Siemens
СРЕДСТВА ПРОМЫШЛЕННОЙ АВТОМАТИЗАЦИИ
официальный партнер Сименс
Каталог СА01 2012
архивный
(4872) 700-366
skenergo@mail.ru

Deprecated: Function eregi() is deprecated in /home/h101150-2/siemens71.ru/docs/new/cat.php on line 184
Set CEM 1

Starting with a mounting frame with sample preparation system, it is possible to add additional units as options. These include:

  • Sampling probe with weather protection hood
  • Heated sample gas line
  • Analyzers
  • Air-conditioning unit
  • NO2/NO converter
  • Sample preparation extension for an additional ULTRAMAT 23 analyzer
  • Single and dual (electrically isolated, not electrically isolated) analog signal processing
  • Power supply modules (115 V, 230 V, 400 V)
  • Outer panels with steel-plate door or with window
  • Single-pole and double-pole fusing
  • Condensation bottle
  • Coalescence filter
Sampling probe

The standard probe is fitted with a DIN flange DN 65, PN 6. The probe is provided with a regulated heater, and has a power consumption of 400 VA. It is supplied with a weather protection hood and 2 µm filter. The maximum dust concentration at the sampling point should not exceed 2 g/m . The sampling pipe is 1 000 mm long, is made of stainless steel, and has dimensions of 20 x 1.5 mm. The sample gas temperature must not exceed 600 °C.

It is also possible to purchase the Set CEM 1 without sample probe.

Heated sample gas line

The temperature of the heated line is regulated at 200 °C by a temperature controller. The power consumption is 100 VA per meter. The internal core is made of PTFE 4/6. The heated line can be up to 35 m in length. Lengths greater than 35 m can be provided upon special request. If desired, the system can also be supplied without a heated sample gas line.

Mounting frame

The basis of each CEM 1 set is the mounting frame with hinged frame (40 HU) for installation of up to five 19" rack units. The mounting frame includes a standardized sample preparation system designed for an ULTRAMAT 23.

The sample preparation system includes a 3/2-way solenoid valve, 3-way switchover ball valve, regulating valve, corrosion-resistant sample gas pump (power consumption 60 VA), condensation trap, room air suction filter with filter element, LOGO for digital display of individual signals in the cabinet, 24 V DC power supply unit (power consumption 70 VA). Also included are a sample gas cooler (power consumption 200 VA) with integral preliminary condensation trap, heat exchanger, hose pump, moisture sensor with flow cell and Teflon filter. Teflon tubes connect the components.

The external dimensions without plinth are 2 000 x 800 x 800 mm (H x W x D). A cabinet depth of 600 mm is also optionally available (not suitable for LDS 6). Hoses and cables can be connected from the left or right. A distance of 500 mm must be provided on the left or right at the installation site for introduction of the hoses and cables.

In addition to the sheet-steel mounting frames for indoor installation, an FRP version is also available for outdoor use. The FRP cabinet is always provided complete with side panels and plinth. The external dimensions are 2 080 x 800 x 600 mm (H x W x D). The GRP cabinet cannot be combined with the LDS 6.

Preparation of sample preparation system for second ULTRAMAT 23

The standard system with sample preparation system and electronics is prepared for one ULTRAMAT 23. If a second ULTRAMAT 23 is to be fitted, this option must be selected so that the sample preparation system and electronics are extended accordingly.

Additional filter

In addition to the fine filter and moisture filter which are always present, a coalescence filter can be optionally fitted in the sample preparation system.

Side panels with doors

Optional outer panels can be selected for the sheet-steel mounting frames. This possibility allows use of the CEM 1 set in analysis cabinets as a rack design on one hand, or on the other as a cabinet design in halls requiring degree of protection IP54. Either a sheet-steel door without window or a glass door can be selected.

Plinth

Plinths with a height of 100 and 200 mm are additionally available.

Cabinet cooling and ventilation

Optionally available are a fan with outlet filter, an air-conditioning unit for indoor installation, and an air-conditioning unit for outdoor installation. The system can be ordered without a fan or air-conditioning unit if the side panels and the door with window are omitted.

