Design, enclosure | |
Weight | 15 kg |
Degree of protection | IP65 (NEMA 4X) |
Mounting | |
Installation on | Post, pipe or wall |
Distance from wall or next chromatograph | 300 mm (12") |
Distance from ceiling or floor | 200 mm (8") |
Explosion protection | ATEX and IEC Ex: II 2 G Ex d IIC T4 Gb Class I, Zone 1, Group IIB + H2 T4 Class I, Div 1, Groups B, C, D T4 Factory Sealed |
Support bracket | |
- Mounting part, dimensions (D x H)
| 380 x 110 mm |
| 8 |
- Bracket for gas connection, dimensions (D x H), bracket on right side, mounted at right angle
| 146 x 110 mm |
Electrical characteristics | |
Auxiliary power | 24 V DC (18.5 ... 30.2 V) |
Power consumption | |
| 18 W |
| 60 W |
| IEC 61010 / DIN VDE 0411 |
EMC immunity | According to IEC 60801/DIN VDE 0843 |
- Conducted interferences on AC supply lines
| |
- According to Part 4 (burst)
| 2 kV |
- According to Part 5 (ms pulses), line against line
| 1 kV |
- According to Part 5 (ms pulses), line against ground
| 2 kV |
- Conducted interferences on signal lines
| |
- According to Part 4 (burst)
| 1 kV |
- Immunity to static discharge
| |
- According to Part 2 (ESD)
| 8 kV |
| |
- According to Part 3 and Part 6
| 10 V/m |
| According to CISPR 11 / EN 55011 / DIN VDE 0875 Limit class B |
| T2.5 A |
Gas inlet conditions | |
Permissible sample pressure | 10 … 60 kPa above atmosphere |
Sample flow | 20 … 100 ml/min |
Max. sample temperature | 120°C |
Solid components | < 0.1 mm |
Climatic conditions | |
Permissible ambient temperature | - 20 ... 50 °C (depending on oven temperature) |
Permissible storage/transport temperature | - 30 ... 70 °C |
Permissible relative humidity | Max. 90% |
Sample and injection | |
Sample streams | 3 |
Calibration sample streams | 1 |
Phase | Gaseous |
Required filtration | Degree of separation 99.99 % for < 0.1 Ојm particles |
Material with which the sample comes into contact | Stainless steel, fused silica, polyimide |
Injection | "Valveless" live injection |
| With multifunctional diaphragm valve |
- Injection volume adjustable using switching times
| 2 ... 50 µl |
- Max. operating temperature
| 165 °C |
Oven | |
Number/type | 1/isothermal |
Purging with N2 | Possible |
Dimensions (DxH) | 160 x 10 mm |
Heating capacity | 20 W |
Temperature range | 60 … 155 °C |
Temperature stability | ± 0.1 K (60 ... 155 °C) |
Temperature accuracy | ± 3 K (60 ... 155 °C) |
Retention time variations per 10 °C change in ambient temperature | Approx. 0.3% |
Heating-up period from 30 ... 100 °C | 10 minutes |
Columns and gases | |
Column type | Capillary columns 0.15 ... 0.25 mm/internal |
Separating column switching | Multidimensional chromatography with backflushing and cut in live system |
Multifunctional diaphragm valve | For injection and backflushing |
Gas connections | Swagelok 1/8" |
Pressure regulators | Max. 4 single-channel electronic pressure regulators |
Solenoid valves for control of diaphragm valve | 2 NC contacts, 2 NO contacts |
Carrier gas | H2, N2, He, Ar |
- Gas purity (minimum requirement)
| > 99.999 % (5.0) |
| < 0.1 Ојm |
| Degree of separation 99.99 % for 0.1 Ојm particles |
| 10 ... 60 ml/min |
| 500 …700 kPa (g) 600 kPa (g) recommended Important A continuous carrier gas supply is required for error-free operation (frequent carrier gas failure has a negative effect on the service life of the detectors and the device-internal pressure regulator). In addition, an external two-layer pressure regulator for the carrier gas pressure is strongly recommended. |
Instrument air | Not required |
Detectors, calibration and performance data | |
Detector type | TCD, max. 8 sensors |
Ambient temperature | Negligible |
Cell volume | 0.02 Ојl |
Calibration | Manual or automatic, single-level or multi-level |
Smallest measuring range | 1 000 ppm (application-dependent) |
Linear range | Typically > 104 |
Cycle time | Typically 30 … 240 s |
Electrical inputs and outputs | |
Basic equipment | |
- Digital outputs
(relay contact 0.4 A / 24 V DC)
| 4, freely usable (expandable by NAU or I/O Extender, see "Communication" in "General information" section) |
- Digital inputs
(24 V to optocoupler)
| 4, 3 freely usable (expandable by NAU or I/O Extender, see "Communication" in "General information" section) |
Interfaces | |
| 1 x Ethernet 10BaseT / TCP/IP |
| 1 x RS 485 or RS 232 / Modbus RTU, OPC (ODPC) over Ethernet |
Electronics | |
Communication and Analytic Controller (CAC) | |
| Intel 586 architecture |
| 128 MB |
| 64 MB |
| Windows CE 5.0 |
| Preinstalled. Modifications or upgrades for operation PC downloadable via network or locally |
Realtime signal processor (RSP) | |
| Motorola 68376, 20 MHz |
| 1 MB |
| 1 MB |
| Forth |
| Preinstalled. Modifications or upgrades downloadable via internal service interface |
Controller | |
| 3 |
- Calibration sample streams
| 1 |
| For: - Supply voltage
- Software sign-of-life
- Operational readiness
- Maintenance request alert
- Fault
- Sample flow
|
| For: - Sample stream: S1, S2, S3, S4
- Sample components: e.g. CO2, propane etc.
- Measured value of sample as numeric value
|
Recommended operator panel | |
| Desktop or laptop |
| At least Pentium III |
| ≥ 800 MHz |
| 1 x Ethernet |
| Windows XP or 7 |
| Gas chromatograph portal, from version 1.02 |