Pressure transmitters for food, pharmaceuticals and biotechnology |
|
Mode of operation |
|
Measuring principle |
piezo-resistive |
Input |
|
Measured variable |
Gage or absolute pressure |
Measuring range |
0 ... 160 mbar (0 ... 2.32 psi) |
Output |
|
Output signal |
|
|
4 ... 20 mA |
|
0 ... 20 mA |
Measuring accuracy |
According to IEC 60770-1 |
Error in measurement at limit setting incl. hysteresis and reproducibility |
0.2 % of full-scale value |
Adjustment accuracy |
±0.2 % of full-scale value |
Step response time |
< 20 ms |
Influence of ambient temperature |
|
On the enclosure |
|
|
< 0.2 %/10 K of full-scale value |
|
< 0.2 %/10 K of full-scale value |
On the process connection (remote seals) |
Zero error (depends on design) |
|
|
|
4.8 mbar/10 K (0.069 psi/10 K) |
|
2.3 mbar/10 K (0.92 psi/10 K) |
|
1.6 mbar/10 K (0.64 psi/10 K) |
|
0.6 mbar/10 K (0.24 psi/10 K) |
|
|
|
9.5 mbar/10 K (0.14 psi/10 K) |
|
4.1 mbar/10 K (0.06 psi/10 K) |
|
3.9 mbar/10 K (0.05 psi/10 K) |
|
3.9 mbar/10 K (0.05 psi/10 K) |
The zero error specified for the process connection should be considered as a guideline for a standard design. We will produce a detailed system calculation on request. Systems with reduced remote seal errors are available on request. |
|
Rated conditions |
|
Installation conditions |
|
|
Any, vertical as standard |
Ambient conditions |
|
|
-10 ... +70 °C (14 ... 158 °F) |
|
-10 ... +90 °C (14 ... 194 °F) |
|
Max. 200 °C (392 °F), depending on design |
|
IP65, optional IP67 |
|
|
|
To EN 50081 Part 1, issue 1993 (residential and industrial areas). The unit has no own emissions. |
|
EN 50082 Part 2, issue March 1995 (industrial areas) |
Design |
|
Weight (without remote seal) |
|
|
460 G ( 1.01 lb) |
|
200 g ( 0.44 lb) |
Enclosure |
|
|
|
|
Stainless steel, mat. no. 1.4404/316L/1.4305 |
Material of union nut |
Polyamide (with electrical connection using plug or cable) Electronics unit potted with silicone Internal ventilation for measuring ranges < 16 bar (< 232 psi), through housing thread or connection cable depending on design |
Process connection |
|
|
See ordering data |
|
Stainless steel, mat. no. 1.4404/316L |
Auxiliary power supply |
|
Terminal voltage on transmitter |
10 ... 30 V DC |
Rated voltage |
24 V DC |
Certificates and approvals |
|
Classification according to pressure equipment directive (PED 97/23/EC) |
For gases of fluid group 1 and liquids of fluid group 1; complies with the requirements of article 3, paragraph 1 (appendix 1); assigned to category III, conformity evaluation module H by the TUV Nord |
Explosion protection |
|
|
TUV 03 ATEX 2099 X |
|
Ex II 2G EEx ib IIC T6 |