Mode of operation | |
Measuring principle | Analog to digital conversion |
Measuring points | - 1 instrument
- Remote monitoring of 100 instruments with PC and RD software
|
Input | |
Measuring range | |
| |
| - 0 V DC … 10 V DC, 1 … 5 V, 0 … 5 V
|
| - Type J: -50 … +750 °C
(-58 … +1 382 °F) - Type K: -50 … +1 260 °C
(-58 … +2 300 °F) - Type E: -50 … +870 °C
(-58 … +1 578 °F) - Type T: -180 … +371 °C
(-292 … +700 °F) - Type T, 0.1 resolution: -180.0 … +371 °C
(-199.9 … +700 °F)
|
| - 100 Ω RTD: -200 … +750 ºC (-328 … +1 382 ºF)
|
Output signal | |
Output | - 4 … 20 mA (optional)
- Modbus RTU
|
Relays | 2 SPDT Form C relays, rated 3 A at 30 V DC or 3 A at 250 V AC, non-inductive, auto-initializing (optional) |
Communications | - RS 232 with PDC or Modbus RTU
- RS 422/485 with PDC or Modbus RTU
|
Accuracy | |
4 … 20 mA optional output | ± 0.1 % FS ± 0.004 mA |
Process input | ± 0.05 % of span ± 1 count, square root: 10 … 100 % FS |
Thermocouple temperature input | - Type J: ± 1 °C (± 2 °F)
- Type K: ± 1 °C (± 2 °F)
- Type E: ± 1 °C (± 2 °F)
- Type T: ± 1 °C (± 2 °F)
- Type T, 0.1° resolution: ± 1 °C (± 1.8 °F)
|
RTD temperature input | - 100 Ω RTD: ± 1 °C (± 1 °F)
|
Rated operating conditions | |
Ambient conditions | |
- Storage temperature range
| -40 … +85 °C (-40 … +185 °F) |
- Operating temperature range
| -40 … +65 °C (-40 … +149 °F) |
Design | |
Weight | 269 g (9.5 oz) (including options) |
Material (enclosure) | - 1/8 DIN, high impact plastic, UL94V-0, color: gray
- Optional plastic, steel and stainless steel (Type 304, EN 1.4301) NEMA 4 enclosures
|
Degree of protection | Type 4X, NEMA 4X, IP65 (front cover); panel gasket provided |
Electrical connection | |
mA output signal | 2-core copper conductor, twisted, shielded, 0.82 … 3.30 mm² (18 … 12 AWG), Belden 8 760 or equivalent is acceptable |
Electrical connection and relay connection | Copper conductor according to local requirements, rated 3 A at 250 V AC |
Power supply | |
Input voltage option 1 | 85 … 265 V AC, 50/60 Hz; 90 … 265 V DC, 20 W max. |
Input voltage option 2 | 12 … 36 V DC; 12 … 24 V AC, 6 W max. |
Transmitter power supply | One or two isolated transmitter power supplies (optional) |
| One 24 V DC ± 10 % at 200 mA max. |
| Two 24 V DC ± 10 % at 200 mA and 40 mA max. |
External loop power supply | 35 V DC max. |
Output loop resistance | - 24 V DC, 10 … 700 Ω max.
- 35 V DC (external),
100 … 1 200 Ω max.
|
Displays and controls | |
Display | - 14 mm (0.56 inch) high LED
- 2X option for 30.5 mm (1.2 inch) high, red LED
- Numeric range from -1 999 … +9 999
- Four digits, automatic lead zero blanking
- Eight intensity levels
|
Memory | - Non-volatile
- Stores settings for minimum of 10 years if power is lost
|
Programming | - Primary: front panel
- Secondary: meter copy or PC with SITRANS RD software
|
Certificates and approvals | CE, UL, CUL |
Options | |
Enclosures | Plastic, steel, and stainless steel (Type 304, EN 1.4301) NEMA 4 and 4X enclosures |
Mounting | - 2 inch (5.08 cm) pipe mounting kit (zinc plated seal)
- 2 inch (5.08 cm) pipe mounting kit (stainless steel, Type 304, EN 1.4301)
|