Climatic conditions |
Ambient temperature | 5 ... 50 °C |
Relative humidity | 70 %, non-condensing |
Corrosive atmosphere | No |
Gas inlet conditions |
| |
| 800 ... 1 100 hPa (absolute) |
| 30 ... 90 l/h (0.5 … 1.5 l/min) |
| |
| 55 000 hPa (absolute) |
| 30 … 90 l/h (0.5 … 1.5 l/min) |
Power supply |
| 200 ... 240 V AC, 47 ... 63 Hz |
| 100 ... 120 V AC, 47 ... 63 Hz |
| 24 V DC for switch amplifiers |
Type of connections |
Pipe material | Stainless steel |
Connections/components | - Metric (6 mm)
- Imperial (?")
|
Cabling |
Electrical design | According to IEC |
Type of cables | Non-armored cables |
Cable ID | No single core labeling |
Installation |
Site | Interior |
Ex-zone analyzer | ATEX II, 3G |
System design |
Version | Cabinet |
Degree of protection | IP54 |
Automatic calibration | No |
Signal outputs | 4 ... 20 mA/isolated contact Max. 24 V AC/DC 1 A |
With sample gas return flow | On request |
Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
Output signal fluctuation | < ± 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (? = 0.25%) |
Zero point drift | < 1%/week of the smallest possible span according to rating plate |
Measured-value drift | < 0.5%/of the smallest possible span according to rating plate |
Repeatability | < 1% of the current measuring range |
Detection limit | 1 % of the current measuring range |
Linearity error | < ± 1 % of the current measuring range |
Influencing variable (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
Ambient temperature | < 1%/10 K referred to smallest possible span according to label |
Accompanying gases | Deviation from zero point |
Sample gas flow | < 0.1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/h within the permissible flow range |
Sample gas pressure | < 1 % of the current measuring range with a pressure change of 100 hPa |
Auxiliary power | < 0.1 % of the current measuring range with rated voltage ± 10 % |