19" rack unit
- 19" rack unit with 3 height units (HU) for installation
- in hinged frames
- in cabinets with or without telescopic rails
- Gas connections for sample gas inlet and outlet: for pipe diameter 6 mm or 1/4"
- Purging gas connections (optional), purging gas connection for 6 mm or 1/4" hose (optional)
Wall-mounted device
- Gas connections for sample gas inlet and outlet: Pipe union for pipe diameter 6 mm or 1/4" (directly on the analyzer modules)
- Purging gas connections: Pipe diameter 6 mm or 1/4" hose (optional)
Display and operator panel
- LCD panel for simultaneous display of:
- Measured value
- Status line
- Measuring ranges
- Menu-driven operation for parameterization, test functions, adjustment
- Operator support in plain text
- Operating software (11 languages)
Display and operator panel of the SIPROCESS GA700 devices
Inputs and outputs
- 8 binary inputs, designed for 24 V, potential-free, freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation)
- 8 relay outputs, with changeover contacts, freely configurable (e.g. for faults, maintenance requests, limit alarms, external solenoid valves)
- Ethernet connection contained in the basic unit (connection on the rear side, Ethernet RJ 45, 100 MBit)
- Service interface (front side); Ethernet RJ 45, 100 MBit.
Interface modules
- Option module 2.1:
one analog output per measured component (max. 6, 0 to 20 mA, 4 to 20 mA or assignable in accordance with NAMUR), plus 6 digital outputs
|
|