General information | Based on DIN EN 61207 / IEC 1207. All data based on digital gas mixture H2 in N2 |
Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Largest possible measuring span | 100 vol.% H2 (for smallest measuring span, see "Function") |
Measuring ranges with suppressed zero point | Any zero point within 0 ... 100 vol.% can be implemented, smallest possible measuring span: 5% H2 |
Operating position | Front wall, vertical |
Conformity | CE mark in accordance with EN 61326/A1 and EN 61010/1 |
Design, enclosure |
Degree of protection | IP20 according to EN 60529 |
Weight | Approx. 10 kg |
Electrical characteristics | |
EMC interference immunity (electromagnetic compatibility) All signal lines must be shielded. Measured value deviations of up to 4% of the smallest measuring range may occur in ranges with strong electromagnetic interference. | In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety | In accordance with EN 61010-1; overvoltage category II |
Auxiliary power (see nameplate) | 100 V -10% ... 120 V +10% AC, 48 ... 63 Hz or 200 V -10% ... 240 V +10% AC, 48 ... 63 Hz |
Power consumption | Approx. 20 VA |
Fuse values | 100 to 120 V: 1.0T/250 200 ... 240 V: 0.63 T/250 |
Gas inlet conditions | |
Sample gas pressure | 800 ... 1 100 hPa (absolute) |
Sample gas flow | 30 ... 90 l/h (0.5 ... 1.5 l/min) |
Sample gas temperature | Min. 0 to max. 50 °C, but above the dew point |
Temperature of the measuring cell | Approx. 60 ?C |
Sample gas humidity | < 90% relative humidity |
Dynamic response | |
Warm-up period | < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90) | < 5 s |
Damping (electrical time constant) | 0 ... 100 s, configurable |
Dead time (purging time of the gas path in the unit at 1 l/min) | Approx. 0.5 s |
Measuring response | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Output signal fluctuation | < ± 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (? = 0.25%) |
Zero point drift | < ± 1%/week of the smallest possible measuring span according to rating plate |
Measured-value drift | < ± 1%/week of the smallest possible measuring span according to rating plate |
Repeatability | < 1% of the current measuring range |
Detection limit | 1% of the current measuring range |
Linearity error | < ± 1% of the current measuring range |
Influencing variables | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Ambient temperature | < 1%/10 K referred to smallest possible measuring span according to rating plate |
Accompanying gases | Deviation from zero point (for influence of interfering gas see paragraph titled "Interference influences") |
Sample gas flow | < 0.2% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure | < 1% of the current measuring range with a pressure change of 100 hPa |
Auxiliary power | < 0.1% of the current measuring range with rated voltage ± 10% |
Electrical inputs and outputs | |
Analog output | 0/2/4 ... 20 mA, floating; load max. 750 ? |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated |
Analog inputs | 2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference |
Digital inputs | 6, designed for 24 V, isolated, freely parameterizable, e.g. for measurement range switchover |
Serial interface | RS 485 |
Options | AUTOCAL function each with 8 additional digital inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions | |
Permissible ambient temperature | -30 … +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity (dew point must not be fallen below) | < 90% relative humidity as annual average, during storage and transportation |