? The Gasmet CEMS is a fully equipped FTIR analysis system comprising several 19'' rack units:
Options for expanding the system include:
Sampling probeThe Gasmet CEMS is supplied as standard without a sampling probe. A suitable stainless steel (SS 316) probe with DN 65 PN 6 flange that can be heated to 180 °C is recommended as a sampling probe for the Gasmet CEMS. This can be offered separately on request. Since particles of dust can usually be expected in sample gas, you are strongly recommended to use a sampling probe with a suitable (pre-)filter element and/or backflushing facility, especially with dust concentrations > 2 g/m3. The sample gas temperature should usually not exceed 600 °C; the sample gas pressure must be between 400 and 6 000 hPa. For performance-tested applications, a sampling probe matched to the Gasmet CEMS must be used (e.g. M&C SP2000H or comparable). This can be offered separately on request. Heated sample gas lineThe Gasmet CEMS is supplied as standard with heated sample gas lines between the sampling unit in the Gasmet CEMS analysis cabinet and the FTIR analyzer. Further internal sample gas lines to the optional ZrO2 and/or FID analyzers can be ordered additionally. The external sample gas line from the sampling probe to the sampling unit in the analysis cabinet of the Gasmet CEMS is not included in the standard scope of delivery of the system. All heated sample gas lines (external and internal) are regulated to 180 °C in the temperature controller integrated into the Gasmet CEMS sampling unit. The external sample gas line can be heated as required and must therefore be designed for approx. 200 °C. It must have a power consumption of 120 VA per meter, and it must use K-type temperature sensors (Ni-Cr-Ni). The length of the heated line must not exceed 30 m. Lengths greater than 30 m can be provided on special request, or they can be implemented via an additional temperature control unit. For the external, heated sample gas line, a replaceable inner lining made from PTFE with a 4 mm inner diameter and 6 mm outer diameter is recommended. The connections must be Swagelok 6 mm. For steric reasons, the outer diameter of the sample gas line including preassembly on the side of the connection to the Gasmet CEMS sampling unit must be less than 43 mm. Gasmet CEMS sampling unitThe sampling unit is designed for hot-extractive sampling of wet, corrosive gas streams, and is integrated into the Gasmet CEMS as a 19" rack unit. It encompasses all valves, flow meters, sample gas pumps and filters required for the measuring task. The temperature of all parts coming into contact with the sample gas (pumps, filters and lines) is stabilized to 180 ?C. The sample gas does not have to be diluted and does not have to be dried beforehand either. The gas flow and the pressure of the instrument air supply can be monitored via the Gasmet CEMS sampling unit by means of an alarm. The sampling unit is connected to the control PC and is controlled using the Calcmet software, like the FTIR and the other (optional) analyzers. Gasmet CEMS sampling unit with removed filter insert Rear with connections of the Gasmet CEMS sampling unit Analysis cabinetThe analysis cabinet of the Gasmet CEMS contains all the essential components in the form of 19'' rack units, including cabling and tubing:
The external dimensions of the analysis cabinet with socket, including the air-conditioning unit mounted on the rear door, are: 2 100 x 600 x (800 + 250) mm (H x W x D). Hose and cable inlets are possible from the top, below, left and right. At the installation location, sufficient space must be provided to the front, the right and the rear of the analysis cabinet to allow opening of the front and rear doors, and facilitate access to the rear door. The air-conditioning unit mounted on the rear also requires sufficient space for air circulation. To be able to pull out the 19'' rack units to the front on the telescopic extensions, the electrical cables and sample gas lines in the cabinet must be sufficiently long and flexible. The same applies to the lines to the CEMS analysis cabinet if it has to be moved on optional rollers for service and maintenance purposes, for example. Minimum space requirements for a Gasmet CEMS for facilitating installation and maintenance work External panels and doorsThe side panels and the rear door of the Gasmet CEMS analysis cabinet are designed in steel plate, and the front door in glass. BaseThe analysis cabinet of the Gasmet CEMS is offered optionally on a plinth or four rollers (two with immobilizing brakes). Air-conditioning unitTo prevent overheating of the electronic components inside the cabinet, the analysis cabinet is supplied as standard with an air-conditioning unit. The air-conditioning unit is attached to the rear door and has a cooling capacity of 1 100 W. Condensation removalThe Gasmet CEMS analysis system is offered as standard with a plastic tube for removal and disposal of condensation by the customer. Power supplyThe system can be designed either for three phases 115 V AC at 60 Hz, 32 A or 400 V AC, or three phases 230 V AC at 50 Hz, 16 A. Three phases, neutral and ground must always be provided by the customer. Analog signal processingThe analog signals are connected as single wires on isolating terminals as standard. AnalyzersThe analysis cabinet is prepared for accommodating one to three analyzers with hot sample feed. As well as the standard Gasmet CEMS FTIR spectrometry unit, the system can also be supplemented with a ZrO2 oxygen analyzer, as well as a FIDAMAT 6. Different measured components and measuring ranges are available for selection. Other combinations of measured components and measuring ranges are available on request, but you must check that the desired certificates and approvals are available. The analyzers, measured components and measuring ranges used are described briefly below. GASMET FTIR spectrometer For measuring 16 infrared active components as standard. A calibrated measuring range is stored as standard per gas component. Further calibrated measuring ranges per gas component are available upon request for an additional charge.
