General information | |
Measuring ranges | 3, automatic measuring range switching |
Detection limit (2?) | < 1% of span |
Smallest possible span | Dependent on order configuration NO: 0 ... 10 / 0 … 20 / 0 … 25 / 0 … 50 vpm NO2: 0 … 101) / 0 ... 20 / 0 … 25 / 0 … 50 vpm SO2: 0 … 101) / 0 ... 20 / 0 … 25 / 0 … 50 vpm H2S: 0 … 25 / 0 … 50 vpm |
Largest possible span | Dependent on order configuration NO, NO2, SO2: 0 … 300 to 0 ... 1 000 vpm H2S: 0 … 500 to 0 ... 1 000 vpm |
UV lamp | |
| EDL, electrodeless discharge lamp |
| ? 2 years (17 500 h) |
Conformity | CE mark |
Design, enclosure | |
Degree of protection | IP40 |
Weight | approx. 17 kg |
Requirements of location of use | |
Installation location | Within closed building |
Atmospheric pressure in the environment | 700 …1 200 hPa |
Relative humidity | 10 ... 95 %, non-condensing |
Permissible contamination | Pollution degree 1 |
Maximum geographic altitude of location of use | 2 500 m above sea level |
Permissible ambient temperature | |
| +5 ... +45 °C (41 ... 113 °F) |
| -10 ... +70 °C (14 ... 158 °F) |
Operating position | Front wall, vertical , max. ± 15° angle for each spatial axis (maximum permissible inclination of the base surface during operation wtih constant operating position) |
Permissible vibrations/ shocks | |
| 0.035 mm (in the range 5 ... 59 Hz) |
- Amplitude of the starting acceleration
| 5 m/s2 (in the range 59 … 160 Hz) |
Electrical characteristics | |
Line voltage (optional, see nameplate) | 93 ... 132 V AC, 186 ... 264 V AC |
Line frequency (AC) | 47 ... 63 Hz |
Permissible overvoltages (transient surges in the power supply network) | Up to overvoltage category II in accordance with IEC 60364-4-443 |
Power consumption | Approx. 50 VA, max. 300 VA |
EMC interference immunity (electromagnetic compatibility) | In accordance with EN 61326-1, EN 61326-2-1, EN 61000-6-2, EN 61000-6-4 and EU Directive 2004/108/EC. In the case of electromagnetic radiation in the frequency range from 750 MHz ± 20 MHz, increased measuring errors can occur for small measuring ranges |
Electrical safety | In accordance with EN 61010-1 |
Internal line fuses | |
| 6.3 A, not replaceable |
| 8 A |
Gas inlet conditions | |
Permissible sample gas pressure | Relative to ambient/atmospheric air pressure: -200 ... +300 hPa (-0.2 … +0.3 bar) |
Sample gas flow | 20 ... 120 l/h (333 ... 2 000 ml/min) |
Sample gas temperature | 5 ... 55 °C |
Measuring response | Based on sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature |
Reference point drift | < ±1%/week of respective span |
Zero point drift | |
- Standard measuring ranges
| < ± 1%/week of respective span |
- Small measuring ranges (? 2x smallest measuring range)
| < ± 2%/week of respective span |
- Measured components NO, NO2, SO2
| < ±1%/day of respective span |
Repeatability (reproducibility) | < ± 1% of respective span |
Linearity error | < ± 1% of respective span |
Electric inputs and outputs | |
Analog output | 4, 0 ... 24 mA ; floating (electrically isolated), residual ripple 0.02 mA, resolution 0.1 % (20 ?A), max. load 500 ?, max. voltage ± 50 V |
Relay outputs | 8, with changeover contacts, max. voltage ± 50 V loading capacity: Max. 30 V AC / max. 48 V DC / max. 500 mA |
Analog inputs | 2, 0 ... 20 mA, reference potential GND, max. signal strength 30 mA, max. overcurrent protection ±1 A, max. voltage ± 50 V |
Digital inputs | 8, switching range 14 ... 42 V (external control voltage), max. voltage ± 50 V |
Serial interface | RS485, Ethernet (LAN) |