Differentiation of cases: Ex zones/danger through flammable sample gas (Series 6)
Zone
Gas type
Sample gas non-flammable below the lower explosive limit LEL (25% of LEL)
Sample gas is flammable and/or is rarely, and then only briefly, above the LEL (25% of LEL)
Sample gas is flammable and/or is occasionally above the LEL (25% of LEL)
Category ATEX II 1G (Zone 0)
Not permitted
Not permitted
Not permitted
Category ATEX II 2G (Zone 1)
Analyzer
Ex analyzer Ex p (certificate ATEX 2022X)
Ex analyzer Ex p (certificate ATEX 2022X)
Ex analyzer Ex p (certificate ATEX 2022X)
Gas path
Pipe gas path
Pipe gas path
Pipe gas path
Flame arrestor
—
Flame arrestor in sample gas inlet and outlet
Flame arrestor in sample gas inlet and outlet
Monitoring
Ex p control device (certificate ATEX E 082)
Ex p control device sample gas pressure < 165 hPa, fail-safe (certificate ATEX E 082)
Ex p control device sample gas pressure < 165 hPa, fail-safe (certificate ATEX E 082)
Pressure switch
—
Differential pressure switch (when sample gas pressure is not controlled fail-safely)
Differential pressure switch (when sample gas pressure is not controlled fail-safely)
Category ATEX II 2G (zone 1)
Operating mode "Continuous purging"
Analyzer
Ex analyzer Ex p (certificate ATEX 1708X)
Ex analyzer Ex p (certificate ATEX 1708X)
Ex analyzer Ex p (certificate ATEX 1708X)
Gas path
Pipe gas path
Pipe gas path
Pipe gas path
Flame arrestor
—
Flame arrestor in sample gas inlet and outlet
Flame arrestor in sample gas inlet and outlet
Monitoring
Ex p control device (certificate DMT 99 ATEX E 082)
Ex p control device (certificate DMT 99 ATEX E 082)
Ex p control device (certificate DMT 99 ATEX E 082)
Pressure switch
—
—
—
Category ATEX II 3G (zone 2)
Analyzer
Standard analyzer in field housing (addition E11: certificate ATEX 1686X)
Standard analyzer in field housing (addition E12: certificate ATEX 1697X)
Standard analyzer in field housing (addition E12: certificate ATEX 1697X)
Gas path
Pipe or hose gas path
Pipe gas path
Pipe gas path
Flame arrestor
—
Flame arrestor in sample gas inlet and outlet
Flame arrestor in sample gas inlet and outlet
Monitoring
—
Ex p control device (certificate DMT 99 ATEX E 082)
Ex p control device (certificate DMT 99 ATEX E 082)
Non-hazardous zone
Analyzer
Analyzer as rack unit or in field housing
Analyzer as rack unit or in field housing
Analyzer as rack unit or in field housing
Gas path
Pipe or hose gas path
Pipe gas path, recommended
Enclosure purging with inert gas (N2) recommended
Pipe gas path, recommended
Enclosure purging with inert gas (N2) recommended
Flame arrestor
—
Flame arrestor in sample gas inlet and outlet
Flame arrestor in sample gas inlet and outlet
Monitoring
—
—
Simplified monitoring of purging recommended
Ex configurations – principle selection criteria (Series 6)
Signal line routing
Within zone 1
From zone 1 to zone 2
From zone 1 to non-hazardous zone
Ex i isolating transformer
Required
Required under certain conditions
(when energy feedback cannot be excluded)
Required under certain conditions
(when energy feedback cannot be excluded)
Isolating relay
Required
Required under certain conditions
(when energy feedback cannot be excluded)
Required under certain conditions
(when energy feedback cannot be excluded)
Additional units, selection criteria (ATEX 2G)
