19" rack unit

ULTRAMAT/OXYMAT 6, 19" rack unit

General information

Operating position

Front wall, vertical

Conformity

CE mark in accordance with EN 50081-1, EN 50082-2

Design, enclosure

Weight

Approx. 21 kg

Degree of protection

IP20 according to EN 60529

Electrical characteristics

EMC (ElectromagneticCompatibility)

In accordance with standard requirements of NAMUR NE21 (08/98)

Electrical safety

According to EN 61010-1, overvoltage category III

Power supply

100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or
200 ... 240 V AC (nominal range of use 180 ... 264 V), 48 ... 63 Hz

Power consumption

Approx. 70 VA

Fuse values

120 ... 120 V: F1/F2 = T 1.6 A
200 ... 240 V: F1/F2 = T 1 A

Electrical inputs and outputs (per channel)

Analog output

0/2/4 ... 20 mA, isolated; max. load 750 

Relay outputs

6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated, non-sparking

Analog inputs

2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of influence of accompanying gas (correction of cross-interference)

Binary inputs

6, designed for 24 V, isolated, freely parameterizable, e.g. for measuring range switchover

Serial interface

RS 485

Options

AUTOCAL function each with 8 additional binary inputs and relay outputs, also with PROFIBUS PA or PROFIBUS DP

Climatic conditions

Permissible ambient temperature

-30 ... +70 °C during storage and transportation, 5 … 45 °C during operation

Permissible humidity

< 90% relative humidity, during storage and transportation (dew point must not be undershot)


Technical data, ULTRAMAT channel

Measuring ranges

4, internally and externally switchable; autoranging is also possible

Smallest possible measuring range

Dependent on the application, e.g.
CO: 0 ... 10 vpm
CO2: 0 ... 5 vpm

Largest possible measuring range

Dependent on the application

Measuring ranges with suppressed zero point

Any zero point within 0 ... 100 vol.% can be implemented; smallest possible span 20 %

Characteristic

Linearized

Influence of interfering gases must be considered separately

Gas inlet conditions

Permissible sample gas pressure

  • Without pressure switch

700 ... 1 500 hPa (absolute)

  • With integrated pressure switch

700 ... 1 300 hPa (absolute)

Sample gas flow

18 ... 90 l/h (0.3 ... 1.5 l/min)

Sample gas temperature

Min. 0 to max. 50 °C, but above the dew point

Sample gas humidity

< 90% (relative humidity), or dependent on measuring task, non-condensing

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90-time)

Dependent on length of analyzer chamber, sample gas line and parameterizable damping

Damping (electrical time constant)

0 ... 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approx. 0.5 ... 5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

Pressure sensor

  • Internal

700 ... 1 200 hPa absolute

  • External

700 ... 1 500 hPa absolute

Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< ± 1% of the smallest possible measuring range according to rating plate

Zero point drift

<± 1% of the current measuring range/week

Measured-value drift

<± 1% of the current measuring range/week

Repeatability

1% of the current measuring range

Detection limit

1 % of the smallest possible measuring range

Linearity error

< 0.5% of the full-scale value

Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

< 1% of current measuring range/10 K (with constant receiver cell temperature)

Sample gas pressure

  • When pressure compensation has been switched on: < 0.15% of the span/1% change in atmospheric pressure
  • When pressure compensation has been switched off: < 1.5% of the span/1% change in atmospheric pressure

Sample gas flow

Negligible

Power supply

< 0.1% of the current measuring range with rated voltage ± 10%

Environmental conditions

Application-specific measuring influences possible if ambient air contains measured component or cross interference-sensitive gases


Technical data, OXYMAT channel

Measuring ranges

4, internally and externally switchable; automatic measuring range switchover also possible

Smallest possible span (relating to sample gas pressure 1 000 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

0.5 vol.%, 2 vol.% or 5 vol.% O2

Largest possible measuring range

100 vol. % O2

Measuring ranges with suppressed zero point

Any zero point within 0 ... 100 vol.% can be implemented, provided that a suitable reference gas is used

Gas inlet conditions

Permissible sample gas pressure

  • With pipes

500 ... 3 000 hPa absolute

  • With hoses

  • Without pressure switch

500 ... 1 500 hPa absolute

  • With pressure switch

500 ... 1 300 hPa absolute

Sample gas flow

18 ... 60 l/h (0.3 ... 1 l/min)

Sample gas temperature

0 ... 50 C

Sample gas humidity

< 90% RH (relative humidity)

Reference gas pressure (high-pressure version)

2 000 ... 4 000 hPa above sample gas pressure, but max. 5 000 hPa

Reference gas pressure (low-pressure version)

Min. 100 hPa above sample gas pressure

Dynamic response

Warm-up period

At room temperature < 30 min (the technical specification will be met after 2 hours)

Delayed display (T90 time)

Min. 1.5 ... 3.5 s, depending on version

Damping (electrical time constant)

0 ... 100 s, parameterizable

Dead time (purging time of the gas path in the unit at 1 l/min)

Approx. 0.5 ... 2.5 s, depending on version

Time for device-internal signal processing

< 1 s

Pressure correction range

Pressure sensor

  • Internal

500 ... 2 000 hPa absolute

  • External

500 ... 3 000 hPa absolute

Measuring response(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Output signal fluctuation

< 0.75% of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25% at 2 )

Zero point drift

< 0.5%/month of the smallest possible measuring span according to rating plate

Measured-value drift

0.5%/month of the current measuring range

Repeatability

1%/month of the current measuring range

Detection limit

1% of the current measuring range

Linearity error

1% of the current measuring range

Influencing variables(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature)

Ambient temperature

  • < 0.5%/10 K referred to smallest possible span according to rating plate
  • With measuring span 0.5%: 1%/10 K

Sample gas pressure (with air (100 hPa) as reference gas, correction of the atmospheric pressure fluctuations is only possible if the sample gas can vent to ambient air)

  • When pressure compensation has been switched off: < 2% of the current measuring range/1% atmospheric pressure change
  • When pressure compensation has been switched on: < 0.2% of the current measuring range/1% atmospheric pressure change

Accompanying gases

Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas

Sample gas flow

< 1% of the smallest possible span according to rating plate with a change in flow of 0.1 l/min within the permissible flow range

Power supply

< 0.1 % of the current measuring range with rated voltage ± 10%
























































Каталог оборудования 2012
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Введение ULTRAMAT 23 ULTRAMAT 6 ULTRAMAT/OXYMAT 6 General information 19" rack unit Documentation Suggestions for spare parts OXYMAT 6 OXYMAT 61 OXYMAT 64 CALOMAT 6 CALOMAT 62 FIDAMAT 6 Непрерывный газовый анализ, in-situ Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо