Ambient temperature | 0 ... 35 °C |
- With heater in sheet-steel cabinet
| Max. -5 °C |
- With heater in FRP cabinet
| Max. -15 °C |
| Max. 45 °C |
Relative humidity | 70 %, non-condensing |
Corrosive atmosphere | No |
Gas inlet conditions | |
Max. sample gas pressure at inlet to sample preparation system | 0.5 barg |
Min. sample gas temperature at inlet to sample preparation system | 180 °C |
Dust content at inlet to sample preparation system | Dust-free |
Sampling probe | Sampling tube 20 x 1.5, 1 000 mm long, stainless steel, flange: DN 65, PN 6 |
Max. sample gas pressure at sampling probe | 0.5 barg |
Max. sample gas temperature at sampling probe | 600 °C |
Max. dust content at sampling probe | 2 g/Nm? |
Sample gas must not be flammable or explosive. |
Power supply | |
Supply 1 | 115 V AC (-15%, +10%) |
Supply 2 | 230 V AC (-15%, +10%) |
Supply 3 | 400 V AC (-15%, +10%) |
Connections | |
Hose material | Teflon |
Cables | Not armored, not halogen-free |
Electrical design | According to IEC |
Cable ID | Individual core labeling as option |
Fusing of electronic consumers | 1-pole; 2-pole as option |
Duplication of analog signals | Not electrically isolated as option Electrically isolated as option |
Installation | |
Site | |
- In sheet-steel cabinet/frame
| Indoor installation |
| Outdoor installation |
Ex zone | Non-Ex area |
System design | |
Version | Mounting frame or cabinet |
Cabinet degree of protection | IP54 |
Automatic calibration | Yes, with ULTRAMAT 23 |
Detailed information on the analyzers Process analytics, topics "Continuous gas analyzers, extractive" ULTRAMAT 23, OXYMAT 6 and "Continuous gas analyzers, in-situ" LDS 6 |
Dimensions (without plinth) | |
Depth of sheet-steel frame | |
| 2 000 x 800 x 800 mm (H x W x D) |
| 2 000 x 800 x 600 mm (H x W x D) |
FRP cabinet (with plinth) | 2 080 x 900 x 600 mm (H x W x D) |
It is necessary to allow a space of 500 mm on the left or right for introduction of the cables. |