Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) | 0.5 vol.%, 2 vol.% or 5 vol.% O2 |
Largest possible measuring range | 100 vol.% O2 |
Measuring ranges with suppressed zero point | Any zero point within 0Â ... 100 vol.% can be implemented, provided that a suitable reference gas is used |
Gas inlet conditions | Â |
Permissible sample gas pressure | Â |
| 500Â ... 3Â 000Â hPa absolute |
| Â |
| 500Â ... 1Â 500Â hPa absolute |
| 500Â ... 1Â 300Â hPa absolute |
Sample gas flow | 18Â ... 60Â l/h (0.3Â ... 1Â l/min) |
Sample gas temperature | 0 ... 50 ºC |
Sample gas humidity | <Â 90% RH (relative humidity) |
Reference gas pressure (high-pressure version) | 2Â 000Â ... 4Â 000Â hPa above sample gas pressure, but max. 5Â 000Â hPa |
Reference gas pressure (low-pressure version) | Min. 100Â hPa above sample gas pressure |
Dynamic response | Â |
Warm-up period | At room temperature < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90-time) | Min. 1.5 ... 3.5 s, depending on version |
Damping (electrical time constant) | 0 ... 100 s, configurable |
Dead time (purging time of the gas path in the unit at 1 l/min) | Approx. 0.5 ... 2.5 s, depending on version |
Time for device-internal signal processing | <Â 1Â s |
Pressure correction range | Â |
Pressure sensor | Â |
| 500Â ... 2Â 000Â hPa absolute |
| 500Â ... 3Â 000Â hPa absolute |
Measuring response | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Output signal fluctuation | < 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2σ) |
Zero point drift | <Â 0.5%/month of the smallest possible measuring span according to rating plate |
Measured-value drift | ≤ 0.5%/month of the current measuring range |
Repeatability | ≤ 1%/month of the current measuring range |
Detection limit | 1% of the current measuring range |
Linearity error | 1% of the current measuring range |
Influencing variables | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Ambient temperature | - <Â 0.5%/10 K referred to smallest possible span according to rating plate
- With measuring span 0.5%: 1%/10Â K
|
Sample gas pressure (with air (100Â hPa) as reference gas, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air) | - With disabled pressure compensation: <Â 2% of the current measuring range /1% change in atmospheric pressure
- With disabled pressure compensation: <Â 0.2% of the current measuring range /1% change in atmospheric pressure
|
Accompanying gases | Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas |
Sample gas flow | <Â 1% of the smallest possible span according to rating plate with a change in flow of 0.1Â l/min within the permissible flow range |
Auxiliary power | < 0.1% of the current measuring range with rated voltage ± 10% |