Siemens
СРЕДСТВА ПРОМЫШЛЕННОЙ АВТОМАТИЗАЦИИ
официальный партнер Сименс
Каталог СА01 2018
(4872) 700-366
skenergo@mail.ru

Point Level Measurement

 

Criteria

Pointek CLS100

Pointek CLS200

Pointek CLS300

Typical applications

Liquids, slurries, powders, granules, applications in constricted spaces

Liquids, slurries, powders, granules, foam, food, and pharmaceuticals, petrochemicals

Liquids, slurries, powders, granules, relatively high pressure, and temperature, hazardous areas

Max. length
including sensor

100 mm (4 inch)

Rod: 5.5 m (18 ft)
Cable: up to 30 m (98 ft)

 

 

Rod: 1 m (40 inch)
Cable: 25 m (82 ft)

 

Process temperature (Temperature ratings are pressure dependent.
See Pressure/Temperature curves for respective product.)

  • Stainless steel process connection:
    -30 ... +100 °C
    (22 ... +212 °F)
  • Fully Synthetic (PPS process connection):
    -10 ... +100 °C
    (14 ... 212 °F)
  • -40 ... +85 °C
    (-40 ... +185 °F)
  • With thermal
    isolator:
    -40 ... +125 °C
    (-40 ... +257 °F)

 

  • -40 ... +200 °C
    (-40 ... +392 °F)
  • HT version:
    -40 ... +400 °C
    (-40 ... +752 °F)

Process pressure
(Pressure ratings are temperature dependent.
See Pressure/Temperature curves for
respective product.)

 

Up to 10 bar g (146 psi g)

  • Rod versions:
    Up to 25 bar g (365 psi g)
  • Cable version:
    Up to 10 bar g (146 psi g)

Up to 35 bar g (511 psi g)

Output

Stainless steel cable or enclosure version:

  • 4 ... 20/20 ... 4 mA,
    2-wire current loop
  • Solid-state output

Fully-synthetic version (PPS)

  • Relay output

Standard:

  • 1 SPDT Form C relay, solid-state switch

Digital:

  • Solid-state switch included

 

 

Standard:

  • 1 SPDT Form C relay, solid-state switch

Digital:

  • Solid-state switch included

 

Communications

 

Standard:

  • 3 LED indicators

Digital:

  • PROFIBUS PA;
    SIMATIC PDM compatible

Standard:

  • 3 LED indicators

Digital:

  • PROFIBUS PA;
    SIMATIC PDM compatible

Power
Specifications

Standard:

  • 12 ... 33 V DC

Intrinsically Safe (Stainless steel version only):

  • 10 ... 30 V DC

Standard:

  • 12 ... 250 V AC/DC, 0 ... 60 Hz,
    2 W max.

Digital:

  • Bus voltage:
    12 ... 30 V DC,
    IS version:
    12 ... 24 V DC
  • Current consumption:
    12.5 mA

Standard:

  • 12 ... 250 V AC/DC, 0 ... 60 Hz,
    2 W max.

Digital:

  • Bus voltage:
    12 ... 30 V DC,
    IS version:
    12 ... 24 V DC
  • Current consumption:
    12.5 mA

Approvals

Stainless steel cable or enclosure version:

CE, CSA, FM, ATEX, RCM, Lloyds Register, WHG

Fully-synthetic version (PPS):

CSA, FM

CSA, FM, CE, ATEX, RCM, Lloyds Register, WHG, Vlarem II

 

CSA, FM, CE, ATEX, RCM, Lloyds Register, WHG, Vlarem II



 


















  © ООО "СК ЭНЕРГО" 2007-2022
  (4872) 700-366  skenergo@mail.ru
Яндекс.Метрика