Analytical performance |
|
|
Measuring range |
Internally adjustable |
|
Detection limit at standardized conditions: |
O2: 200 ppm CO: 0.6 ppm |
|
Linearity (under standard conditions) |
Better than 1% |
|
Repeatability (under standard conditions) |
O2: 1% of the measuring range CO: 0.5 % of the measuring range |
|
General information |
|
|
Design |
Transmitter and detector units, connected by a sensor cable |
|
Materials |
|
|
Installation |
In-situ or bypass |
|
Concentration units |
ppm, vol. %, mg/Nm3 |
|
Display |
Digital concentration display (4 digits with floating decimal point) |
|
Laser protection class |
Class 1, safe to the eye |
|
Explosion protection |
Optionally, according to
|
|
Design, enclosure |
|
|
Degree of protection |
IP65 according to EN 60529 |
|
Dimensions |
For each unit (transmitter, detector)
|
|
Purging tube |
Length, outer diameter, inner diameter: |
|
Weights |
|
|
6.0 kg |
|
5.2 kg |
|
|
|
5.3 kg |
|
Approx. 12 kg |
|
Connection dimension customer flange |
DN 50/PN 25 or ANSI 4"/150 lbs |
|
Electrical characteristics |
|
|
Power supply |
24 V DC nominal (18 ... 30.2 V DC) |
|
Power consumption, maximum |
10 VA |
|
EMC |
In accordance with EN 61326-1 |
|
Electrical safety |
In accordance with EN 61010-1 |
|
Fuse specifications |
T1.6L250V |
|
Dynamic performance |
|
|
Warm-up time at 20 °C ambient temperature |
Approx. 15 min |
|
Response time (T90) |
Approx. 2 s, depends on application |
|
Integration time |
0 ... 100 s, selectable |
|
Influencing variables |
|
|
Variations in ambient temperature |
< 0.5%/10 K of the measuring range |
|
Process gas temperature |
With compensation: < 1 %/100 K of the measuring range |
|
Variations in ambient pressure |
Negligible |
|
Process gas pressure |
O2: With compensation: < 1%/4 000 hPa of the measuring range CO: Negligible |
|
Variations in supply voltage |
Negligible |
|
Electrical inputs and outputs |
|
|
Number of measurement channels |
1 |
|
Analog outputs |
2 outputs, 4 ... 20 mA, floating, ohmic resistance max. 660 . External isolating power supplies may have to be provided by the customer. |
|
Analog inputs |
2 inputs, designed for 4 ... 20 mA, 120 |
|
Binary outputs |
2 outputs, with switchover contacts, configurable, 24 V/0.5 A, floating, single pole double throw (SPDT) |
|
Binary input |
1 input, designed for 24 V, floating, configurable |
|
Service port |
Ethernet 10BaseT (RJ-45) |
|
RS 485 PROFIBUS DPV0 version |
Two-wire interface, up to 3 Mbit/s, -7 … 12 V |
|
RS 485 Modbus version |
Two-wire interface, up to 115 200 bit/s, -7 … 12 V |
|
Cable to customer interface (not included in standard delivery, ATEX or optional) |
|
|
Analog connection cable |
10 x 2, with shielding in twisted-pair configuration (depending on type and number of I/Os used) |
|
PROFIBUS DP connection cable |
1 x 2 + 4 (PROFIBUS DP hybrid cable) |
|
Modbus connection cable |
1 x 2 + 3, with shielding in twisted-pair configuration |
|
Cable length for ATEX configuration |
3 m |
|
Conductor cross-section |
Min. 0.34 mm |
|
Cable diameter |
8 ... 12 mm or 13 ... 18 mm |
|
Minimum bending radius ATEX-PROFIBUS |
110 mm |
|
Sensor cable (not included in standard delivery, ATEX or optional) |
|
|
Sensor cable type |
4 x 2, with shielding, in twisted-pair configuration |
|
Conductor cross-section |
Min. 0.34 mm |
|
Cable sheath |
PUR (polyurethane) |
|
Dimensions |
|
|
Minimum bending radius |
ATEX: 85 mm |
|
Climatic conditions |
|
|
Ambient temperature range |
|
|
Temperature range on the sensor side of the process interface (connection plate) |
-20 ... +70 °C |
|
Atmospheric pressure |
800 ... 1100 hPa |
|
Humidity |
< 100 % rel. humidity |
|
Measuring conditions |
|
|
Measurement path |
0.3 ... 8 m (other lengths: please contact Siemens) |
|
Process gas pressure, temperature |
|
|
Dust load |
The influence of a high dust load is complex, and depends on the optical path length and particle size distribution. |
|
Purging |
|
|
Purging gas |
|
|
O2 application: Purity better than 99.7 % in order to achieve full performance. For oxygen measurements, an O2 content < 0.01 vol. % in the purging gas is recommended. |
|
< -10 °C, condensation on the optics must be avoided |
|
Sensor purging |
|
|
500 hPa |
|
0 ... +55 °C |
|
O2 application: When commissioning a sensor enclosure previously filled with air: 3 ... 5 l/min (for at least 15 min), subsequently: at least 0.25 l/min |
|
Purging on process side (optional) |
|
|
2 000 ... 8 000 hPa |
|
Dependent on process gas pressure, process gas velocity, dust load, moisture, etc. up to max. 50 l/min |