Set CEM CERTTested component designThe complete system consists of the following tested individual components: - Sampling probe: M&C, type: SP2000; Buhler/Siemens, type: GAS222/7MB1943-2F
- Heated sample gas line: Winkler/Siemens, type: 7MB1943-2A
- Temperature controller: Siemens, type: SIRIUS
- Two-stage compressor gas cooler: M&C, type: CSS; Buhler, type: EGK 2-19
- Sample gas pump: Buhler/Siemens, type: P2.3/7MB1943-3C
- NOX converter: M&C, type: CG-2; Buhler/Siemens, type: BUENOX/7MB1943-2D
Performance-tested measuring rangesComponent | Analyzer module | Smallest certified measuring range | Additional measuring ranges | CO | | U23-7MB2355 | 0 ... 200 mg/m? | 0 ... 750 / 3 000 mg/m? | | | | U23-7MB2357 | 0 ... 200 mg/m? | 0 ... 750 / 3 000 mg/m? | | | U23-7MB2358 | 0 ... 250 mg/m? | 0 ... 1 250 mg/m? | NO | | U23-7MB2355 | 0 ... 150 mg/m? | 0 ... 750 / 2 000 mg/m? | | | | U23-7MB2357 | 0 ... 150 mg/m? | 0 ... 750 / 2 000 mg/m? | | | U23-7MB2358 | 0 ... 400 mg/m? | 0 ... 2 000 mg/m? | | | (in test phase) | UV600 | 0 ... 50 mg/m? | 0 ... 100 / 1 000 / 2 000 mg/m? | SO2 | | U23-7MB2355 | 0 ... 400 mg/m? | 0 ... 2 000 mg/m? | | | | U23-7MB2357 | 0 ... 400 mg/m? | 0 ... 2 000 mg/m? | | | U23-7MB2358 | 0 ... 400 mg/m? | 0 ... 2 000/7 000 (in operation) mg/m? | | | (in test phase) | UV600 | 0 ... 75 mg/m? | 0 ... 130 / 1 500 / 2 000 mg/m? | NO2 | (in test phase) | UV600 | 0 ... 50 mg/m? | 0 ... 500 mg/m? | O2, (paramag.) | U23 | 0 ... 25 vol % | | O2, (electrochem.) | U23 | 0 ... 25 vol % | |
|
|