The fan with outlet filter has a power consumption of 60 VA, and is fitted in the cabinet wall. The delivery also includes a thermostat with a power consumption of 25 VA.

The air-conditioning unit has a cooling power of 820 VA.

Frost protection heater

The power consumption of the optional cabinet heater is 500 VA. The delivery includes a thermostat with a power consumption of 25 VA for controlling the frost protection heater.

Fusing of the analog signals

In addition to single-pole fusing of the electronic consumers, it is possible to provide double-pole fusing.

The double-pole fuse is mainly required in Benelux countries.

Removal of condensation

A 19 liter condensation bottle can be provided as an option. It is also possible to order the system without a condensation bottle if the condensation can be removed on-site.

NO2/NO converter

The mounting frame and cabinets can be optionally extended by a 19" rack unit with NO2/NO converter with carbon cartridge. The power consumption is 520 VA. The flow rate is 90 l/h. An NO2/NO converter is required if the share of NO2 in the total NOx is greater than 5% and/or if total NOx is to be always determined.

Power supply

The system can be designed either for 115 V AC, 230 V AC or 400 V AC (-15 %, +10 %) with 50 or 60 Hz.

Three phases, neutral and ground must be provided by the customer at 400 V AC.

Analog signal processing

As standard, the analog signals are simply connected to isolating terminals. As an option, the analog signals can be processed twice without electrical isolation by a diode module, or twice with electrical isolation.

Analyzers

The standardized set is prepared for an ULTRAMAT 23. The system can be supplemented by a second ULTRAMAT 23, OXYMAT 6 and/or LDS 6. Various measured components and measuring ranges can be selected. Other combinations of components and ranges are available on request. In such a case it must be checked whether the desired certificates and approvals are available. The analyzers used, the measured components and the measuring ranges are described briefly below.

Details on the analyzers, alternative components and ranges can be found under the topics "Continuous gas analyzers, extractive" and "Continuous gas analyzers, in-situ".

ULTRAMAT 23: CO, NO

For measuring two infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 150 mg/Nm

0 ... 750 mg/Nm

NO

0 ... 100 mg/Nm

0 ... 500 mg/Nm


One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

ULTRAMAT 23: CO, NO, SO2

For measuring three infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 250 mg/Nm

0 ... 1 250 mg/Nm

NO

0 ... 400 mg/Nm

0 ... 2 000 mg/Nm

SO2

0 ... 400 mg/Nm

0 ... 2 000 mg/Nm


One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

ULTRAMAT 23: CO, NO, CO2

For measuring three infrared components.

Component

Smallest tested measuring range

Switchable to

CO

0 ... 250 mg/Nm

0 ... 1 250 mg/Nm

NO

0 ... 400 mg/Nm

0 ... 2 000 mg/Nm

CO2

0 ... 5 %

0 ... 25 %


One or two measuring ranges can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

The component CO2 has not been type approved by the TUV.

ULTRAMAT 23: CO2

For measuring one infrared component.

Component

Smallest measuring range

Largest measuring range

CO2

0 ... 5 %

0 ... 25 %


One or two limits can be freely set within the limits. The ULTRAMAT 23 carries out automatic self-calibration with ambient air. The power consumption is 60 VA.

The component CO2 has not been type approved by the TUV.

The ULTRAMAT 23 analyzers can be optionally equipped with an electrochemical oxygen sensor.

O2: Tested measuring ranges 0 to 10 / 25%

FIDAMAT 6: Ctotal (THC)

Component

Smallest tested measuring range

Ctotal

0 ... 10 ppm


The following is supplied:

  • Universal filter heated up to 180 °C. Heater divided in two with heated inlets. Mounted on a base plate with thermally insulated hood, including temperature regulator adjustable from 0 to 180 °C with overtemperature limiter and undertemperature alarm.
  • Filter body of carbon-reinforced PTFE, with ultrafine glass fiber filter element (0.1 µm)
  • Combustion air preparation device for connection to instrument air approximately 6 bar g with pressure regulator and maintenance unit.
  • Hydrogen limitation for installation with cylinder pressure reducer for H2
  • 2/2-way Ex solenoid valve of stainless steel for shutting off the H2 infeed in the event of loss of seal
  • Gas monitor
  • Evaluation instrument
  • Test gas set in plastic case

OXYMAT 6: O2

For paramagnetic measurement of oxygen. Instead of ULTRAMAT 23 with electrochemical cell.

