Measuring ranges
|
4, switchable internally and externally; autoranging is also possible
|
Smallest possible measuring range
|
Depending on application, e.g. CO: 0 to 10 vpm, CO2: 0 to 5 vpm
|
Largest possible measuring range
|
Depending on application
|
Measuring range with suppressed zero
|
Any zero point within 0 to 100 % can be achieved; smallest possible span 20 %
|
Heated version
|
65 °C
|
Characteristic
|
Linearized
|
Position of use
|
Front panel vertical
|
Conformity
|
CE identification EN 50081-1, EN 50082-2
|
Design, enclosure
|
|
Weight
|
Approx. 32 kg
|
Degree of protection
|
IP65 according to EN 60529, restricted breathing to EN 50021
|
Electrical characteristics
|
|
Power supply
|
100 to 120 V AC (rated range 90 to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 to 264 V), 48 to 63 Hz
|
Power consumption
|
Approx. 35 VA; approx. 330 VA with heated version
|
Electromagnetic compatibility (EMC)
|
According to standard requirements of NAMUR NE21 (08/98)
|
Electrical safety
|
According to EN 61010-1
|
|
overvoltage category II
|
|
overvoltage category III
|
Fuses (unit without heater)
|
|
|
F3: 1T/250; F4: 1T/250
|
|
F3: 0.63T/250; F4: 0.63T/250
|
Fuses (unit with heater)
|
|
|
F1: 1T/250; F2: 4T/250
F3: 4T/250; F4: 4T/250
|
|
F1: 0.63T/250; F2: 2.5T/250
F3: 2.5T/250; F4: 2.5T/250
|
Gas inlet conditions
|
|
Perm. sample gas pressure
|
|
- for analyzers with hoses (without pressure switch)
|
600 to 1500 hPa (absolute)
|
- for analyzers with pipes (without pressure switch)
|
600 to 1500 hPa (absolute)
|
- Ex (leakage compensation)
|
600 to 1160 hPa (absolute)
|
|
600 to 1500 hPa (absolute)
|
Purging gas pressure
|
|
|
< 165 hPa above ambient
|
|
250 hPa above ambient
|
Sample gas flow
|
18 to 90 l/h (0.3 to 1.5 l/min)
|
Sample gas temperature
|
0 to 50 °C; with heated version: 0 to 80 °C
|
Sample gas humidity
|
< 90% RH (relative humidity) or depending on application
|
Time response
|
|
Warm-up period
|
With amb. temperature < 30 min (maximum accuracy achieved after 2 hours); heated version: approx. 90 min
|
Response time (T90 time)
|
Dependent on length of analyzer cell, sample gas line and programmed damping
|
Damping (electric time constant)
|
0 to 100 s, programmable
|
Dead time (purging time of gas path in analyzer at 1 l/min)
|
Approx. 0.5 to 5 s, depending on version
|
Time for internal signal processing
|
< 1 s
|
Pressure correction range
|
|
Pressure sensor
|
|
|
600 to 1200 hPa absolute
|
|
600 to 1500 hPa absolute
|
Measuring response (maximum accuracy achieved after 2 hours)
|
Output signal fluctuation
|
± 0.1% to ± 1% of smallest possible measuring range specified on rating plate, depending on application with the unit specific electronic time constant (corresponds to ±0.33% at 2?)
|
Zero drift
|
< 1% of measuring range/week
|
Measured-value drift
|
< 1% of measuring range/week
|
Repeatability
|
Between 0.1% and 1% of respective measuring range
|
Minimum detectable quantity
|
1% of smallest measuring range
|
Linearity error
|
< 0.5% of full-scale value
|
Influencing variables (referred to 1000 hPa absolute sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature)
|
Ambient temperature
|
< 1% of measuring range/10 K (for a constant temperature of the reception cell)
|
Sample gas pressure
|
With pressure compensation: < 0.15% of setpoint/1% change in atmospheric pressure
|
Sample gas flow
|
Negligible
|
Power supply
|
< 0.1% of output signal span at rated voltage ± 10%
|
Ambient conditions
|
Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases
|
Electric inputs and outputs
|
|
Analog output
|
0/2/4 to 20 mA. floating; max. load 750 ?
|
Relay outputs
|
6, with changeover contacts, freely selectable, e.g. for range identification; loading capacity: 24 V AC/DC / 1 A floating, non sparking
|
Analog inputs
|
2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross-interference)
|
Binary inputs
|
6, designed for 24 V, floating, freely selectable, e.g. for range switching
|
Serial interface
|
RS 485
|
Options
|
Autocal function with 8 additional binary inputs and 8 relay outputs, also with PROFIBUS PA and PROFIBUS DP
|
Ambient conditions
|
|
Permissible ambient temperature
|
-30 to +70 °C during storage and transport, +5 to +45 °C during operation
|
Permissible humidity
|
< 90 % RH (relative humidity) as annual average, during storage and transport (dew point must not be fallen below)
|