General |
Weight | Approx. 3 kg |
Measuring ranges |
Number of measuring ranges | Max. 4; parameters can be assigned freely |
Parameters can be assigned in the measuring ranges | |
| 0.5% H2 in N2 |
- Largest possible measuring span
| 100% H2 in N2 |
- Smallest possible span with suppressed zero point
| 5% (e.g. 95% to 100%) H2 in N2 |
Gas inlet conditions |
Sample gas pressure | 700 to 1200 hPa (abs.) |
Pressure drop between sample gas inlet and sample gas outlet | < 50 hPa at 1.5 l/min |
Sample gas flow | 30 to 90 l/h (0.5 to 1.5 l/min) |
Sample gas temperature | 0 to 70 °C |
Sample gas humidity (rel. humidity) | < 90% (condensation inside the gas path is to be avoided) |
Sample chamber temperature |
Standard version | Approx. 72 °C |
Time response |
Warm-up period at room temperature | < 30 min (max. accuracy after 2 h) |
Response characteristics | |
- Delay display T90 with device-internal signal damping (low pass filter) of 1 s
| < 2.5 s |
- Dead time (T10) at 1 l/min
| < 0.5 s |
- Adjustable signal damping range
| 0 to 100 s |
Measuring response |
Output signal fluctuation with device-internal signal damping of 1 s | ? ± 0.5% of the smallest span acc. to nameplate (? < ± 8.33 vpm H2) |
Detection limit | ? 1% of the smallest measuring span according to nameplate |
Measured-value drift | ? ± 1%/week of smallest span according to nameplate or ? 50 vpm H2 / week, whichever is greater |
Repeatability | ? ± 1% of the current measuring span or 100 vpm H2 |
Linearity error | ? ± 1% of the current measuring span or 100 vpm H2 |
Influencing variables |
Ambient temperature | ? ± 0.5%1)/10 K of the current measuring span or ? ± 50 vpm H2 / 10 K |
Sample gas pressure | ? ± 0.5 %1) of the current measuring span/1% pressure variation or ? ± 50 vpm H2 / 1% pressure change |
Sample gas flow | ? ± 0.2% of the smallest possible measuring span with a change in flow of 1 dl/min within the permissible flow range |
Accompanying gases (interference gases) | The interference gas sensitivity depends on the application and must be determined in each case except for applications with blast furnace gas / converter gas / wood gasification (pre-adjusted). |
Supply voltage | ? ± 0.1% of full-scale value (within the nominal range of use) |
Electrical inputs and outputs |
Analog and digital interfaces | See base unit |
Climatic conditions |
Storage and transport | -30 to 70 °C |
Permissible ambient temperature (during operation in base unit)2) | 0 to 50 °C |
Relative humidity (RH) during storage, transport or operation | < 90% (condensation on the installed components is to be avoided) |
Gas connections | |
Connection fittings | Pipe connection with 6 mm outer diameter |
Materials of wetted parts |
Gas connection | Stainless steel material no. 1.4571 |
Clamping rings and union nut (set) | Stainless steel material no. 1.4401 |
Sample gas pipes | Stainless steel material no. 1.4404 |
Sensor mounting block | Stainless steel material no. 1.4571 |
Sensor | Si, SiOxNy, Au, epoxy resin, glass |
Gasket, contained in the sensor module | Perfluorelastomere FFKM |