General information
19" rack unit
  • 19" rack unit with 4 HU for installation
    • in hinged frame
    • in cabinets with or without telescopic rails
  • Front plate can be swung down for servicing purposes (laptop connection)
  • Internal gas paths: hose made of FKM (Viton) or pipe made of titanium or stainless steel
  • Gas connections for sample gas inlet and outlet: pipe diameter 6 mm or 1/4"
  • Flow indicator for sample gas on front plate (option)
  • Sample chamber (OXYMAT channel) – with or without flow-type compensation branch – made of stainless steel (mat. no. 1.4571) or of tantalum for highly corrosive sample gases (e.g. HCl, Cl2, SO2, SO3, etc.)
  • Monitoring (option) of sample gas and/or reference gas (both channels)
Display and control panel
  • Large LCD panel for simultaneous display of:
    • Measured value (digital and analog displays)
    • Status bar
    • Measuring ranges
  • Contrast of LCD panel adjustable using menu
  • Permanent LED backlighting
  • Washable membrane keyboard with five softkeys
  • Menu-driven operation for parameterization, test functions, adjustment
  • User help in plain text
  • Graphic display of concentration trend; programmable time intervals
  • Bilingual operating software:
    German/English, English/Spanish, French/English, Italian/English, Spanish/English
Inputs and outputs (per channel)
  • One analog output for each measured component
  • Two analog inputs freely configurable (e.g. correction of cross-interference or external pressure sensor)
  • Six binary inputs freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation)
  • Six relay outputs freely configurable e.g. for fault, maintenance request, limit alarm, external solenoid valves)
  • Expansion by eight additional binary inputs and eight additional relay outputs e.g. for autocalibration with up to four calibration gases
Communication

RS 485 present in the basic unit (connection at the rear; for the rack unit also behind the front plate).

Options

  • AK interface for the automotive industry with extended functions
  • RS 485/RS 232 converter
  • RS 485/Ethernet converter
  • RS 485/USB converter
  • Connection to networks via PROFIBUS DP/PA interface
  • SIPROM GA software as the service and maintenance tool

ULTRAMAT/OXYMAT 6, membrane keyboard and graphic display

Designs – Parts touched by sample gas, standard

Gas path ULTRAMAT channel

19" rack unit

With hoses

Bushing

Stainless steel, mat. no. 1.4571

Hose

FKM (e.g. Viton)

Sample chamber:

  • Body

Aluminum

  • Lining

Aluminum

  • Fitting

Stainless steel, mat. no. 1.4571,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

  • Window

CaF2, adhesive: E353,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

With pipes

Bushing

Titanium

Pipe

Titanium,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

Sample chamber:

  • Body

Aluminum

  • Lining

Tantalum (only for cell length 20 mm to 180 mm)

  • Window

CaF2, adhesive: E353,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

With pipes

Bushing

Stainless steel, mat. no. 1.4571

Pipe

Stainless steel, mat. no. 1.4571,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

Sample chamber:

  • Body

Aluminum

  • Lining

Aluminum or tantalum (Ta: only for cell length 20 mm to 180 mm)

  • Window

CaF2, adhesive: E353,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

Flow indicator

Measurement pipe

Duran glass

Variable area

Duran glass

Suspension boundary

PTFE (Teflon)

Angle pieces

FKM (e.g. Viton)

Pressure switch

Membrane

FKM (e.g. Viton)

Enclosure

PA 6.3T


Options

Gas path ULTRAMAT channel

19" rack unit

Flow indicator

Measurement pipe

Duran glass

Variable area

Duran glass

Suspension boundary

PTFE (Teflon)

Angle pieces

FKM (e.g. Viton)

Pressure switch

Membrane

FKM (e.g. Viton)

Enclosure

PA 6.3T


Versions – Parts wetted by sample gas, special applications (examples)

Gas path ULTRAMAT channel

19" rack unit

With pipes

Bushing

e.g. Hastelloy C22

Pipe

e.g. Hastelloy C22,

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)

Sample chamber:

  • Body

e.g. Hastelloy C22

  • Window

CaF2, without adhesive

O-ring: FKM (e.g. Viton) or FFKM (Kalrez)


Designs – Parts touched by sample gas, standard

Gas path OXYMAT channel

19" rack unit

With hoses

Bushing

Hose

Sample chamber

Fittings for sample chamber

Restrictor

O-rings

Stainless steel, mat. no. 1.4571

FKM (e.g. Viton)

Stainless steel, mat. no. 1.4571 or tantalum

Stainless steel, mat. no. 1.4571

PTFE (e.g. Teflon)

FKM (e.g. Viton)

With pipes

Bushing

Pipe

Sample chamber

Restrictor

O-rings

Titanium

Titanium

Stainless steel, mat. no. 1.4571 or Tantalum

Titanium

FKM (Viton) or FFKM (Kalrez)

With pipes

Bushing

Pipe

Sample chamber

Restrictor

O-rings

Stainless steel, mat. no. 1.4571

Stainless steel, mat. no. 1.4571

Stainless steel, mat. no. 1.4571 or Tantalum

Stainless steel, mat. no. 1.4571

FKM (Viton) or FFKM (Kalrez)

With pipes

Bushing

Pipe

Sample chamber

Restrictor

O-rings

Hastelloy C 22

Hastelloy C 22

Stainless steel, mat. no. 1.4571 or Tantalum

Hastelloy C 22

FKM (e.g. Viton) or FFKM (e.g. Kalrez)


Options

Gas path ULTRAMAT channel
and OXYMAT channel

19" rack unit

Flow indicator

Measurement pipe

Duran glass

Variable area

Duran glass

Suspension boundary

PTFE (Teflon)

Angle pieces

FKM (e.g. Viton)

Pressure switch

Membrane

FKM (e.g. Viton)

Enclosure

PA 6.3T


Gas path

Legend for the gas path figures

1

Sample gas inlet (OXYMAT channel)

10

Connection of pressure sensor (ULTRAMAT channel)

2

Sample gas outlet (OXYMAT channel)

11

Restrictor (in reference gas inlet)

3

Not used

12

O2 physical system

4

Reference gas inlet

13

Pressure sensor

5

Sample gas inlet (ULTRAMAT channel)

14

Pressure switch in sample gas path (option)

6

Sample gas outlet (ULTRAMAT channel)

15

Flow indicator in sample gas path (option)

7

Reference gas outlet (ULTRAMAT channel, option)

16

IR physical system

8

Reference gas inlet (ULTRAMAT channel, option)

17

Filter

9

Purging gas

18

Pressure switch (reference gas) (option)


ULTRAMAT/OXYMAT 6, gas path (example) IR channel without flow-type reference side

ULTRAMAT/OXYMAT 6, gas path (example) IR channel with flow-type reference side























































Каталог оборудования 2012
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 ULTRAMAT 6 ULTRAMAT/OXYMAT 6 General information 19" rack unit Documentation Suggestions for spare parts OXYMAT 6 OXYMAT 61 OXYMAT 64 CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо

  © ООО "СК ЭНЕРГО" 2007-2022
  (4872) 700-366
Яндекс.Метрика