Climatic conditions
|
|
|
Ambient temperature
|
+5° ... +40 °C (standard)
|
- With heater in sheet-steel cabinet
|
Min. -5 °C
|
- With heating in GRP cabinet
|
Min. -15 °C
|
|
|
Max. +52 °C
|
|
Relative humidity
|
75 % (annual average), non-condensing
|
|
Sample gas conditions
|
|
|
Max. sample gas pressure at inlet to sample preparation system
|
500 hPa (mbar)
|
|
Max. moisture content in sample gas
|
17 vol % (cooler type: CSS) 25 vol % (cooler type: EGK 2-19)
|
|
Sample gas temperature
|
Max. 200 °C at cabinet entry
|
|
Sample gas flow
|
Approx. 60 l/h per analyzer module
|
|
Sampling probe (standard)
|
Dust load: < 2 g/m3
|
|
Mounting flange: DN 65, PN 6, form B
|
|
Including temperature controller with PT100
|
|
With internal sampling tube, stainless steel, length: 1 m (can be shortened)
|
|
With filter in probe, to 600 °C
|
|
Sampling probe for higher temperatures and in other materials (optional)
|
On request
|
|
Sample gas line, electrically heated
|
Max. 35 m, longer lengths on request
|
|
Sample gas must not be flammable or explosive.
|
|
Power supply
|
|
|
Supply 1
|
115 V AC, 50 … 60 Hz (-15 %, +10 %)
|
|
Supply 2
|
230 V AC, 50 … 60 Hz (-15 %, +10 %)
|
|
Supply 3
|
400 V AC, 50 … 60 Hz (-15 %, +10 %)
|
|
Power
|
4 000 VA max.
|
|
System design
|
|
|
Fusing of electronic consumers
|
1-pole or 2-pole (selectable)
|
|
Sample gas cooler
|
2-stage
|
|
Output signals
|
4 … 20 mA (corresponding to the analyzer information)
|
|
Additional digital inputs and outputs via PLC (LOGO)
|
|
Color
|
RAL 7035
|
|
Weight
|
Approx. 160 kg
|
|
Sheet-steel cabinet/frame
|
Indoor installation
|
|
GRP cabinet
|
Outdoor installation
|
|
Explosion protection classification
|
Installation outside the Ex zone
|
|
Degree of protection
|
IP 54 (avoid direct sunlight for outdoor installation)
|
|
Calibration
|
Semi-automatic; AUTOCAL freely adjustable on U23 up to max. 24-hour interval
|
|
Detailed information on the analyzers
You can find detailed information on the analyzers:
- in the "Continuous Gas Analyzers, extractive" chapter
- ULTRAMAT 23
- SIPROCESS UV600
|
|
Dimensions
|
|
|
Sheet-steel cabinet (with base) for indoor installation
|
2 000 x 800 x 800 mm (H x W x D)
|
|
GRP cabinet (with base) for indoor installation
|
2 060 x 900 x 800 mm (H x W x D)
|
|
500 mm spacing on the right or left must be provided for the cable inlet and connection of the heated sample gas line.
|