Measured variable
|
Total mercury
|
|
Measuring ranges
|
0 ... 45, 0 ... 75 to 0 ... 500 ?g/Nm3
|
|
Measuring principle
|
UV absorption
|
|
Sample gas temperature
|
0 ... 250 °C
|
|
Sample gas pressure
|
-50 ... +50 hPa
|
|
Duct diameter
|
> 0.5 m
|
|
Ambient temperature
|
5 °C ... 40 °C
|
|
Degree of protection
|
IP40 (IP54)
|
|
Measured value outputs
|
2 x 0/4 ... 20 mA, 500 ?
|
|
Relative expanded uncertainty
|
5.5 % CR
|
|
Digital outputs
|
4 relay outputs, permissible load 250 V, 100 VA
|
|
Digital inputs
|
1 isolated input
|
|
Detection limit
|
< 0.1 ?g/Nm3
|
|
Reference point drift
|
< 2 % ZRE
|
|
Zero point drift
|
< 0.5 % ZRE automatic zero offset
|
|
Supply voltage
|
230/400 VAC, 50 Hz, 3 x L, N, PE
|
|
|
1 200 VA
|
|
|
650 VA
|
|
|
100 VA/m
|
- Heated sample pipe 0.6 m/ 1.0 m/ 1.5 m
|
600 VA/800 VA/1 200 VA
|
|
Dimensions (H x W x D) control cabinet
|
1 700 x 800 x 500 mm
|
|
Weight
|
220 kg
|
|
Purge air supply, compressed air
|
6 ... 8 bar (for reference gas generator only)
|