O2: Tested measuring ranges 0 to 10 / 0 to 25%

Sample chamber without flow-type compensation branch, made of stainless steel 1.4571.

LDS 6: HCl

Component

Smallest tested measuring range

HCl

0 ... 15 mg/Nm


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 2 000 mm, applies to gases with a methane content < 15 mg/m . Necessary gas temperature between 120 and 210 °C.

LDS 6: HCl / H2O

Component

Smallest tested measuring range

HCl

0 ... 15 mg/Nm

H2O

0 ... 30 %


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 2 000 mm, applies to gases with a methane content < 15 mg/m . Necessary gas temperature between 120 and 210 °C.

LDS 6: HF

HF: Smallest possible measuring range depends on the gas composition.

Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA. The HF measurement has not been type approved by the TUV.

Limitation:
Component has not been type approved by TUV. Necessary gas temperature between 0 and 150 °C.

LDS 6: HF/H2O

HF: Smallest possible measuring range depends on the gas composition.

H2O: Smallest tested measuring range 0 to 30%

Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA. The HF measurement has not been type approved by the TUV.

Limitation:
Component has not been type approved by TUV. Necessary gas temperature between 0 and 150 °C.

LDS 6: NH3

Component

Smallest tested measuring range

NH3

0 ... 20 mg/Nm


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 1 250 mm. Necessary gas temperature between 0 and 150 °C.

LDS 6: NH3/ H2O

Component

Smallest tested measuring range

NH3

0 ... 20 mg/Nm

H2O

0 ... 15 %


Application for channel 1: Emission monitoring

The power consumption is 50 VA. Suitable for connection of non-Ex sensors, including non-Ex-protected sensor electronics.

The delivery includes a pair of sensors for instrument air or N2 on the process side. The pair of sensors is designed for a moderate flow rate of 0 to 120 l/min. The 400 mm long purging tubes are made of stainless steel. The process connection is DN 65, PN 6. The power consumption is 2 VA.

Limitation:
Applies to measurement paths > 1 250 mm. Necessary gas temperature between 0 and 150 °C.

Hybrid cable

A hybrid cable is required to connect a central unit to one pair of sensors. Versions for 5, 10, 25, 40 and 50 m are available. Cable lengths cannot be combined. Lengths greater than 50 m can be ordered on request.

Sensor cable

A sensor cable is required to connect one pair of sensors. Versions for 5, 10 and 25 m are available. Cable lengths cannot be combined. Lengths greater than 25 m can be ordered on request.

Electrical preparation for dust measurement

Electrical preparation for connection of an external dust measurement to the system.

Electrical preparation for flow measurement

Electrical preparation for connection of an external flow measurement to the system.

Electrical preparation for pressure measurement

Electrical preparation for connection of an external pressure measurement to the system.

Electrical preparation for temperature measurement

Electrical preparation for connection of an external temperature measurement to the system.

Electrical preparation for emission data memory on rail module

Electrical preparation for connection of an emission data memory on rail module to the system.

Electrical preparation for emission data memory in 19" rack unit

Electrical preparation for connection of an emission data memory in 19" rack unit to the system.

Additional LOGO for four or more 19" rack units

Sets with more than three 19" rack units integrated require a LOGO extension module. The delivery also includes connection and programming.

Core end labeling

It is optionally possible to order core end labeling according to the Siemens standard (VDE 0100 Part 200).

Documentation

The Siemens standard documentation is available in German, English or French.

The documentation includes gas path diagram, circuit diagram, terminal diagram, installation diagram, consumable materials list, signal list, cable list, and parts list. Also included are technical data sheets and manuals for the components and devices used. The documentation language for parts provided by other suppliers may deviate. Plant description, LOGO program and test certificates are also included in the documentation.

The documentation contains no customer-specific/project-specific drawings, and consists of two folders and one CD per set.

Figure contains options
























































  © ООО "СК ЭНЕРГО" 2007-2022
  (4872) 700-366
Яндекс.Метрика