General information | В |
Measuring ranges | 4, internally and externally switchable; autoranging is also possible |
Smallest possible span (relating to sample gas pressure 1В 000В hPa absolute, 0.5 l/min sample gas flow and 25В В°C ambient temperature) | 2В vol.% or 5В vol.% O2 |
Largest possible measuring span | 100В vol.% O2 |
Measuring ranges with suppressed zero point | Any zero point within 0В ... 100 vol.% can be implemented, provided that a suitable reference gas is used |
Operating position | Front wall, vertical |
Conformity | CE mark in accordance with ENВ 50081-1 and ENВ 50082-2 |
Design, enclosure | В |
Degree of protection | IP20 according to ENВ 60529 |
Weight | Approx. 13 kg |
Electrical characteristics | В |
Auxiliary power | 100 … 120 V AC (nominal range of use 90 … 132 V), 48 … 63 Hz or 200 … 240 V AC (nominal range of use 180 … 264 V), 48 … 63 Hz |
Power consumption | Approx. 45В VA |
EMC (electromagnetic compatibility) | In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety | According to EN 61010-1, overvoltage categoryВ III |
Fuse values | 100В ...В 120 V: 1.0 T/250 200В ...В 240 V: 0.63 T/250 |
Gas inlet conditions | В |
Permissible sample gas pressure | В |
- External reference gas supply
| 800 … 1 200 hPa absolute |
| Atmospheric pressure В±50В hPa |
Sample gas flow | 18 … 60 l/h (0.3 … 1 l/min) |
Sample gas temperature | Min. 0 to max. 50В В°C, but above the dew point |
Sample gas humidity | < 90% relative humidity |
Reference gas pressure (high-pressure version) | 2В 000В to 4В 000 hPa above sample gas pressure, but max. 5В 000В hPa absolute (version without reference gas pump) |
Reference gas pressure (low-pressure version) with external pump | Min. 100В hPa above sample gas pressure |
Dynamic response | В |
Warm-up period | At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90) | 3.5В s |
Damping (electrical time constant) | 0В ...В 100 s, configurable |
Dead time (purging time of the gas path in the unit at 1 l/min) | Approximately 0.5 ... 2.5 s, depending on version |
Time for device-internal signal processing | <В 1В s |
Pressure correction range | В |
Pressure sensor internal | 500 … 2 000 hPa, absolute (see gas inlet conditions for permissible sample gas pressure) |
Measuring response | Based on sample gas pressure 1В 013В hPa absolute, 0.5 l/min sample gas flow and 25В В°C ambient temperature |
Output signal fluctuation | <В В±В 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1В s (corresponds to В±0.25% at 2В Пѓ) |
Zero point drift | < В±В 0.5%/month of the smallest possible span according to rating plate |
Measured-value drift | <В В±0.5%/month of the current measuring range |
Repeatability | <В 1% of the current measuring range |
Detection limit | 1% of the current measuring range |
Linearity error | <В 1% of the current measuring range |
Influencing variables | Based on sample gas pressure 1В 013В hPa absolute, 0.5 l/min sample gas flow and 25В В°C ambient temperature |
Ambient temperature | < 1 % of the current measuring range/ 10В K Zero offset: < 0.1 vol.% O2 absolute/10 K |
Sample gas pressure (with air (100В hPa) as internal reference gas supply, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air.) | - With disabled pressure compensation: <В 2% of the current measuring range /1% pressure change
- With enabled pressure compensation: <В 0.2% of the current measuring range /1% pressure change
|
Accompanying gases | Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas (see table) |
Sample gas flow at zero point | <В 1% of the current measuring range according to rating plate with a change in flow of 0.1В l/min within the permissible flow range |
Auxiliary power | <В 0.1% of the current measuring range with rated voltageВ В±В 10% |
Electrical inputs and outputs | В |
Analog output | 0/2/4 … 20 mA, floating; max. load 750 Ω |
Relay outputs | 6, with changeover contacts, freely configurable, e.g. for measuring range identification; load: 24В V AC/DC/1В A, floating |
Analog inputs | 2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and accompanying gas influence correction (correction of cross-interference) |
Digital inputs | 6, designed for 24 V, floating, freely configurable, e.g. for measuring range switchover |
Serial interface | RSВ 485 |
Options | AUTOCAL function each with 8 additional digital inputs and relay outputs, also with PROFIBUSВ PA or PROFIBUSВ DP |
Climatic conditions | В |
Permissible ambient temperature | -30 … +70 °C during storage and transportation 5 … 45 °C during operation |
Permissible humidity | < 90% relative humidity as annual average, during storage and transportation (must not fall below dew point) |