General | |
Measuring ranges | 4, switchable internally and externally; autoranging is also possible |
Smallest possible measuring range | Depending on application, e.g., CO: 0 … 10 vpm, CO2: 0 … 5 vpm |
Largest possible measuring span | Depending on application |
Measuring range with suppressed zero | Every zero possible within 0 to 100 Vol.%, smallest possible measuring span 20% |
Characteristic | Linearized |
Position of use | Front panel vertical |
Conformity | CE identification EN 50081-1, EN 50082-2 |
Design, enclosure | |
Weight | Approx. 15 kg (with one IR channel), approx. 21 kg (with two IR channels) |
Degree of protection | IP20 according to EN 60529 |
Electrical characteristics | |
Electromagnetic compatibility (EMC) | According to standard requirements of NAMUR NE21 (08/98) |
Electrical safety | According to EN 61010-1, overvoltage category III |
Power supply | 100 to 120 V AC (rated range 90 to 132 V), 48 to 63 Hz or 200 to 240 V AC (rated range 180 to 264 V), 48 to 63 Hz |
Power consumption | 1-channel unit: approx. 40 VA 2-channel unit: approx. 70 VA |
Fuses | |
| 1T/250 (7MB2121), 1.6T/250 (7MB2123) |
| 0.63T/250 (7MB2121), 1T/250 (7MB2123) |
Gas inlet conditions | |
Perm. sample gas pressure | |
| |
| 600 to 1500 hPa (absolute) |
| 600 to 1300 hPa (absolute) |
- for analyzers with pipes (without pressure switch)
| 600 to 1500 hPa (absolute) |
Sample gas flow | 18 to 90 l/h (0.3 to 1.5 l/min) |
Sample gas temperature | 0 to 50 °C |
Sample gas humidity | < 90 % RH (relative humidity) or depending on application, non condensing |
Time response | |
Warm-up period | With amb. temperature < 30 min (maximum accuracy achieved after 2 hours) |
Response time (T90 time) | Dependent on length of analyzer cell, sample gas line and damping |
Damping (electric time constant) | 0 to 100 s, programmable |
Dead time (purging time of gas path in analyzer at 1 l/min) | Approx. 0.5 to 5 s, depending on version |
Time for internal signal processing | < 1 s |
Pressure correction range | |
Pressure sensor | |
| 600 to 1200 hPa absolute |
| 600 to 1500 hPa absolute |
Measuring response (maximum accuracy achieved after 2 hours) |
Output signal fluctuation | ± 0.1 % to ± 1 % of smallest possible measuring range specified on rating plate depending on application with the unit specific electronic time constant (corresponds to ± 0.33% at 2?) |
Zero drift | < 1% of measuring range/week |
Measured-value drift | < 1% of measuring range/week |
Repeatability | ? 1% of respective measuring range |
Minimum detectable quantity | 1% of smallest measuring range |
Linearity error | < 0.5% of full-scale value |
Influencing variables (referred to 1000 hPa sample gas pressure, 0.5 l/min sample gas flow and 25 °C ambient temperature) |
Ambient temperature | < 1% of measuring range/10 K (for a constant temperature of the reception cell) |
Sample gas pressure | With pressure compensation: < 0.15% of span/1% change in atmospheric pressure without pressure compensation: < 1.5% of span/1% change in atmospheric pressure |
Sample gas flow | Negligible |
Power supply | < 0.1% of output signal span with rated voltage ± 10% |
Ambient conditions | Application-dependent influencing of measurement if ambient air contains measured component or cross-sensitive gases |
Electric inputs and outputs | |
Analog output | 0/2/4 to 20 mA. floating; load ? 750 ? |
Relay outputs | 6, with changeover contacts, freely parameterizable, e.g. for range identification; loading capacity: 24 V AC/DC / 1 A floating, non sparking |
Analog inputs | 2, designed for 0/2/4 to 20 mA, for external pressure sensor and correction of influence of residual gas (correction of cross interference) |
Binary inputs | 6, designed for 24 V, floating, freely parameterizable, e.g. for range switching |
Serial interface | RS 485 |
Options | Autocal function with 8 additional binary inputs and 8 relay outputs, also with PROFIBUS PA and PROFIBUS DP |
Ambient conditions | |
Permissible ambient temperature | -30 to +70 °C during storage and transport, +5 to +45 °C during operation |
Permissible humidity | < 90 % RH (relative humidity) as annual average, during storage and transport (dew point must not be fallen below) |