flow | 0.2% of the current measuring span with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure | < 1% of the current span with a change in pressure of 100В hPa |
Auxiliary power | <В 0.1% of the current span with rated voltage В±В 10% |
Electrical inputs and outputs | В |
Analog output | 0/2/4В ...В 20 mA, floating; max. load 750В О© |
Relay outputs | 6, with changeover contacts, freely configurable, e.g. for measuring range identification; load: 24В V AC/DC/1В A, isolated |
Analog inputs | 2, dimensioned for 0/2/4 ... 20 mA for external pressure sensor and correction of cross-interference |
Digital inputs | 6, designed for 24В V, isolated, freely configurable, e.g. for measuring range switchover |
Serial interface | RSВ 485 |
Options | AUTOCAL function with 8 additional digital inputs and relay outputs, also with PROFIBUSВ PA (on request) or PROFIBUSВ DP (on request) |
Climatic conditions | В |
Permissible ambient temperature | -40В ...В +70В В°C during storage and transportation, 5В ...В 45В В°C in operation |
Permissible humidity (dew point must not be fallen below) | < 90% relative humidity as annual average, during storage and transportation |