General information19" rack unit- With 4 HU for installation
- in hinged frame
- In cabinets with or without telescope rails
- Front plate for service purposes can be pivoted down (laptop connection)
- Internal gas paths: stainless steel pipe (mat. no. 1.4571)
- Gas connections for sample gas inlet and outlet and for purging gas: fittings, pipe diameter of 6 mm or ?"
Field device- Two-door enclosure (IP65) with gas-tight separation of analyzer and electronics sections
- Individually purgeable enclosure halves
- Stainless steel gas path and stubs (mat. no. 1.4571)
- Purging gas connections: pipe diameter 10 mm or 3/8"
- Gas connections for sample gas inlet and outlet: clamping ring connection for a pipe diameter of 6 mm or ?"
Display and control panel- Large LCD panel for simultaneous display of:
- Measured value (digital and analog displays)
- Status bar
- Measuring ranges
- Contrast of LCD panel adjustable using menu
- Permanent LED backlighting
- Washable membrane keyboard with five softkeys
- Menu-driven operation for parameterization, test functions, adjustment
- User help in plain text
- Graphic display of concentration trend; programmable time intervals
- Bilingual operating software German/English, English/Spanish, French/English, Spanish/English, Italian/English
Input and outputs- One analog output per medium (from 0, 2, 4 to 20 mA; NAMUR parameterizable)
- Two analog inputs configurable (e.g. correction of cross-interference or external pressure sensor)
- Six digital inputs freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation)
- Six relay outputs, freely configurable (e.g. failure, maintenance request, threshold alarm, external magnetic valves)
- Expansion by eight additional digital inputs and eight additional relay outputs each (e.g. for autocalibration with up to four calibration gases)
CommunicationRS 485 present in basic unit (connection from the rear; for the rack unit also behind the front plate). Options - RS 485/RS 232 converter
- RS 485/Ethernet converter
- RS 485/USB converter
- Connection to networks via PROFIBUS DP/PA interface
- SIPROM GA software as the service and maintenance tool
 CALOMAT 6, membrane keyboard and graphic display Designs – parts wetted by sample gasGas path | 19" rack unit | Field device | Field device Ex |
|---|
With pipes | Bushing | Stainless steel, mat. no. 1.4571 | | | Pipe | Stainless steel, mat. no. 1.4571 | | | Sample cell body | Stainless steel, mat. no. 1.4571 | | | O-rings | FFKM-Chemraz | | | Sensor | Si, SiOxNy, AU, epoxy resin, glass | | | Tightness | Leakage < 1 ?l/s |
 CALOMAT 6, 19" rack unit, gas path  CALOMAT 6, field device, gas path
|
|