Tested component design
The complete system consists of the following tested individual components:
- Sampling probe: M&C, type: SP2000; Buhler/Siemens, type: GAS222/7MB1943-2F
- Heated sample gas line: Winkler/Siemens, type: 7MB1943-2A
- Temperature controller: Siemens, type: SIRIUS
- Two-stage compressor gas cooler: M&C, type: CSS; Buhler, type: EGK 2-19
- Sample gas pump: Buhler/Siemens, type: P2.3/7MB1943-3C
- NOX converter: M&C, type: CG-2; Buhler/Siemens, type: BUENOX/7MB1943-2D
Performance-tested measuring ranges
|
Component
|
Analyzer module
|
Smallest certified measuring range
|
Additional measuring ranges
|
|
CO
|
(in test phase)
|
U23-7MB2355
|
0 ... 150 mg/m?
|
0 ... 750 / 3 000 mg/m?
|
| |
(in test phase)
|
U23-7MB2357
|
0 ... 150 mg/m?
|
0 ... 750 / 3 000 mg/m?
|
| |
U23-7MB2358
|
0 ... 250 mg/m?
|
0 ... 1 250 mg/m?
|
|
NO
|
(in test phase)
|
U23-7MB2355
|
0 ... 100 mg/m?
|
0 ... 750 / 2 000 mg/m?
|
| |
(in test phase)
|
U23-7MB2357
|
0 ... 100 mg/m?
|
0 ... 750 / 2 000 mg/m?
|
| |
U23-7MB2358
|
0 ... 400 mg/m?
|
0 ... 2 000 mg/m?
|
| |
(in test phase)
|
UV600
|
0 ... 50 mg/m?
|
0 ... 100 / 1 000 / 2 000 mg/m?
|
|
SO2
|
(in test phase)
|
U23-7MB2355
|
0 ... 400 mg/m?
|
0 ... 2 000 mg/m?
|
| |
(in test phase)
|
U23-7MB2357
|
0 ... 400 mg/m?
|
0 ... 2 000 mg/m?
|
| |
U23-7MB2358
|
0 ... 400 mg/m?
|
0 ... 2 000/7 000 (in operation) mg/m?
|
| |
(in test phase)
|
UV600
|
0 ... 75 mg/m?
|
0 ... 130 / 1 500 / 2 000 mg/m?
|
|
NO2
|
(in test phase)
|
UV600
|
0 ... 50 mg/m?
|
0 ... 500 mg/m?
|
|
O2, (paramag.)
|
U23
|
0 ... 25 vol %
|
|
|
O2, (electrochem.)
|
U23
|
0 ... 25 vol %
|
|
|
|