Gauge pressure input | |
Measured variable | Gauge pressure |
Span (fully adjustable) or nominal measuring range, max. operating pressure (in accordance with 2014/68/EU Pressure Equipment Directive) and max. test pressure (pursuant to DIN 16086) (for oxygen measurement, max. 100 bar/10 MPa/1450 psi and 60 °C (140 °F) ambient temperature/process temperature) | HART | PROFIBUS PA/ FOUNDATION Fieldbus | | |
Span | Nominal measuring range | Max. perm. operating pressure MAWP (PS) | Max. permissible test pressure |
8.3 ... 250 mbar 0.83 ... 25 kPa 0.12 ... 3.6 psi | 250 mbar 25 kPa 3.6 psi | 4 bar 400 kPa 58 psi | 6 bar 600 kPa 87 psi |
0.01 ... 1 bar 1 ... 100 kPa 0.15 ... 14.5 psi | 1 bar 100 kPa 14.5 psi | 4 bar 400 kPa 58 psi | 6 bar 600 kPa 87 psi |
0.04 ... 4 bar 4 ... 400 kPa 0.58 ... 58 psi | 4 bar 400 kPa 58 psi | 7 bar 0.7 Mpa 102 psi | 10 bar 1 MPa 145 psi |
0.16 ... 16 bar 16 ... 1600 kPa 2.3 ... 232 psi | 16 bar 1600 kPa 232 psi | 21 bar 2.1 MPa 305 psi | 32 bar 3.2 MPa 464 psi |
0.63 ... 63 bar 63 ... 6300 kPa 9.1 ... 914 psi | 63 bar 6300 kPa 914 psi | 67 bar 6.7 MPa 972 psi | 100 bar 10 MPa 1450 psi |
1.6 ... 160 bar 0.16 ... 16 MPa 23 ... 2321 psi | 160 bar 16 MPa 2321 psi | 167 bar 16.7 MPa 2422 psi | 250 bar 25 MPa 3626 psi |
4 ... 400 bar 0.4 ... 40 MPa 58 ... 5802 psi | 400 bar 40 MPa 5802 psi | 400 bar 40 MPa 5802 psi | 600 bar 60 MPa 8702 psi |
Lower measuring limit (for 250 mbar/25 kPa/3.6 psi measuring cells, the lower measuring limit is 750 mbar a/75 kPa a/10.8 psi a. The measuring cell is vacuum-resistant up to 30 mbar a/3 kPa a/0.44 psia.) | |
- Measuring cell with silicone oil filling
| 30 mbar a/3 kPa a/0.44 psia |
- Measuring cell with inert fill fluid
| 30 mbar a/3 kPa a/0.44 psia |
Upper measuring limit | 100% of the max. measuring span (for oxygen measurement max. 100 bar/10 MPa/ 1450 psi and 60 °C (140 °F) ambient temperature/process temperature) |
Absolute pressure input | |
Measured variable | Absolute pressure |
Span (fully adjustable) or nominal measuring range, max. operating pressure (in accordance with 2014/68/EU Pressure Equipment Directive) and max. test pressure (pursuant to DIN 16086) | HART | PROFIBUS PA/ FOUNDATION Fieldbus | | |
Span | Nominal measuring range | Max. perm. operating pressure MAWP (PS) | Max. permissible test pressure |
8.3 ... 250 mbar a 0.83 ... 25 kPa a 3 ... 100 inH2O a | 250 mbar a 25 kPa a 100 inH2O a | 1.5 bar a 150 kPa a 21.8 psia | 6 bar a 600 kPa a 87 psia |
43 ... 1300 mbar a 4.3 ... 130 kPa a 17 ... 525 inH2O a | 1300 mbar a 130 kPa a 525 inH2O a | 2.6 bar a 260 kPa a 37.7 psia | 10 bar a 1 MPa a 145 psia |
160 ... 5000 mbar a 16 ... 500 kPa a 2.32 ... 72.5 psia | 5000 mbar a 500 kPa a 72.5 psia | 10 bar a 1 MPa a 145 psia | 30 bar a 3 MPa a 435 psia |
1 ... 30 bar a 0.1 ... 3 MPa a 14.5 ... 435 psia | 30 bar a 3 MPa a 435 psia | 45 bar a 4.5 MPa a 653 psia | 100 bar a 10 MPa a 1450 psia |
Low measuring limit | |
- Measuring cell with silicone oil filling
| 0 mbar a/3 kPa a/0.44 psia |
- Measuring cell with inert liquid
| |
- for process temperature -20 °C < ? ? +60 °C
(-4 °F < ? ? +140 °F)
| 30 mbar a/0 kPa a/0 psia |
- for process temperature 60 °C < ? ? +100 °C (max. 85 °C for measuring cell 30 bar) (140 °F < ? ? +212 °C (max. 185 °C for measuring cell 435 psi))
| 30 mbar a + 20 mbar a . ( ? - 60 °C)/°C 3 kPa a + 2 kPa a . ( ? - 60 °C)/°C 0.44 psi a + 0.29 psi a . ( ? - 108 °F)/°F |
Upper measuring limit | 100% of the max. measuring span (for oxygen measurement max. 100 bar/10 MPa/ 1450 psi and 60 °C (108 °F) ambient temperature/process temperature) |
