General information
|
Measuring ranges
|
3, automatic measuring range switching
|
Detection limit (2?)
|
< 1% of span
|
Smallest possible span
|
Dependent on order configuration NO: 0 ... 10 / 0 … 20 / 0 … 25 / 0 … 50 vpm NO2: 0 … 101) / 0 ... 20 / 0 … 25 / 0 … 50 vpm SO2: 0 … 101) / 0 ... 20 / 0 … 25 / 0 … 50 vpm H2S: 0 … 25 / 0 … 50 vpm
|
Largest possible span
|
Dependent on order configuration NO, NO2, SO2: 0 … 300 to 0 ... 1 000 vpm H2S: 0 … 500 to 0 ... 1 000 vpm
|
UV lamp
|
|
|
EDL, electrodeless discharge lamp
|
|
? 2 years (17 500 h)
|
Conformity
|
CE mark
|
Design, enclosure
|
Degree of protection
|
IP40
|
Weight
|
approx. 17 kg
|
Requirements of location of use
|
Installation location
|
Within closed building
|
Atmospheric pressure in the environment
|
700 …1 200 hPa
|
Relative humidity
|
10 ... 95%, non-condensing
|
Permissible contamination
|
Pollution degree 1
|
Maximum geographic altitude of location of use
|
2 500 m above sea level
|
Permissible ambient temperature
|
|
|
+5 ... +45 °C (41 ... 113 °F)
|
|
-10 ... +70 °C (14 ... 158 °F)
|
Operating position
|
Front wall, vertical , max. ± 15° angle for each spatial axis (maximum permissible inclination of the base surface during operation wtih constant operating position)
|
Permissible vibration/shock
|
|
|
0.035 mm (in the range 5 ... 59 Hz)
|
- Amplitude of the starting acceleration
|
5 m/s2 (in the range 59 … 160 Hz)
|
Electrical characteristics
|
Line voltage (optional, see nameplate)
|
93 ... 132 V AC, 186 ... 264 V AC
|
Line frequency (AC)
|
47 ... 63 Hz
|
Permissible overvoltages (transient surges in the power supply network)
|
Up to overvoltage category II in accordance with IEC 60364-4-443
|
Power consumption
|
Approx. 50 VA, max. 300 VA
|
EMC interference immunity (electromagnetic compatibility)
|
In accordance with EN 61326-1, EN 61326-2-1, EN 61000-6-2, EN 61000-6-4 and EU Directive 2004/108/EC. In the case of electromagnetic radiation in the frequency range from 750 MHz ± 20 MHz, increased measuring errors can occur for small measuring ranges
|
Electrical safety
|
In accordance with EN 61010-1
|
Internal line fuses
|
|
|
6.3 A, not replaceable
|
|
8 A
|
Gas inlet conditions
|
Permissible sample gas pressure
|
Relative to ambient/atmospheric air pressure:
-200 ... +300 hPa (-0.2 ... +0.3 bar)
|
Sample gas flow
|
20 ... 120 l/h (333 ... 2 000 ml/min)
|
Sample gas temperature
|
5 ... 55 °C
|
Measuring response
|
(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
|
Reference point drift
|
< ±1%/week of respective span
|
Zero point drift
|
|
- Standard measuring ranges
|
< ± 1%/week of respective span
|
- Small measuring ranges (? 2x smallest measuring range)
|
< ± 2%/week of respective span
|
- Measured components NO, NO2, SO2
|
< ±1%/day of respective span
|
Repeatability (reproducibility)
|
< ± 1% of respective span
|
Linearity error
|
< ± 1% of respective span
|
Electric inputs and outputs
|
Analog output
|
4, 0 ... 24 mA; floating (electrically isolated), residual ripple 0.02 mA, resolution 0.1% (20 ?A), max. load 500 ?, max. voltage ± 50 V
|
Relay outputs
|
8, with changeover contacts, max. voltage ± 50 V loading capacity: Max. 30 V AC / max. 48 V DC / max. 500 mA
|
Analog inputs
|
2, 0 ... 20 mA, reference potential GND, signal strength max. 30 mA, overcurrent protection max. ± 1 A, voltage max. ± 50 V
|
Digital inputs
|
8, switching range 14 ... 42 V (external control voltage), max. voltage ± 50 V
|
Serial interface
|
RS485, Ethernet (LAN)
|