Differentiation of cases: Ex zones/danger through flammable sample gas (OXYMAT 7)
Wall-mounted device
Zone
Gas type
Sample gas non-flammable below the lower explosive limit LEL (25% of LEL)
Sample gas is flammable and/or is rarely, and then only briefly, above the LEL (25% of LEL)
Sample gas is flammable and/or is occasionally above the LEL (25% of LEL)
Category ATEX II 1G (zone 0)
Not permitted
Not permitted
Not permitted
Category ATEX II 2G (Zone 1) Operating mode "continuous purging"
Analyzer
Ex analyzer Ex p (Certificate BVS 14 ATEX E 153 X and IECEx BVS 14.0104X)
Ex analyzer Ex p (Certificate BVS 14 ATEX E 153 X and IECEx BVS 14.0104X)
Ex analyzer Ex p (Certificate BVS 14 ATEX E 153 X and IECEx BVS 14.0104X)
Flame arrestor
Not required
Flame arrester in sample gas and reference gas paths:
Flame arrester in sample gas and reference gas paths:
Monitoring
Permitted px monitoring system
Permitted px monitoring system
Permitted px monitoring system
Category ATEX II 3G (Zone 2) Operating mode "continuous purging"
Analyzer
Ex analyzer Ex p (Certificate BVS 14 ATEX 153 E X und IECEx BVS 14.0104X)
Ex analyzer Ex p (Certificate BVS 14 ATEX 153 E X und IECEx BVS 14.0104X)
Ex analyzer Ex p (Certificate BVS 14 ATEX E 153 X and IECEx BVS 14.0104X)
Flame arrestor
Not required
Flame arrester in sample gas and reference gas paths:
Flame arrester in sample gas and reference gas paths:
Monitoring
Permitted pz monitoring system
Permitted px monitoring system
Permitted px monitoring system
Category ATEX II 3G (Zone 2) Operating mode "gas-fume-proof"
Analyzer
Ex analyzer Ex nR (Certificate BVS 14 ATEX E 153/154 X and IECEx BVS 14.0104X)
Not permitted
Not permitted
Flame arrestor
Not required
Not required
Not required
Ex configurations – principle selection criteria
Field device
Zone
Gas type
Sample gas non-flammable below the lower explosive limit LEL (25% of LEL)
Sample gas is flammable and/or is rarely, and then only briefly, above the LEL (25% of LEL)
Sample gas is flammable and/or is occasionally above the LEL (25% of LEL)
Category ATEX II 1G (zone 0)
Not permitted
Not permitted
Not permitted
Category ATEX II 2G (Zone 1) Operating mode "pressurized enclosure"
Analyzer
Ex analyzer Ex d (Certificate BVS 15 ATEX E 038 X and IECEx BVS 15.0030X)
Ex analyzer Ex d (Certificate BVS 15 ATEX E 038 X and IECEx BVS 15.0030X)
Ex analyzer Ex d (Certificate BVS 15 ATEX E 038 X and IECEx BVS 15.0030X)
Flame arrestor
Integrated
Integrated
Integrated
Ex configurations – principle selection criteria
Rack-mounted device
Zone
Gas type
Sample gas non-flammable below the lower explosive limit LEL (25% of LEL)
Sample gas is flammable and/or is rarely, and then only briefly, above the LEL (25% of LEL)
Sample gas is flammable and/or is occasionally above the LEL (25% of LEL)
Category ATEX II 1G (zone 0)
Not permitted
Not permitted
Not permitted
Category ATEX II 2G (zone 1)
Not permitted
Not permitted
Not permitted
Category ATEX II 3G (Zone 2) Type of protection "nA"
Analyzer (Installation in a suitable enclosure IP54)
Ex analyzer Ex nA (Certificate BVS 15 ATEX E 007 X and IECEx BVS 15.0007X)
Ex analyzer Ex nA (Certificate BVS 15 ATEX E 007 X and IECEx BVS 15.0007X)
Not permitted
Flame arrestor
Not required
Not required
Not required
Ex configurations – principle selection criteria
Use of OXYMAT 6 in hazardous area and/or for measurement of flammable gases
Article No.