Start of scale | Between the measuring limits (continuously adjustable) |
Input of gauge pressure, with flush-mounted diaphragm | |
Measured variable | Gauge pressure, flush-mounted |
Span (fully adjustable) or nominal measuring range, max. permissible operating pressure and max. test pressure | HART | PROFIBUS PA/ FOUNDATION Fieldbus | | |
Span | Nominal measuring range | Max. perm. operating pressure MAWP (PS) | Max. permissible test pressure |
0.01 ... 1 bar 1 ... 100 kPa 0.15 ... 14.5 psi | 1 bar 100 kPa 14.5 psi | 4 bar 400 kPa 58 psi | 6 bar 600 kPa 87 psi |
0.04 ... 4 bar 4 ... 400 kPa 0.58 ... 58 psi | 4 bar 400 kPa 58 psi | 7 bar 0.7 Mpa 102 psi | 10 bar 1 MPa 145 psi |
0.16 ... 16 bar 16 ... 1600 kPa 2.3 ... 232 psi | 16 bar 1600 kPa 232 psi | 21 bar 2.1 MPa 305 psi | 32 bar 3.2 MPa 464 psi |
0.63 ... 63 bar 63 ... 6300 kPa 9.1 ... 914 psi | 63 bar 6300 kPa 914 psi | 67 bar 6.7 MPa 972 psi | 100 bar 10 MPa 1450 psi |
Low measuring limit | 100 mbar a (1.45 psia) |
- Measuring cell with silicone oil
| 100 mbar a/10 kPa a/1.45 psia |
- Measuring cell with inert liquid
| 100 mbar a/10 kPa a/1.45 psia |
- Measuring cell with neobee
| 100 mbar a/10 kPa a/1.45 psia |
Upper measuring limit | 100 % of max. span |
Input of absolute pressure, with flush-mounted diaphragm | | |
Measured variable | Absolute pressure, flush-mounted |
Span (fully adjustable) or nominal measuring range and max. permissible test pressure | HART | PROFIBUS PA/ FOUNDATION Fieldbus | | |
Span | Nominal measuring range | Max. perm. operating pressure MAWP (PS) | Max. permissible test pressure |
43 ... 1300 mbar a 4.3 ... 130 kPa a 17 ... 525 inH2O a | 1300 mbar a 130 kPa a 525 inH2O a | 2.6 bar a 260 kPa a 37.7 psia | 10 bar a 1 MPa a 145 psia |
160 ... 5000 mbar a 16 ... 500 kPa a 2.32 ... 72.5 psia | 5000 mbar a 500 kPa a 72.5 psia | 10 bar a 1 MPa a 145 psia | 30 bar a 3 MPa a 435 psia |
1 ... 30 bar a 0.1 ... 3 MPa a 14.5 ... 435 psia | 30 bar a 3 MPa a 435 psia | 45 bar a 4.5 MPa a 653 psia | 100 bar a 10 MPa a 1450 psia |
| Depending on the process connection, the span may differ from these values |
Low measuring limit | 0 mbar a/0 kPa a/0 psia |
Upper measuring limit | 100% of maximum measuring span |
Output | HART | PROFIBUS PA/ FOUNDATION Fieldbus |
Output signal | 4 ... 20 mA | Digital PROFIBUS PA signal |
Physical bus | - | IEC 61158-2 |
Protection against polarity reversal | Protected against short-circuit and polarity reversal. Each connection against the other with max. supply voltage. |
Electrical damping (step width 0.1 s) | Set to 2 s (0 ... 100 s) |
Measuring accuracy for gauge pressure | According to IEC 60770-1 |
Reference conditions | - Rising characteristic curve
- Start of scale 0 bar
- Seal diaphragm stainless steel
- Measuring cell with silicone oil
- Room temperature 25 °C (77 °F)
|
Measuring span ratio (spread, Turn-Down) | r = maximum measuring span/set measuring span or nominal measuring range |
Error in measurement at limit setting including hysteresis and reproducibility | |
| |
| r ? 1.25 : 1.25 < r ? 30 : | ? 0.075 % ? (0.008 · r + 0.065) % |
- 1 bar/100 kPa/3.6 psi
4 bar/400 kPa/58 psi 16 bar/1.6 MPa/232 psi 63 bar/6.3 MPa/914 psi 160 bar/16 MPa/2321 psi
| r ? 5 : 5 < r ? 100 : | ? 0.075 % ? (0.005 · r + 0.05) % |
| r ? 3 : 3 < r ? 10 : 10 < r ? 100 : | ? 0.075 % ? (0.0029 · r + 0.071) % ? (0.005 · r + 0.05) % |
Influence of ambient temperature (in percent per 28 °C (50 °F)) | |
