General information |
Weight | Approx. 5.5 kg (standard version) |
Measuring ranges |
Number of measuring ranges | Max. 4; parameters can be assigned freely |
Parameters can be assigned in the measuring ranges | |
- Smallest possible measuring spans
| 0.5%, 1%, 2% or 5% O2 |
- Largest possible measuring span
| 100% O2 |
Gas inlet conditions |
Sample gas pressure | |
- Standard devices with hoses
| 500 … 1 500 hPa (abs.) |
- Tolerance within permitted pressure range
| ± 150 hPa |
- Standard devices with hoses and ext. RG pump
| 700 to 1200 hPa (abs.) |
- Tolerance within permitted pressure range
| ± 150 hPa |
- Standard devices with pipes
| 500 to 3 000 hPa (abs.); short-term < 5000 hPa (abs.) |
- Tolerance within permitted pressure range
| ± 150 hPa |
Reference gas pressure | |
| 2000 hPA above sample gas pressure (within permitted reference gas pressure range 2500 to 5000 hPa, abs.) |
- Low-pressure connection with external reference gas pump
| |
- Without vibration compensation
| 100 hPa above sample gas pressure |
- With vibration compensation
| 150 hPa above sample gas pressure |
Pressure loss between sample gas inlet and sample gas outlet | < 100 hPa at 1 l/min |
Sample gas flow | 18 … 60 l/h (0.3 … 1 l/min) |
Sample gas temperature | 0 … 60 °C |
Sample gas humidity (rel. humidity) | < 90% (condensation inside the gas path is to be avoided) |
Sample chamber temperature |
Standard version | Approx. 72 °C |
Time response |
Warm-up period at room temperature | < 2 h |
Response characteristics | |
- Display delay T90 with an electronic damping setting of 0 s and a sample gas flow of 1 Nl/min.
| ? 1.9 s |
| ? 1.1 s |
Measuring response |
Output signal fluctuation with static damping constant of 0 s and dynamic noise suppression of 5% / 10 s | ? ±0.5% of smallest measuring span (noise bandwidth corresponds to 1% = 6? value or 0.333% = 2? value), with vibration compensation activated: < 1.5 times the value |
Detection limit | ? 1% of smallest measuring span according to nameplate (with vibration compensation activated: < 1.5 times the value) |
Measured-value drift | |
| ? ±0.5% of the smallest span/month or ? ±50 vpm O2/month, whichever is greater |
| ? ±0.5% of the current measuring span/month or ? ±50 vpm O2/month, whichever is greater |
Repeatability | |
| ? ±0.5% of the smallest measuring span/month or ? ±50 vpm O2/month, whichever is greater |
| d? ±0.5% of the current measuring span/month or ? ±50 vpm O2, whichever is greater |
Linearity error with dry ambient air1) | < 0.1% |
Influencing variables |
Ambient temperature | |
| ? 0.5% of the smallest measuring span / 10 K or ? 50 vpm O2/10 K, whichever is greater |
- Deviation of the span gas
| ? 0.5% of the current measuring span / 10 K or ? 50 vpm O2/10 K, whichever is greater |
Sample gas pressure | |
| ? 0.2% of the smallest measuring span / 1% pressure variation or ? 50 vpm O2/1% pressure variation, whichever is greater |
- Deviation of the span gas
| ? 0.2% of the current measuring span / 1% pressure variation or ? 50 vpm O2/1% pressure variation, whichever is greater |
Sample gas flow | |
| ? 1% of smallest measuring span per 0.1 l/min change in flow or ? 50 vpm O2 per 0.1 l/min change in flow within the permissible flow range (0.3 to 1 l/min), whichever is greater |
- Deviation of the span gas
| ? 1% of current measuring span per 0.1 l/min change in flow or ? 50 vpm O2 per 0.1 l/min change in flow within the permissible flow range (0.3 to 1 l/min), whichever is greater |
Accompanying gases | Zero point deviation (cross-sensitivity) in accordance with Table A.1 of EN 61207-3 |
Supply voltage | < 0.1% of the current measuring span (within the nominal range of use) |
Electrical inputs and outputs |
Analog and digital interfaces | See base unit |
Gas connections |
| Pipe connection with 6 mm outer diameter |
Climatic conditions |
Storage and transport | ?30 … 70°C |
Permissible ambient temperature (during operation in base unit)2) | 0 … 50? |
Relative humidity (RH) during storage, transport or operation | < 90% (condensation from the installed components is to be avoided) |
Materials of wetted parts |
Sample chamber | Stainless steel: - Plates: Mat. No. 1.4571 (X6CrNiMoTi 17-12-2)
- Screw-in glands: Mat. No. 1.4404 (X2CrNiMo17-12-2)
|
Internal gas path | |
| FPM (e.g. Viton), gas connections of stainless steel material no. 14404 (X2CrNiMo 17-12-2) |
| Stainless steel: - Pipes: Mat. No. 1.4571 (X6CrNiMoTi 17-12-2)
- Gas connections: Mat. No. 1.4404 (X2CrNiMo 17-12-2)
|
Sealing material | FPM (e.g. Viton) or FFKM Compound 2035 (e.g. Kalrez 2035 (see device certificate)) |
Special applications |
Gas path | |
| Materials adapted to the application |