Gasmet CEMS
|
General data
|
Measuring principle
|
- Fourier transform infrared (FTIR) spectroscopy
- for oxygen: ZrO2 measuring cell (optional)
- for total carbon (Ctotal, THC, TOC)
- FTIR, not performance-tested (standard)
- FID analyzer, also performance-tested in accordance with EN 14181 if required (option)
|
Performance capability
|
Simultaneous analysis of up to 50 gas components Standard: 16 components
|
T90 time
|
< 180 s, with 20 m sample gas line
|
Climatic conditions
|
|
- Ambient temperature during operation
|
0 ... 40 °C, non-condensing
|
- Ambient conditions during storage and transportation
|
-20 ... 60 °C, 0 ... 95% relative humidity, no condensation!
|
|
Indoors, dust-free, clean ambient air, no external vibrations
|
Ex zone
|
Non-Ex area (IEC General Purpose)
|
System design
|
|
|
Cabinet configuration
|
- Cabinet degree of protection
|
IP54
|
Electrical connections
|
|
- Supply voltage (alternative 1)
|
3 phases 115 V AC (-15 %, +10 %) / 60 Hz, 32 A + neutral + ground (TN-S)
|
- Supply voltage (alternative 2)
|
400 V AC (-15 %, +10 %) /50 Hz, 3 phases 230 V AC (-15 %, +10 %) / 50 Hz, 16 A + neutral + ground (TN-S)
|
|
Gasmet CEMS with 20 m heated sample gas line: approx. 7.5 kW
|
Dimensions (with plinth)
|
2 100 x 600 x (800 + 250) mm (H x W x D)
|
Weight
|
Gasmet CEMS with FTIR and O2 analyzers: Approx. 320 kg *)
|
Electrical inputs/outputs
|
|
Analog inputs/outputs
|
|
|
16 channels, 4 … 20 mA, isolated, freely programmable
|
|
8 channels, 4 … 20 mA, isolated, freely programmable
|
Digital inputs/outputs
|
|
|
16 channels, 24 V DC, freely programmable
|
|
System alarm, maintenance requirement, maintenance, measured value overshoot, valid measured value
|
|
16 channels, floating contacts
|
|
Temperature alarm sampling probe, temperature alarm analysis cabinet, alarm air-conditioning unit on the analysis cabinet, alarm zero gas pressure, alarm O2 analyzer, activated scan gas infeed via sampling probe, activated system standby
|
Optional expansions
|
Up to 255 terminal strips
|
Interfaces (optional)
|
RS 232, RS 422/485, Modbus, PROFIBUS
|
Sampling unit
|
|
Temperature controller for heated sample gas lines
|
0 … 180 °C
|
Type of temperature sensor in sample gas lines
|
K type (Ni-Cr-Ni element)
|
Gas inlet conditions
|
|
- Max. sample gas pressure at inlet to sample preparation system
|
1 500 hPa
|
- Sample gas temperature at inlet to sample preparation system
|
180 … 200 °C, non-condensing
|
- Dust content at inlet to sample preparation system
|
Dust-free
|
|
Oil-free and dust-free instrument air, dew point < ?40 °C
|
|
- Sample gas: 2 ... 5 l/min
- Zero gas: Approx. 0.5 l/min
- Span gas: Approx. 0.5 l/min
- Instrument air to rinse the FTIR spectrometer:
- Normal mode: Approx. 0.5 l/min
- Safety purging: Approx. 4 l/min
- Reference air for optional O2 measurement: Approx. 0.7 l/min
|
|
Sample gas (heated to 180 °C), span gas, zero gas, instrument air
|
|
FTIR analyzer (heated), O2 analyzer (heated), span gas for sample probe, interferometer purging, reference air for O2 analyzer
|
|
Swagelok 6 mm, for sample gas lines with 4/6 mm internal/external diameters of the Teflon core. The external diameter of the sample gas line including preassembly on the side of the connection to the sampling unit must be no more than 43 mm.
|
Sample gas pump
|
|
|
Stainless steel 316
|
|
PTFE
|
|
4 l/min
|
|
For particles > 2 µm
|
Valves
|
|
|
Sample gas / backflushing, span gas via sample outlet, span gas via sample probe, zero gas, instrument air (safety circuit)
|
|
Sample gas/backflushing, gas infeed for maintenance purposes
|
FTIR spectrometer unit
|
|
Measuring principle
|
Fourier transform infrared (FTIR) spectroscopy
|
Spectral resolution
|
Standard: 8 cm-1
|
Scan rate
|
10 scans / s
|
Detector
|
Peltier cooled MCT
|
Infrared light source
|
SiC, 1 550 K
|
Beam splitter and window material
|
ZnSe
|
Spectral range
|
4 200 … 900 cm-1 (2.4 … 11 µm)
|
FTIR gas cuvette
|
Multiple path gas cell with fixed optical path length: 5.0 m
|
|
Standard: 100% rhodium-coated aluminum
|
|
Gold coating
|
|
0.4 liters
|
|
Standard: Viton O rings, others: On request
|
Window material
|
BaF2
|
Interferometer purging
|
- Oil-free and particle-free instrument air, dew point < ?40 °C, 0.5 l/min
- Nitrogen (at least 5.0)
|
Measuring performance
|
|
|
Recommendation: Automatic "zero point calibration" with nitrogen every 24 hours (Quality of the nitrogen: 5.0 or higher)
|
|
< 2 % of the measuring range of the "zero point calibration"
|
|
< 2 % of the measuring range
|
|
< 2 % of the measuring range per 10 K temperature change
|
|
1 % of the measured value at 1% sample gas pressure change. Ambient pressure fluctuations are measured and compensated for
|
Oxygen analyzer (optional)
|
|
Measuring principle
|
ZrO2 measuring cell
|
Detection limit
|
< 1 ppm O2
|
Reference gas
|
- Oil-free and particle-free instrument air, dew point < ?40 °C
- Span gas
|
Sensor material
|
Stainless steel 316 (1.4571)
|
Gas flow
|
Minimum 0.1 l/min
|
Measuring ranges
|
Min. 0 … 2 vol.% O2 Type 0 … 25 vol.% O2
|
Accuracy
|
± 0.2% of the measuring range
|
FIDAMAT 6 (optional)
|
See "Continuous gas analysis, extractive"
|
Measuring principle
|
FID analyzer
|