Standard components and measuring ranges Gasmet CEMS Gasmet CEMS FTIR rack unit, front and rear ZrO2 oxygen analyzer
Electrical and gas connections of the ZrO2 analyzer FIDAMAT 6 The FIDAMAT 6 is used for FID-based measurement of the total carbon content (Ctotal, THC, TOC value).
You can find details of the optional FIDAMAT 6 analyzer in the "Continuous gas analyzers, extractive" chapter of this catalog.
Performance-tested measuring ranges of the Gasmet CEMS analysis system Other combinations of measured components and measuring ranges are available on request, as well as other sample gas components. In this case, you must check that the desired certificates and approvals are available. Control and analysis unitAn industrial PC with Windows operating system and RAID hard drives is used in the Gasmet CEMS analysis system for controlling the sampling unit and the analyzers, as well as for data analysis and forwarding of the analysis data to the process control system. The GASMET Calcmet software is used as the control and analysis software. Operator interface of the Calcmet software of the Gasmet CEMS analysis system The standard configuration enables control and monitoring of the analyzers via the serial interface, watchdog support, as well as remote monitoring and maintenance options for the service and maintenance personnel. Siemens provides automation and drive products with industrial security functions that support the secure operation of plants or machines. They are an important component in a holistic industrial security concept. With this in mind, our products undergo continuous development. We therefore recommend that you keep yourself informed with respect to our product updates. Please find further information and newsletters on this subject at: http://support.automation.siemens.com To ensure the secure operation of a plant or machine it is also necessary to take suitable preventive action (e.g. cell protection concept) and to integrate the automation and drive components into a state-of-the-art holistic industrial security concept for the entire plant or machine. Any third-party products that may be in use must also be taken into account. Please find further information at: http://www.siemens.com/industrialsecurity Zero gas and calibration gas supplyThe Gasmet CEMS analysis system is equipped as standard with two solenoid valves for the automatic infeed of zero gas and calibration gas. Either nitrogen (quality: at least 5.0) or purified instrument air is used as the zero gas for the FTIR analyzer, which must be fed in at a pressure of 2 000 hPa. The span gas must contain one or more of the gas components to monitored, depending on requirements and/or official regulations. The span gas must also be fed in at a pressure of approx. 2 000 hPa. Instrument air supplyThe FTIR interferometer of the Gasmet CEMS is purged continuously with instrument air at an even flow rate of approx. 0.5 l/min. All internal gas routes of the Gasmet CEMS upstream of the FTIR spectrometer must be kept free of all contamination using clean purge gas before switching on the analyzer with sample gas infeed, or in good time before switching off temperature stabilization of the sample gas lines and the analyzer. This prevents condensation of sample gas components, airborne pollutants, and atmospheric humidity, and/or the accumulation of particles inside the FTIR spectrometer. This prevents irreversible damage to the internal gas cuvette. Either nitrogen (quality: at least 5.0) or completely oil-free and particle-free instrument air is used as the purge gas. To enable a "safety purge" of the internal gas routes in particular, the instrument air must always be available at a pressure of approx. 6 bar. Options/expansions (on request)
|
skener.ru Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Арматура DENDOR Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Датчики и измерители Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Регуляторы и регистраторы Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Пневматическое оборудование Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Краны и Клапаны Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Измерительные приборы Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Системы беспроводного управления «умный дом» Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Бесконтактные выключатели Конечные выключатели Оптические датчики Энкодеры Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 SKW-FS - Установка умягчения Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30 SKW-FK - Установка обезжелезивания |
Каталог 2018 | Каталог 2017 | Каталог 2016 | Каталог 2015 | Каталог 2014 | Каталог 2013 | Каталог 2012 | Сертификат | Контакты | Карта сайта | Поиск |