Certification (short codes)
Additional unit
Gas
Dust
Purging unit
Flame arrestor
Pressure switch
Ex i isolating transformer
Ex i isolating relay
Category
Operating mode
Ex zone
Ex zone
7MB8000-
7MB8000-
7MB8000-
7MB8000-
7MB8000-
ATEX II 2G (Zone 1)
Leakage compensation
7MB2011-***0*-2***
X
—
2BB
6BA/6BB1)
5AA1)
3AB1)
4AB1)
7MB2011-***0*-3***
X
—
2BA
6BA/6BB1)
5AA1)
3AB1)
4AA1)
Continuous purging
7MB2011-***0*-6***
X
—
2CB
6BA/6BB1)
o
3AB1)
4AB1)
7MB2011-***0*-7***
X
—
2CA
6BA/6BB1)
o
3AB1)
4AA1)
ATEX II 3G (Zone 2)
Flammable gases
7MB2011-***0*-0***
E12
—
2CB
6BA/6BB1)
o
o
o
7MB2011-***0*-1***
E12
—
2CA
6BA/6BB1)
o
o
o
7MB2011-***0*-0***
E42
—
6BA/6BB1)
o
o
o
7MB2011-***0*-1***
E42
—
6BA/6BB1)
o
o
o
Non-flammable gases
7MB2011-***0*-0***
E11
—
o
o
o
o
o
7MB2011-***0*-1***
E11
—
o
o
o
o
o
7MB2011-***0*-0***
E41
—
o
o
o
o
7MB2011-***0*-1***
E41
—
o
o
o
o
Non-hazardous zone
Non-hazardous gas zone
7MB2011-***0*-0***
E40
—
o
o
o
o
7MB2011-***0*-1***
E40
—
o
o
o
o
7MB2011-***0*-0***
—
—
o
o
o
o
o
7MB2011-***0*-1***
—
—
o
o
o
o
o
7MB2021-****0-****
X
—
o
o
o
o
o
CLASS 1
Div 2
Flammable and non-flammable gases
7MB2011-***0*-0***
E20
—
1AA
6BA/6BB
o
o
o
7MB2011-***0*-1***
E20
—
1AA
6BA/6BB
o
o
o
7MB2021-*****-****2)
E20
—
1AA
6BA/6BB
o
o
o
— Combination not allowed
X Possible combination, no additional data required
o Not required
Ex configurations, possible combinations
1) Required under certain conditions: see table of Ex configurations, selection criteria
2) Installation in additional housing required
Use of ULTRAMAT 6 in hazardous area and/or for measurement of flammable gases
Article No.
Certification and short codes
Additional unit
Gas
Dust
Purging unit
Flame arrestor
Pressure switch
Ex i isolating transformer
Ex i isolating relay
Category
Operating mode
7MB2111- 7MB2112-
Ex zone
Ex zone
7MB8000-
7MB8000-
7MB8000-
7MB8000-
7MB8000-
ATEX II 2G (Zone 1)
Leakage compensation
*****-2*A*
X
—
2BB
—
5AA1)
3AB1)
4AB1)
*****-3*A*
X
—
2BA
—
5AA1)
3AB1)
4AA1)
Continuous purging
*****-6*A*
X
—
2CB
—
o
3AB1)
4AB1)
*****-7*A*
X
—
2CA
—
o
3AB1)
4AA1)
ATEX II 3G (Zone 2)
Flammable gases
*****-0*A*
E42
—
—
o
o
o
*****-1*A*
E42
—
—
o
o
o
*****-0*A*
E12
—
2CB
o
o
o
o
*****-1*A*
E12
—
2CA
o
o
o
o
Non-flammable gases
*****-0*A*
E41
—
o
o
o
o
*****-1*A*
E41
—
o
o
o
o
*****-0*A*
E11
—
o
o
o
o
o
*****-1*A*
E11
—
o
o
o
o
o
Non-hazardous zone
Non-hazardous gas zone
*****-0*A*
X
E40
—
o
o
o
o
*****-1*A*
X
E40
—
o
o
o
o
CLASS 1
Div 2
Flammable and non-flammable gases
7MB2111-*****-0*A*
E20
—
1AA3)
6BA/6BB3)
o
o
o
7MB2111-*****-1*A*
E20
—
1AA3)
6BA/6BB3)
o
o
o
7MB212*-*****-****2)
E20
—
1AA3)
6BA/6BB3)
o
o
o
— Combination not allowed
X Possible combination, no additional data required
o Not required
1) Required under certain conditions; see table of Ex configurations, selection criteria.
