General information
|
|
Ambient temperature
|
-30 ... 55 °C (with heating)
|
|
Explosion protection
|
ATEX Category II, 2G, T3
|
|
Supply voltage
|
230 V AC, 115 V AC or 24 V DC
|
|
Max. permissible pressure at input of high-pressure reduction
|
16 000 kPa
|
|
Max. permissible pressure at input of sample preparation system with pressure regulator
|
16 000 kPa
|
|
Max. permissible pressure at output of sample preparation system
|
10 … 60 kPa above atmospheric pressure
Notice: Sample must not contain ethine!
|
|
Sampling
|
|
Lance
|
Outer diameter 6 mm and inner diameter 2 mm or outer diameter 12 mm and inner diameter 8 mm Length 1 000 mm
|
|
Special probe with BTU diaphragm and pressure reduction, non-retractable
|
Immersion depth approx. 380 mm
Protection pipe outer diameter 25.4 mm
|
|
Process connection
|
At flange DN 65 PN 16 Form C, max. 1 600 kPa gas pressure
At flange DN 65 PN 160 Form E, max. 16 000 kPa gas pressure
At flange ANSI, 2-inch, 300 lbs RF
At flange ANSI, 2-inch, 2 500 lbs RF
|
|
High-pressure reduction in casing with pressure regulators, optional heating
|
Casing with dimensions 385 x 485 x 380 mm
Primary pressure 16 000 kPa, output pressure 100/170 kPa (power consumption 150 W)
|
|
Pipe base
|
2-inch pipe base for free-standing transmitter box, 1 700 mm high
|
|
Heated line
|
Heating power is 38 W per meter Self-regulating heating up to approx. 80 °C Outer material is PE corrugated hose with 43 mm outer diameter
|
|
Sample preparation for 1 to 3 sample gas flow, plus calibration gas flow
|
|
Basic configuration
|
0.5 µm filter, 3/2-way solenoid valve, flowmeter, overflow valve and stopcock
|
|
Pressure regulator, optional heating
|
Primary pressure 16 000 kPa, output pressure 100/170 kPa (power consumption 150 W)
|
|
Simple sample injection
|
One 3/2-way solenoid valve per sample gas stream
|
|
DB&B sample injection
|
Two 3/2-way solenoid valves per sample gas flow
|
|
On mounting plate
|
652 x 422 x 3 mm
|
|
In protective casing
|
750 x 520 x 430 mm
|
|
Protective casing heating
|
Power consumption 300 W
Adjustable between 10 and 40 °C, in increments of 5
|
|
Pipe base
|
2-inch pipe base, 1 700 mm high
|
|
Additional filters
|
Aerosol, glycol
|
|
System components
|
|
On mounting plate
|
650 x 422 x 3 mm
|
|
In protective casing
|
750 x 520 x 430 mm
|
|
Protective casing heating
|
Power consumption 300 W Adjustable between 10 and 40 °C, in increments of 5.
|
|
Terminal box
|
340 x 170 x 91 mm
|
|
Pipe base
|
2-inch pipe base, 1700 mm high
|
|
Heated line
|
Heating power is 38 W per meter
|
| |
Self-regulating heating up to approximately 80 °C
|
| |
Outer material is PE corrugated hose with 43 mm outer diameter
|
|
Gas supply
|
|
Transfer station
|
Max. cylinder pressure 20 000 kPa Output pressure 50 ... 1 000 kPa
|
|
Pipe base
|
2-inch pipe base, mounted Height 2 200 mm
|
|
Gas cylinder cabinet
|
Gas cylinder cabinet for two 50 l carrier gas cylinders and one 10 l calibration gas cylinder. Dimensions are 1 250 x 400 x 2 050 mm
|
|
Cylinder heating
|
The cylinder heating system is dimensioned for 10 l calibration gas cylinders and ensures that no condensation occurs within the gas cylinder.
|
|
Cylinder pressure reducer
|
Cylinder pressure reducer for reducing primary pressure of max. 30 000 kPa to back pressure 0 … 400 kPa
|
|
Heated line
|
Heating power is 38 W per meter Self-regulated heating up to approximately 80 °C Outer material is PE corrugated hose with 43 mm outer diameter
|
|
Calibration gas
|
|
Gas mixture 1 ... 6
|
The gas mixture is stable for 36 months.
|