General information

General information

Ambient temperature

-30 ... 55 °C (with heating)

Explosion protection

CENELEC Category 2G, T3

Supply voltage

230 V AC, 115 V AC or 24 V DC

Max. permissible pressure at input of high-pressure reduction

160 000 hPa

Max. permissible pressure at input of sample preparation system with pressure regulator

160 000 hPa

Max. permissible pressure at output of sample preparation system

10 ... 500 kPa, min. 200 kPa below carrier gas pressure

Optimum carrier gas pressure 600 ... 700 kPa

Sampling

Lance

Outer diameter 6 mm and inner diameter 2 mm
or outer diameter 12 mm and inner diameter 2 mm
Length 1 000 mm

Special probe with BTU diaphragm and pressure reduction, non-retractable

Immersion depth approx. 380 mm

Protection pipe outer diameter 25.4 mm

Process connection

At flange DN65 PN16 Form C, max. 16 000 hPa gas pressure

At flange DN65 PN160 Form C, max. 160 000 hPa gas pressure

At flange ANSI, 2-inch, 300 lbs RF

At flange ANSI, 2-inch, 2 500 lbs RF

High-pressure reduction in casing with pressure regulators, optional heating

Casing with dimensions 385 x 485 x 380 mm

Primary pressure 160 000 hPa, output pressure 1 000 / 1 700 hPa (power consumption 150 W)

Pipe base

2-inch pipe base for free-standing transmitter casing, 1 700 mm high

Heated line

Heating power is 38 W per meter
Self-controlled heating up to approx. 80 °C
Outer material is PE corrugated hose with 45 mm outer diameter

Sample preparation for 1 to 3 sample gas streams, plus calibration gas stream

Basic configuration

0.5 µm filter, 3/2-way solenoid valve, flowmeter, overflow valve and stop valve

Pressure regulator, optional heating

Primary pressure 160 000 hPa, output pressure 1 000 / 1 700 hPa (power consumption 150 W)

Simple sample injection

One 3/2-way solenoid valve per sample gas stream

DB&B sample injection

Two 3/2-way solenoid valves per sample gas stream

On mounting plate

652 x 422 x 3 mm

In protective casing

750 x 520 x 430 mm

Protective casing heating

Power consumption 300 W

Adjustable between 10 and 40 °C, in increments of 5

Pipe base

2-inch pipe base, 1 700 mm high

Additional filters

Aerosol, glycol

SITRANS CV system components

On mounting plate

650 x 422 x 3 mm

In protective casing

750 x 520 x 430 mm

Protective casing heating

Power consumption 300 W
Adjustable between 10 and 40 °C, in increments of 5.

Terminal box

340 x 170 x 91 mm

Pipe base

2-inch pipe base, 1700 mm high

Heated line

Heating power is 38 W per meter

Self-controlled heating up to approximately 80 °C

Outer material is PE corrugated hose with 45 mm outer diameter

Gas supply

Transfer station

Max. cylinder pressure 200 000 hPa
Output pressure 500 ... 10 000 hPa

Pipe base

2-inch pipe base, mounted
Height 2 200 mm

Gas cylinder cabinet

Gas cylinder cabinet for two 50 l carrier gas cylinders and one 10 l calibration gas cylinder. Dimensions are 1 250 x 400 x 2 050 mm

Cylinder heating

The cylinder heating system is dimensioned for 10 l calibration gas cylinders and ensures that no condensation occurs within the gas cylinder.

Cylinder pressure reducer

Cylinder pressure reducer for reducing primary pressure of max. 300 000 hPa to output pressure 0 … 4 000 hPa

Heated line

Heating power is 38 W per meter
Self-controlled heating up to approximately 80° C
Outer material is PE corrugated hose with 45 mm outer diameter

Calibration gas

Gas mixture 1

Min. storage temperature is -10 °C
The gas mixture remains stable for 36 months.

Gas mixture 2

Min. storage temperature is 0 … 5 °C
The gas mixture remains stable for 36 months.

Gas mixture 3

Min. storage temperature is -20 °C
The gas mixture remains stable for 36 months.


SITRANS CV

Dimensions and weights, installation

Diameter

25 cm (10")

Height

23 cm (9")

Weight

15 kg (35 lbs)

Max. permissible ambient temperature

-20 … 55 °C (-4 … 130 °F)

Installation on

Post, pipe or wall

Gas connections

Swagelok 1/8"

Power supply

Power supply

24 V DC

Power consumption

Typically 18 VA, max. 50 VA

Certificates

Degree of protection

IP65, NEMA 4X

Explosion protection

ATEX II 2 G EEx d IIC T4

FM Class I, Div 1, Groups B, C, D T4

FM Class I, Zone 1, Group II B+H2 T4

CSA Class I, Div 1, Groups B, C, D T4

Communication

Communication

Ethernet, RS485, RS232, ASCII

Protocols

TCP/IP, Modbus RTU

Storage duration

Reports

100 days

Hourly averages

1 year

Daily averages

2 years

Analysis

Sample streams

3 measuring streams, 1 calibration gas stream

Repeatability for calorific value and density

< 0,01 %

Accuracy for calorific value and density

< 0,1 %

Minimum detection limit for neopentane

< 10 ppm

Cycle time

< 180 s

Carrier gas

He

Gas purity

99,9 % (5,0)

Required filtration

Degree of separation 99.99 % for 0.1  m particles

Consumption

10 ... 40 ml/min

Permissible carrier gas pressure

500 … 700 kPA

Perm. sample pressure

10 ... 500 kPA, but min. 200 kPa under carrier gas pressure

Notice: Sample must not contain ethine (acetylene).

Sample flow

20 ... 100 ml/min

Max. sample temperature

120 °C

Phase

Gaseous

Solid components

< 0.1 µm

Required filtration

Degree of separation 99.99 % for 0.1  m particles

You can find further information on SITRANS CV under "Process gas chromatographs".


Measurements can be made within the following working ranges:

Component

Checked working range (in %)

Possible working range (in %)

Methane

57 ... 100

50 ... 100

Nitrogen1)

0 ... 22

0 ... 25

Carbon dioxide

0 ... 12

0 ... 20

Ethane

0 ... 14

0 ... 20

Propane

0 ... 5

0 ... 15

i-butane

0 ... 0,9

0 ... 10

n-butane

0 ... 1,8

0 ... 10

Neopentane

0 ... 0,1

0 ... 1

i-pentane

0 ... 0,12

0 ... 1

n-pentane

0 ... 0,12

0 ... 1

Hexane+2)

0 ... 0.08

0 ... 3

Helium

Concentration can be entered as a fixed value in the components list

No measured component

H2S

< 500 ppm

No measured component


Standard measuring ranges of natural gas components

1) Any oxygen or carbon monoxide present in the sample will be detected along with the nitrogen and, therefore, taken into account when the nitrogen concentration is determined.

2) Hexane+ = iso/n-hexane to iso/n-nonane

Analyses within the checked working range as well as the quality parameters resulting from these (upper and lower calorific values, density, relative density and Wobbe index) correspond to the requirements listed below. Measurements outside these specifications (Table 1, right-hand column) are possible, but checking of the repeatability and correctness has not been carried out.

Concentration range (mol.%)

Repeatability according to ISO 6974-5 (2001); molar fraction (%) absolute

50 < xi < 100

0.03

1 < xi < 50

0.011

0.1 < xi <1

0.006

xi < 0.1

0.006


The repeatability of the measured components satisfies the criteria of ISO 6974-5 (2001)