Climatic conditions
|
|
Ambient temperature
|
5 ... 50 °C
|
|
Relative humidity
|
70 %, non-condensing
|
|
Corrosive atmosphere
|
No
|
|
Gas inlet conditions
|
|
|
|
|
|
800 ... 1 100 hPa (absolute)
|
|
|
30 ... 90 l/h (0.5 … 1.5 l/min)
|
|
|
|
|
|
55 000 hPa (absolute)
|
|
|
30 … 90 l/h (0.5 … 1.5 l/min)
|
|
Power supply
|
|
|
200 ... 240 V AC, 47 ... 63 Hz
|
|
|
100 ... 120 V AC, 47 ... 63 Hz
|
|
|
24 V DC for switch amplifiers
|
|
Type of connections
|
|
Pipe material
|
Stainless steel
|
|
Connections/components
|
- Metric (6 mm)
- Imperial (?")
|
|
Cabling
|
|
Electrical design
|
According to IEC
|
|
Type of cables
|
Non-armored cables
|
|
Cable ID
|
No single core labeling
|
|
Installation
|
|
Site
|
Interior
|
|
Ex-zone analyzer
|
ATEX II, 3G
|
|
System design
|
|
Version
|
Cabinet
|
|
Degree of protection
|
IP54
|
|
Automatic calibration
|
No
|
|
Signal outputs
|
4 ... 20 mA/isolated contact Max. 24 V AC/DC 1 A
|
|
With sample gas return flow
|
On request
|
|
Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
|
|
Output signal fluctuation
|
< ± 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (? = 0.25%)
|
|
Zero point drift
|
< 1%/week of the smallest possible span according to rating plate
|
|
Measured-value drift
|
< 0.5%/of the smallest possible span according to rating plate
|
|
Repeatability
|
< 1% of the current measuring range
|
|
Detection limit
|
1 % of the current measuring range
|
|
Linearity error
|
< ± 1 % of the current measuring range
|
|
Influencing variable (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)
|
|
Ambient temperature
|
< 1%/10 K referred to smallest possible span according to label
|
|
Accompanying gases
|
Deviation from zero point
|
|
Sample gas flow
|
< 0.1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/h within the permissible flow range
|
|
Sample gas pressure
|
< 1 % of the current measuring range with a pressure change of 100 hPa
|
|
Auxiliary power
|
< 0.1 % of the current measuring range with rated voltage ± 10 %
|