General (based on EN 61207/IEC 1207. All data refers to the binary mixture H2 in N2) |
Measuring ranges | 4, internally and externally switchable; automatic measuring range switchover also possible |
Largest possible measuring span | 100 vol.% H2 (for smallest measuring span, see "Function") |
Measuring ranges with suppressed zero point | Any zero point within 0 ... 100 vol.% can be implemented, smallest possible measuring span: 5 % H2 |
Operating position | Front wall, vertical |
Conformity | CE mark in accordance with EN 61326/A1 and EN 61010/1 |
Design, enclosure |
Degree of protection | IP20 according to EN 60529 |
Weight | Approx. 10 kg |
Electrical characteristics |
EMC (Electromagnetic Compatibility) (All signal lines must be shielded. Measured value deviations of up to 4 % of the smallest measuring range may occur in ranges with strong electromagnetic interference.) | In accordance with standard requirements of NAMUR NE21 (08/98) |
Electrical safety | In accordance with EN 61010-1; overvoltage category II |
Auxiliary power (see nameplate) | 100 V -10% ... 120 V +10% AC, 48 ... 63 Hz or 200 V -10% ... 240 V +10% AC, 48 ... 63 Hz |
Power consumption | Approx. 20 VA |
Fuse values | 100 ... 120 V: 1.0T/250 200 ... 240 V: 0.63 T/250 |
Gas inlet conditions |
Sample gas pressure | 800 ... 1 100 hPa (absolute) |
Sample gas flow | 30 ... 90 l/h (0.5 ... 1.5 l/min) |
Sample gas temperature | Min. 0 to max. 50 °C, but above the dew point |
Temperature of the measuring cell | Approx. 60 ?C |
Sample gas humidity | < 90% relative humidity |
Dynamic response |
Warm-up period | < 30 min (the technical specification will be met after 2 hours) |
Delayed display (T90) | < 5 s |
Damping (electrical time constant) | 0 ... 100 s, parameterizable |
Dead time (purging time of the gas path in the unit at 1 l/min) | Approx. 0.5 s |
Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
Output signal fluctuation | < ± 0.75 % of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (? = 0.25%) |
Zero point drift | < ± 1 %/week of the smallest possible measuring span according to rating plate |
Measured-value drift | < ± 1 %/week of the smallest possible measuring span according to rating plate |
Repeatability | < 1 % of the current measuring range |
Detection limit | 1 % of the current measuring range |
Linearity error | < ± 1 % of the current measuring range |
Influencing variable (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
Ambient temperature | < 1%/10 K referred to smallest possible measuring span according to rating plate |
Accompanying gases | Deviation from zero point (for influence of interfering gas see paragraph titled "Interference influences") |
Sample gas flow | < 0.2 % of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure | < 1 % of the current measuring range with a pressure change of 100 hPa |
Auxiliary power | < 0.1% of the current measuring range with rated voltage ± 10% |
Electrical inputs and outputs |
Analog output | 0/2/4 ... 20 mA, isolated; load max. 750 ? |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated |
Analog inputs | 2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference |
Binary inputs | 6, designed for 24 V, isolated, freely parameterizable, e.g. for measurement range switchover |
Serial interface | RS 485 |
Options | AUTOCAL function with 8 additional binary inputs and relay outputs each, also with PROFIBUS PA or PROFIBUS DP |
Climatic conditions |
Permissible ambient temperature | -30 … +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity (dew point must not be fallen below) | < 90% relative humidity as annual average, during storage and transportation |