- 19“ unit with 4 HU for installation
- in swing frames
- in cabinets, with or without slide rails
- Front panel for service can be hinged down
(laptop connection)
- Gas connections for sample gas input and output: pipe diameter 6 mm or 1/4"
- Gas and electrical connections at the rear.
Display and control panel
- Large LCD panel for simultaneous display of:
- Measured value
- Status line
- Measuring ranges
- Contrast of LCD panel adjustable via menu
- Permanent LED backlighting
- Washable membrane keyboard with five softkeys
- Menu-based operation for configuration, test functions and calibration
- User help in plain text
- Graphic display of concentration trend; programmable time intervals.
Inputs and outputs
- Six binary inputs freely configurable (e.g. for range switching, processing of external signals from sample preparation)
- Six relay outputs freely configurable (failure, maintenance request, maintenance switch, limit alarm, external solenoid valves)
- Two programmable analog inputs (e.g. cross interference correction, external pressure sensor)
- Extension with eight additional binary inputs and eight additional relay outputs for automatic calibration with up to four calibration gases.
Communication
- RS 485 included in basic unit
(connection at the rear).
Options
- RS 485/RS 232 converter
- RS 485/Ethernet converter
- Linking to networks via PROFIBUS DP/PA interface
- SIPROM GA software as service and maintenance tool.
OXYMAT 61, membrane keyboard and graphic display
Executions of the wetted parts
|
Executions of wetted parts
|
|
|
Gas path
|
19“ unit
|
|
with hoses
|
Nipple
Hose
Sample cell
Stub sample cell
Restrictor
O-rings
Hose coupling
|
SS, type No. 1.4571
FKM (Viton)
SS, type No. 1.4571
SS, type No. 1.4571
PTFE (Teflon)
FKM (Viton)
Polyamide 6
|
Options
|
Options
|
|
|
|
Flowmeter
|
Metering pipe
Float
Float limit
Elbows
|
Duran glass
Duransglass, black
PTFE (Teflon)
FKM (Viton)
|
|
Sample gas pressure switch
|
Membrane
Enclosure
|
FKM (Viton)
PA 6.3T
|
Gas path
|
Keys to the gas path figures
|
|
1
|
Sample gas inlet
|
8
|
Pressure switch in sample gas path (option)
|
|
2
|
Sample gas outlet
|
9
|
Purging gas
|
|
3
|
Not used
|
10
|
Restrictor im in reference gas path (outlet)
|
|
4
|
Reference gas inlet
|
11
|
Pressure switch for reference gas monitoring
|
|
5
|
Restrictor in reference gas path
|
12
|
Pump
|
|
6
|
O2 physics
|
13
|
Filter
|
|
7
|
Restrictor in sample gas path
|
14
|
Flowmeter in sample gas path (option)
|
Gas path OXYMAT 61 with integrated reference gas pump (connection for 100 hPa)
Gas path OXYMAT 61 with reference gas connection 3000 to 4000 hPa
|
|