2) Installation in additional housing required
3) When using flammable gases, only in coordination with relevant authority
Use of CALOMAT 6 in hazardous area and/or for measurement of flammable gases
Article No.
Certification
Additional unit
Gas
Dust
Purging unit
Flame arrestor
Pressure switch
Ex i isolating transformer
Ex i isolating relay
Category
Operating mode
7MB8000-
7MB8000-
7MB8000-
7MB8000-
7MB8000-
ATEX II 2G (Zone 1)
Leakage compensation
7MB2511-***0*-0AE*
X
—
2BB
6BA/6BB1)
5AA1)
3AB1)
4AB1)
7MB2511-***0*-1AE*
X
—
2BA
6BA/6BB1)
5AA1)
3AB1)
4AA1)
Continuous purging
7MB2511-***0*-0AF*
X
—
2CB
6BA/6BB1)
o
3AB1)
4AB1)
7MB2511-***0*-1AF*
X
—
2CA
6BA/6BB1)
o
3AB1)
4AA1)
ATEX II 3G (Zone 2)
Flammable gases
7MB2511-***0*-*AJ*
X
X
—
6BA/6BB1)
o
o
o
7MB2511-***0*-*AC*
X
—
2CA/2CB
6BA/6BB1)
o
o
o
7MB2521-***0*-*AB*2)
X
—
In accord. w/certificate
6BA/6BB1)
o
o
o
Non-flammable gases
7MB2511-***0*-*AH*
X
X
—
o
o
o
o
7MB2511-***0*-*AB*
X
—
o
o
o
o
o
7MB2521-***0*-*AB*
X
—
o
o
o
o
o
Non-hazardous zone
Non-hazardous gas zone
7MB2511-***0*-*AG*
—
X
—
o
o
o
o
CLASS 1 Div 2
Flammable and non-flammable gases
7MB2511-***0*-*AD*
X
—
1AA3)
6BA/6BB3)
o
o
o
7MB2521-***0*-*AD*2)
X
—
1AA3)
6BA/6BB3)
o
o
o
X Possible combination, no additional data required
— Combination not allowed
o Not required
Ex configurations, possible combinations
1) Required under certain conditions; see table of Ex configurations, selection criteria
2) Installation in additional housing required
3) When using flammable gases, only in coordination with relevant authority
Use of ULTRAMAT 23 in hazardous area and/or for measurement of flammable gases
Article No.
Certification and short codes
Additional unit
Gas
Dust
Purging unit
Flame arrestor
Pressure switch
Ex i isolating transformer
Category
Operating mode
7MB233*-
Ex zone
Ex zone
7MB8000-
7MB8000-
7MB8000-
7MB8000-
7MB8000-
ATEX II 3G (zone 2)
Non-flammable gases 2)
*****-****
E20
—
o
o
o
o
o
CLASS 1 Div 2
Flammable and non-flammable gases 2)
*****-****
E20
—
1AA
o1)
o
o
o
— Combination not allowed
o Not required/not defined
1) Required under certain conditions
2) Installation in additional housing required
Use of OXYMAT 7 in hazardous area and/or for measurement of flammable gases
Certification
Additional unit
Gas
Dust
Purging unit
Flame arrestor
Category
Operating mode
Article No.
Ex zone
Ex zone
7MB8000-
Zone 1
Wall-mounted device Ex p
Flammable gases
7MB3000-(3/4)A*0*-**(E/G)0
X
X1)
px monitoring system
6BA/6BB
7MB3020-2**10-*AA0
Field device Ex d
Flammable gases
7MB3000-6AX00-(0/6)*H0
X
—
o
Integrated
7MB3020-(4/5)(C/D)*10-1AA0
Zone 2
Wall-mounted device Ex p
Non-flammable gases
7MB3000-(3/4)A*0*-**(E/G)0
X
X1)
pz monitoring system
o
7MB3020-2(C/D)*10-*AA0
Wall-mounted device Ex nR
Non-flammable gases
7MB3000-(3/4)A*0*-**(D/G)0
X
X1)
o
o
7MB3020-2(C/D)*10-*AA0
19" rack unit Ex nA 2)
Flammable gases
7MB3000-0A*0*-**C0
X
—
—
o
7MB3020-2(C/D)*10-*AA0
— Combination not allowed
X Possible combination, no additional data required
o Not required/not defined
1) Variant "G" only
2) Installation in an IP54 enclosure in accordance with EN/IEC 60079-0 and EN/IEC 60079-15
Overview of the Ex certificates (Series 6)
Ex approval
ATEX
CLASS I Div 2
ATEX
2G-LC
2G - CP
3G burn.