| ? (0.16 · r + 0.1) % |
- 1 bar/100 kPa/3.6 psi
4 bar/400 kPa/58 psi 16 bar/1.6 MPa/232 psi 63 bar/6.3 MPa/914 psi 160 bar/16 MPa/2321 psi 400 bar/40 MPa/5802 psi
| ? (0.07 · r + 0.08) % |
Long-term stability (temperature change ±30 °C (±54 °F)) | |
| ? (0.16 · r) % per year |
- 1 bar/100 kPa/3.6 psi
4 bar/400 kPa/58 psi
| ? (0.25 · r) % in 5 years |
- 16 bar/1.6 MPa/232 psi
63 bar/6.3 MPa/914 psi 160 bar/16 MPa/2321 psi 400 bar/40 MPa/5802 psi
| ? (0.125 · r) % in 5 years |
Influence of mounting position | ? 0.05 mbar/0.005 kPa/0.000725 psi per 10° incline (zero point correction is possible with position error compensation) |
Effect of auxiliary power (in percent per voltage change) | 0.005 % per 1 V |
Measured-value resolution for PROFIBUS PA and FOUNDATION Fieldbus | 3 · 10-5 of nominal measuring range |
Measuring accuracy for absolute pressure | According to IEC 60770-1 |
Reference conditions (All error data refer always refer to the set span) | - Rising characteristic curve
- Start of scale 0 bar
- Seal diaphragm stainless steel
- Silicone oil filling
- Room temperature 25 °C (77 °F)
|
Measuring span ratio r (spread, Turn-Down) | r = maximum measuring span/set measuring span or nominal measuring range |
Error in measurement at limit setting including hysteresis and reproducibility | |
| |
| ? 0.1 % |
| ? 0.2 % |
Influence of ambient temperature (in percent per 28 °C (50 °F)) | |
- 250 mbar a/25 kPa a/3.6 psia
| ? (0.15 · r + 0.1) % |
- 1300 mbar a/130 kPa a/18.8 psia
5 bar a/500 kPa a/72.5 psia 30 bar a/3000 kPa a/435 psia
| ? (0.08 · r + 0.16) % |
Long-term stability (temperature change ±30 °C (±54 °F)) | ? (0.25 · r) % in 5 years |
Effect of mounting position (in pressure per change of angle) | ? 0.05 mbar/0.005 kPa/0.000725 psi per 10° incline (zero point correction is possible with position error compensation) |
Effect of auxiliary power (in percent per voltage change) | 0.005 % per 1 V |
Measured-value resolution for PROFIBUS PA and FOUNDATION Fieldbus | 3 · 10-5 of nominal measuring range |
Measuring accuracy for gauge and absolute pressure, with flush-mounted diaphragm | According to IEC 60770-1 |
Reference conditions (All error data refer always refer to the set span) | - Rising characteristic curve
- Start of scale 0 bar
- Seal diaphragm stainless steel
- Silicone oil filling
- Room temperature 25 °C (77 °F)
|
Measuring span ratio r (spread, Turn-Down) | r = maximum measuring span/set measuring span or nominal measuring range |
Error in measurement at limit setting including hysteresis and reproducibility | |
| Gauge pressure with flush-mounted diaphragm | Absolute pressure with flush-mounted diaphragm |
| ? 0.075 % | - |
| ? (0.005 · r + 0.05) % | - |
| - | ? 0.2 % |
| - | ? 0.4 % |
Influence of ambient temperature (in percent per 28 °C (50 °F)) | ? (0.08 · r + 0.16) % | ? (0.16 · r + 0.24) % |
Influence of the process temperature (in pressure per temperature unit) | |
- Temperature difference between process temperature temperature and ambient temperature
| 3 mbar/0.3 kPa/0.04 psi per 10 K |
Long-term stability (temperature change ±30 °C (±54 °F)) | ? (0.25 · r) % in 5 years |
Effect of mounting position (in pressure per change of angle) | 0.4 mbar/0.04 kPa/0.006 per 10° incline (zero point correction is possible with position error compensation) |
Effect of auxiliary power (in percent per voltage change) | 0.005 % per 1 V |
Measured-value resolution for PROFIBUS PA and FOUNDATION Fieldbus | 3 · 10-5 of nominal measuring range |