Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/new/cat.php on line 184
Climatic conditions |
|
Ambient temperature |
|
|
-20 … 30 °C |
|
-20 … 50 °C |
|
5 … 30 °C |
Relative humidity |
< 90%, non-condensing |
Corrosive atmosphere |
No |
Auxiliary media |
|
Compressed air |
Approx. 6 000 hPa, purified instrument air, free of oil, water and dust |
Volume flow |
1 200 l/h |
Power supply |
|
Supply voltage |
|
Frequency |
48 … 63 Hz |
Type of connections |
|
Pipe material |
Stainless steel |
Connections |
|
Cabling |
|
Electrical design |
According to IEC |
Type of cables |
Non-armored cables |
Cable ID |
No labeling of individual cores |
Installation |
|
Site |
Interior |
Ex-zone analyzer |
ATEX II 3G |
System design |
|
Version |
|
Degree of protection |
IP54 |
Automatic calibration |
Optional |
Signal outputs |
4 … 20 mA / floating contact |
With sample gas return flow |
No |
Add-on electronics (option) |
AUTOCAL extension
|
Measuring response |
See technical specifications of FIDAMAT 6 |
Influencing variables |
See technical specifications of FIDAMAT 6 |
Analyzer ID |
Ex II 3G EEx n T3 |
FIDAMAT 6 without pump, with heated oven, with combustion air connection |
|||||
---|---|---|---|---|---|
Gases |
Inlet pressure |
Sample/calibration gas operating pressure |
Flow through FID |
Flow through bypass |
|
Without |
With |
||||
hPa (abs.) |
hPa (abs.) |
hPa |
|||
Combustion gas |
3 000 ... 5 000 |
2 000 ± 20 |
~ 25 |
— |
|
Combustion air |
3 000 ... 5 000 |
1 485 ± 5 |
— |
~ 320 |
~ 300 |
Sample gas |
~ 2 000 |
— |
1 500 ± 2 |
~ 3 |
~ 500 |
Zero gas |
~ 2 000 |
— |
1 500 ± 2 |
~ 3 |
~ 500 |
Calibration gas |
~ 2 000 |
— |
1 500 ± 2 |
~ 3 |
~ 500 |
Gas inlet conditions