Ambient temperature
|
0 ... 35 °C
|
- With heater in sheet-steel cabinet
|
Max. -5 °C
|
- With heater in FRP cabinet
|
Max. -15 °C
|
|
|
Max. 45 °C
|
|
Relative humidity
|
70 %, non-condensing
|
|
Corrosive atmosphere
|
No
|
|
Gas inlet conditions
|
|
|
Max. sample gas pressure at inlet to sample preparation system
|
0.5 barg
|
|
Min. sample gas temperature at inlet to sample preparation system
|
180 °C
|
|
Dust content at inlet to sample preparation system
|
Dust-free
|
|
Sampling probe
|
Sampling tube 20 x 1.5, 1 000 mm long, stainless steel, flange: DN 65, PN 6
|
|
Max. sample gas pressure at sampling probe
|
0.5 barg
|
|
Max. sample gas temperature at sampling probe
|
600 °C
|
|
Max. dust content at sampling probe
|
2 g/Nm?
|
|
Sample gas must not be flammable or explosive.
|
|
Power supply
|
|
|
Supply 1
|
115 V AC (-15%, +10%)
|
|
Supply 2
|
230 V AC (-15%, +10%)
|
|
Supply 3
|
400 V AC (-15%, +10%)
|
|
Connections
|
|
|
Hose material
|
Teflon
|
|
Cables
|
Not armored, not halogen-free
|
|
Electrical design
|
According to IEC
|
|
Cable ID
|
Individual core labeling as option
|
|
Fusing of electronic consumers
|
1-pole; 2-pole as option
|
|
Duplication of analog signals
|
Not electrically isolated as option
Electrically isolated as option
|
|
Installation
|
|
|
Site
|
|
- In sheet-steel cabinet/frame
|
Indoor installation
|
|
|
Outdoor installation
|
|
Ex zone
|
Non-Ex area
|
|
System design
|
|
|
Version
|
Mounting frame or cabinet
|
|
Cabinet degree of protection
|
IP54
|
|
Automatic calibration
|
Yes, with ULTRAMAT 23
|
|
Detailed information on the analyzers
Process analytics, topics "Continuous gas analyzers, extractive" ULTRAMAT 23, OXYMAT 6 and "Continuous gas analyzers, in-situ" LDS 6
|
|
Dimensions (without plinth)
|
|
|
Depth of sheet-steel frame
|
|
|
|
2 000 x 800 x 800 mm (H x W x D)
|
|
|
2 000 x 800 x 600 mm (H x W x D)
|
|
FRP cabinet (with plinth)
|
2 080 x 900 x 600 mm (H x W x D)
|
|
It is necessary to allow a space of 500 mm on the left or right for introduction of the cables.
|