3G nbrn.
FM
CSA
3D (dust)
See base article number
Z + E12
Z + E11
Z + E20
Z + E20
Z + E4X
except CALOMAT
except CALOMAT
Field device
U6F
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1686 X
3016050
1526657
ATEX 2278 X
U6F-S
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1686 X
3016050
1526657
—
O6F
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1686 X
3016050
1526657
ATEX 2278 X
O6F-S
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1686 X
3016050
1526657
—
C6F
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1697 X
3018862
1526660
ATEX 2278 X
C6F-S
ATEX 2022 X
ATEX 1708 X
ATEX 1697 X
ATEX 1697 X
o
1526657
—
C62F
ATEX 2022 X
ATEX 1708 X
—
—
—
—
—
C62F-S
ATEX 2022 X
ATEX 1708 X
—
—
—
—
—
19" rack unit
U6E
—
—
—
3016050
1526657
—
U6E-S
—
—
—
3016050
1526657
—
O6E
—
—
—
3016050
1526657
—
O6E-S
—
—
—
3016050
1526657
—
OU6E
—
—
—
3016050
1526657
—
OU6E-S
—
—
—
3016050
1526657
—
C6E
—
—
ATEX 1873 X
3018862
1526660
—
C6E-S
—
—
ATEX 1873 X
o
1526660
—
C62E
—
—
—
—
—
—
C62E-S
—
—
—
—
—
—
O61
—
—
—
—
—
—
O64
—
—
—
—
—
—
F6
—
—
—
(SET)
—
—
—
U23
—
—
ATEX 0027 x
3035269
2133209
—
U23 O2p
—
—
ATEX 0027 x
3035269
2133209
—
U23 H2S
—
—
ATEX 0027 x
3035269
2133209
—
CP = Continuous Purging
LC = Leakage compensation
… -S = Special application
burn. = Flammable gases
nbrn. = Non-flammable gases
o = In progress
Possible combinations of PROFIBUS with Ex applications
ATEX 2022 X
PROFIBUS PA Ex-i
FM/CSA
PROFIBUS PA or DP
ATEX 1708 X
PROFIBUS PA Ex-i
ATEX 1697 X
PROFIBUS PA Ex-i
ATEX 2278 X
PROFIBUS DP - Non-hazardous gas installation
(Z + E40)
ATEX 1686 X
PROFIBUS DP
ATEX 2278 X
PROFIBUS DP in combination ATEX 1686 X
(Z + E41)
ATEX 1873 X
PROFIBUS DP
ATEX 2278 X
PROFIBUS PA Ex-i in combination with ATEX 1697 X
(Z + E42)
skener.ru
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Àðìàòóðà DENDOR
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Äàò÷èêè è èçìåðèòåëè
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Ðåãóëÿòîðû è ðåãèñòðàòîðû
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Ïíåâìàòè÷åñêîå îáîðóäîâàíèå
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Êðàíû è Êëàïàíû
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Èçìåðèòåëüíûå ïðèáîðû
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Ñèñòåìû áåñïðîâîäíîãî óïðàâëåíèÿ «óìíûé äîì»
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Áåñêîíòàêòíûå âûêëþ÷àòåëè Êîíå÷íûå âûêëþ÷àòåëè Îïòè÷åñêèå äàò÷èêè Ýíêîäåðû
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FS - Óñòàíîâêà óìÿã÷åíèÿ
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 23
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/kip/kip.php on line 30
SKW-FK - Óñòàíîâêà îáåçæåëåçèâàíèÿ