Measuring instrument D-R 290
|
|
|
Measurements
|
|
|
Measuring ranges
|
|
|
|
0 ... 20 to 0 ... 100 %
|
|
|
0 ... 0.1 to 0 ... 1.6
|
|
|
0 ... 80 mg/m3 to 0 ... 4 000 mg/m3
|
|
Measuring principle
|
Transmission
|
|
Sample gas temperature
|
Above dew point up to 250 °C, optional up to 1 000 °C, depending on application
|
|
Sample gas pressure
|
-50 ... +20 hPa, optional higher
|
|
Duct diameter
|
1 ... 12 m, optional up to 18 m
|
|
Ambient temperature
|
-20 ... +50 °C, optional higher
|
|
Degree of protection
|
IP65, Ex optional
|
|
Measured value outputs
|
2 x 0/4 ... 20 mA, 500 ? Manual or automatic measurement range switching optional Modbus RTU, PROFIBUS DP
|
|
Digital outputs
|
6 programmable relay outputs, permissible load 48 V/0.5 A
|
|
Digital inputs
|
6 programmable isolated inputs
|
|
Accuracy
|
< 1 % of measuring range
|
|
Detection limit
|
0.75 % at extinction 0–0.1
|
|
Reference point drift
|
< 0.4 % of measuring range / month
|
|
Zero point drift
|
< 0.4 % of measuring range / month
|
|
Auxiliary power
|
95 ... 264 VAC, 47 ... 63 Hz, 30 VA
|
|
Dimensions (H x W x D) measuring head
|
363 x 185 x 398 mm
|
|
Weight
|
17 kg
|
|
Purge air supply
|
|
|
Purge air quantity
|
Approx. 80 m3/h
|
|
Auxiliary power
|
115/230 VAC, 50/60 Hz, 0.37/0.43 kW
|
|
Dimensions (H x W x D)
|
350 x 550 x 500 mm
|
|
Weight
|
12 kg
|
|
Degree of protection
|
IP55
|