General information
|
Weight
|
Approx. 5.5 kg (standard version)
|
Measuring ranges
|
Number of measuring ranges
|
Max. 4; parameters can be assigned freely
|
Parameters can be assigned in the measuring ranges
|
|
- Smallest possible measuring spans
|
0.5% (? 1% for high-temperature model), 2% or 5% O2
|
- Largest possible measuring span
|
100% O2
|
Gas inlet conditions
|
Sample gas pressure
|
|
|
500 … 1 500 hPa (abs.)
|
|
|
- Without vibration compensation
|
500 to 3 000 hPa (abs.); short-term max. 5 000 hPa (abs.)
|
- With vibration compensation
|
500 to 2 500 hPa (abs.); short-term max. 5 000 hPa (abs.)
|
Correction of the internal pressure sensor
|
|
|
500 … 1 450 hPa (abs.)
|
|
500 … 2 450 hPa (abs.)
|
Reference gas pressure
|
|
|
0.2 to 0.4 MPa above the sample gas pressure, but a maximum of 0.5 MPa (absolute)
|
- Without vibration compensation
|
2 000 … 3 500 hPa above sample gas pressure; max. 5 000 hPa (abs.)
|
- With vibration compensation
|
2 500 … 4 000 hPa above sample gas pressure; max. 5 000 hPa (abs.)
|
- Low-pressure connection with external reference gas pump (only for sample gas pressure 500 to 1 500 hPa (absolute))
|
100 hPa above the sample gas pressure
|
Pressure loss between sample gas inlet and sample gas outlet
|
< 100 hPa at 1 l/min
|
Sample gas flow
|
18 … 60 l/h (0.3 … 1 l/min)
|
Sample gas temperature
|
0 … 60 °C
|
Sample gas humidity (rel. humidity)
|
< 90% (condensation inside the gas path is to be avoided)
|
Sample chamber temperature
|
Standard version
|
Approx. 72?
|
Time response
|
Warm-up period at room temperature
|
< 2 h
|
Dead time (T10)
|
< 0.5 s at 1 l/min
|
Signal rise time or fall time for a flow rate of 1 l/min, a static attenuation constant and a dynamic attenuation constant of 0 s
|
< 1 s
|
Time for device-internal signal processing
|
approx. 1 s
|
Delayed display T90
|
T90 < T10 + rise or fall time + signal processing time
|
Measuring response
|
Output signal fluctuation
|
? 0.5 % of the current measuring span (6 ? value) for a static attenuation constant of 0 s and a dynamic attenuation setting of 5 % / 10 s (with activated vibration compensation: 1.5 times the value
|
Detection limit
|
? 1% of smallest measuring span according to nameplate (with vibration compensation activated: 1.5 times the value)
|
Measured-value drift
|
? 0.5%/month of current measuring span or ? 50 vpm oxygen, whichever is larger
|
Repeatability
|
? 0.5 % of current measuring span
|
Linearity error
|
? 0.1 % with ambient air as reference gas
|
Influencing variables
|
Ambient temperature
|
|
|
? 0.5% of smallest measuring span according to nameplate/10 K or ? 50 vpm O2/10 K, whichever is larger
|
|
? 0.5% of the current measuring span/10 K or ? 50 vpm O2/10 K, whichever is larger
|
Sample gas pressure
|
|
- Without pressure compensation
|
Deviation approx. 2% of current measuring span/1% pressure variation
|
- With pressure compensation switched on
|
? 0.2% of the current measuring span/1% pressure variation or ? 50 vpm O2/1% pressure variation, whichever is larger
|
Sample gas flow
|
? 1% of the current measuring span with a flow rate change of 0.1 l/min within the permissible flow range (0.3 to 1 l/min)
|
Accompanying gases
|
Zero point deviation (cross-sensitivity) in accordance with Table A.1 of EN 61207-3
|
Supply voltage (fluctuations of the supply voltage of the basic unit*) in the range of 90 to 253 V AC/47 to 63 Hz)
|
? 0.1% of full-scale value of characteristic
|
Electrical inputs and outputs
|
Analog and digital interfaces
|
See basic unit
|
Gas connections
|
With hoses
|
Plastic screw connection for plastic pipe or tube 4 mm/6 mm
|
With pipes
|
Connection for threaded joint; ISO female thread 1/8"
|
Climatic conditions
|
Storage and transport
|
?30 … 70°C
|
Permissible ambient temperature (for operation in basic unit)
|
0 … 50?
|
Relative humidity (RH) during storage, transport or operation
|
< 90% (condensation from the installed components is to be avoided)
|
Materials of wetted parts
|
Sample chamber
|
Stainless steel:
- Plates: Mat. No. 1.4571 (X6CrNiMoTi 17-12-2)
- Screw-in glands: Mat. No. 1.4404 (X2CrNiMo17-12-2)
Hastelloy C22:
- Plates: Mat. No. 2.4602 (NiCr21Mo14W)
- Screw-in glands: Mat. No. 2.4819 (NiMo16Cr15W)
|
Gas path
|
|
|
FPM (e.g. Viton), connections PVDF
|
|
Stainless steel:
- Pipes: Mat. No. 1.4571 (X6CrNiMoTi 17-12-2)
- Gas connections: Mat. No. 1.4404 (X2CrNiMo 17-12-2)
Hastelloy C22:
- Pipes: Mat. No. 2.4602 (NiCr21Mo14W)
- Gas connections: Mat. No. 2.4819 (NiMo16Cr15W)
|
Sealing material
|
FPM (e.g. Viton) or FFKM Compound 2035 (e.g. Kalrez 2035 (see device certificate))
|
Special applications
|
Gas path
|
|
|
Materials adapted to the application
|