MAXUM edition II classic oven
|
General
|
Smallest measuring ranges (depending on application)
|
- Thermal conductivity: 0 ... 500 ppm
- Flame ionization: 0 ... 1 ppm
|
Temperature range in oven
|
Application-specific, depending on temperature class 5 ... 350 °C depending on oven version and temperature class
|
Temperature control
|
± 0.02 °C
|
EMI/RFI design
|
- CE-compatible; certified according to 2004/108/EC (EMC directive)
- CE-compatible; certified according to 2006/95/EC (low-voltage directive)
- Tested according to EN 61010-1 / IEC 1010-1
|
Calibration
|
Comparison measurement with external standard
|
|
Manual or automatic
|
|
Automatic baseline correction
|
|
Standard sample cylinder (single or multipoint calibration possible)
|
Design, enclosure
|
Mounting
|
- Spacing on left: 460 mm from walls and other devices
- Spacing on right: 460 mm in all cases
- Spacing at front: 654 mm in all cases
- Wall-mounted units
- Center-to-center: 1 120 mm in all cases
|
Weight
|
77 kg
|
Degree of protection
|
IP54, Category 2
|
Danger class
|
Standard configurations:
- Certified according to ATEX with air or nitrogen purging for Zones 1 and 2 (EEx pyedmib IIB + H2)
- Suitable for use in non-hazardous areas and with non-dangerous conditions
- Certified according to CSA C/US for use in Class 1, Div. 1, Groups B, C, D with air or nitrogen purging
- Certified according to CSA C/US for use in Class 1, Div. 2, Groups B, C, D.
Important note!
Use in non-hazardous areas requires purging of the electronics area with air or nitrogen. PDD is not certified for hazardous areas.
|
Configuration
|
Oven options
|
- Single isothermal oven or divided oven with two independent isothermal zones
- Single oven or two independent, airless ovens. The dual version has two separate oven areas with separate doors which operate completely independently.
|
Detector modules for
|
thermal conductivity, flame ionization, flame photometry, helium ionization, photo-ionization and electron capture
|
Number of detector modules
|
- 1, 2 or 3 in any combination of detector module types for airbath ovens
- 1, or 2 in any combination of detector module types for airless ovens, up to 3 in special configurations
|
Sampling and column valves
|
Diaphragm valves, diaphragm piston valves, sliding vane rotary valves, slider valves, or liquid injection valve
|
Valveless option
|
Live switching
|
Columns
|
Packed, micropacked or capillary columns
|
Regulation of gas supply
|
Up to 8 electronic pressure regulator channels and up to 6 mechanical pressure regulators
|
Electrical characteristics
|
Power supply
|
- Single-phase AC, 100 ... 130 V or 195 ... 260 V (selectable), 47 … 63 Hz
- Single oven: max. 14 A
- Dual oven: 2 circuits, max. 14 A each
|
Gas inlet conditions
|
Sample flow
|
5 ... 100 ml/min (depending on application)
|
Sample filter size
|
0.1 ... 5 µm with gaseous samples depending on type of valve Max. 0.3 µm with liquid samples
|
Minimum sample pressure
|
35 kPa, standard
|
Maximum sample pressure
|
200 kPa standard, higher pressure as option
|
Maximum sample temperature
|
121 °C standard; higher temperature as option
|
Materials wetted by sample
|
Stainless steel and Teflon; other materials as option
|
Liquid injection (valve)
|
Vaporization temperature
|
60 ... 350 °C depending on application and temperature class
|
Injection volume
|
0.1 ... 9.5 ?l
|
Sample temperature
|
-20 ... +150 °C
|
Material of parts wetted by sample
|
Stainless steel, mat. no. 1.4571, Hastelloy, Monel or special materials
|
Control pressure
|
400 ... 600 kPa
|
Sample pressure
|
Max. 6 000 kPa, recommended 50 ... 100 kPa
|
Connections for pipe
|
3.14 mm (1/8") outer diameter
|
Measuring response
|
Sensitivity (depending on application)
|
± 0.5% of span
|
Linearity (depending on application)
|
± 2% of span
|
Effects of vibrations
|
Negligible
|
Repeatability in % of full span between
|
2 and 100%: ± 0.5 %;
0.05 and 2%: ± 1 %;
50 and 500 ppm: ± 2 %;
5 and 50 ppm: ± 3 %;
0.5 and 5 ppm: ± 5 %
|
Detection limits
|
See detectors
|
Influencing variables
|
Effects of ambient temperature
|
None with electronic pressure control
Different effects with mechanical pressure control (depending on application)
|
Electrical inputs and outputs
|
Standard input and output
|
- 2 analog outputs;
- 4 digital outputs (1 for output of system faults, 3 are user-configurable);
- 4 digital inputs;
- 3 serial outputs
|
Card slots for optional inputs and outputs via internal I2C bus
|
2
|
Input and output cards
|
A IO 8: 8 analog outputs, 8 analog inputs, 2 digital inputs
D IO: 6 digital inputs and 8 digital outputs
AD I/O: 4 digital inputs and 4 digital outputs, 4 analog inputs and 4 analog outputs
|
Digital inputs
|
Optocoupler with internal power supply (12 … 24 V DC); switchable by dry contacts.
Alternative: switchable by external power supply 12 … 24 V DC (only dry relay contacts), external power supply, negative connection linked to ground, for a specific digital input.
|
Digital outputs
|
Dry changeover contacts, max. contact rating:
1 A with 30 V DC.
Diode bypass suppression should be used for inductive loads.
|
Analog inputs
|
-20 ... +20 mA into 50 ? or
-10 ... +10 V Rin = 0.1 M?, alternate insulation up to 100 V
|
Analog outputs
|
0/4 ... 20 mA into max. 750 ?, common negative pole, electrically isolated from ground; freely-connectable to ground
|
Termination
|
Screw terminal for shielded or solid cable with a maximum area of 18 AWG or 0.82 mm2
|
Climatic conditions
|
Ambient temperature
|
-18 ... 50 °C application-dependent
|
Gas supply
|
Instrument air
|
- At least 350 kPa for units with valves Model 11 or Valco
- At least 825 kPa for units with valves Model 50
- At least 175 kPa for airbath ovens; 85 l/min per oven
- No instrument air for fan-free ovens
|
Carrier gas
|
- Nitrogen or helium in compressed gas cylinder, purity 99.998%, or hydrogen with a purity of 99.999% (depending on application).
- Typical consumption quantity: 5 100 l/month per detector module
|
Combustion gas
|
- Hydrogen with a purity of 99.999%
- Typical consumption quantity: 2 000 l/month per detector module
|
Combustion air
|
- Reference air (< 1 ppm THC, O2 concentration 20 … 21%). Supply through instrument air with catalytic purification (optional).
- Typical consumption quantity: 26 000 l/month
|
Corrosion protection
|
- Purging with dry air to protect the electronics
- Air bath oven with stainless steel lining
- Airless oven made of aluminum
- Steel lining painted on outside (epoxy powder coating)
|
Communication
|
Serial output
|
RS 232, RS 485, e.g. Modbus
|
Ethernet
|
Standard 10/100 BaseT Ethernet with 4 RJ 45 connectors e.g. Modbus TCP IP or OPC
Optional
ESBF board
Fiber-optic 100Base FX multimode with ST connection
3 x RJ45 and 1 x optical or
Scalance network components e.g. for redundant connections.
|
|
|