General(based on DIN EN 61207/IEC 1207. All data refers to the binary gas mixture H2 in N2) |
|
Measuring ranges |
4, internally and externally switchable; automatic measuring range switchover also possible |
Span |
Application-dependent (see ordering data) |
Measuring ranges with suppressed zero point |
Application-dependent (see ordering data) |
Operating position |
Front wall, vertical |
Conformity |
CE marking in accordance with EN 50081-1/EN 50081-2 and RoHS |
Design, enclosure |
|
Degree of protection |
IP65 according to EN 60529 |
Weight |
Approx. 25 kg |
Electrical characteristics |
|
EMC (ElectromagneticCompatibility) |
In accordance with standard requirements of NAMUR NE21 (08/98) and EN 61326 |
Electrical safety |
In accordance with EN 61010-1; overvoltage category II |
Power supply (see nameplate) |
100 AC -10 % ... 120 V AC +10 %, 47 ... 63 Hz or 200 AC -10 % ... 240 V AC +10 %, 47 ... 63 Hz |
Power consumption |
|
Fuse values (gas connection unheated) |
100 ... 120 V F3 1T/250 F4 1T/250 200 ... 240 V F3 0.63T/250 F4 0.63T/250 |
Fuse values (gas connection heated) |
100 ... 120 V F1 1T/250 F2 4T/250 F3 4T/250 F4 4T/250 200 ... 240 V F1 0.63T/250 F2 2.5T/250 F3 2.5T/250 F4 2.5T/250 |
Gas inlet conditions |
|
Sample gas pressure |
800 ... 1 100 hPa (absolute) |
Sample gas flow |
30 ... 90 l/h |
Sample gas temperature |
Min. 0 to max. 50 °C, but above the dew point |
Temperature |
|
|
70 C |
|
80 C (heated) |
Sample gas humidity |
< 90% relative humidity |
Purging gas pressure |
|
|
165 hPa above ambient pressure |
|
Max. 250 hPa above ambient pressure |
Dynamic response(the dynamic and measuring response refers to the measurement of H2 in N2) |
|
Warm-up period |
< 30 min at room temperature (the technical specification will be met after 2 hours) |
Delayed display (T90) |
Approx. 35 s (including dead time) |
Electrical damping |
0 ... 100 s, parameterizable |
Dead time (the diffusion to the probes is the determining variable) |
Approx. 34 s |
Measuring response(the dynamic and measuring response refers to the measurement of H2 in N2) |
|
Output signal fluctuation (3 value) |
< ± 1% of the smallest possible span according to rating plate with electronic damping constant of 1 s |
Zero point drift |
< ± 1% of the current span/week |
Measured-value drift |
< ± 1% of the smallest possible span (according to rating plate)/week |
Repeatability |
< ± 1% of the current span |
Detection limit |
1% of the smallest possible span according to rating plate |
Linearity error |
< ± 1% of the current span |
Influencing variables(referred to sample gas pressure 1 000 hPa absolute, sample gas flow 0.5 l/min, and ambient temperature 25 °C) |
|
Ambient temperature |
< 2%/10 K referred to smallest possible span according to rating plate |
Accompanying gases |
Deviation from zero point (for influence of interfering gas, see section "Cross-interference") |
Sample gas flow |
0.2% of the current measuring span with a change in flow of 0.1 l/min within the permissible flow range |
Sample gas pressure |
< 1 % of the span with a change in pressure of 100 hPa |
Power supply |
< 0.1% of the output signal span with rated voltage ± 10% |
Electrical inputs and outputs |
|
Analog output |
0/2/4 ... 20 mA, isolated; max. load 750 |
Relay outputs |
6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, isolated |
Analog inputs |
2, dimensioned for 0/2/4 … 20 mA for external pressure sensor and correction of cross-interference |
Binary inputs |
6, designed for 24 V, isolated, freely parameterizable, e.g. for measuringrange switchover |
Serial interface |
RS 485 |
Options |
AUTOCAL function with 8 additional binary inputs and 8 additional relay outputs, also with PROFIBUS PA (on request) or PROFIBUS DP (on request) |
Climatic conditions |
|
Permissible ambient temperature |
-40 ... +70 °C during storage and transportation, 5 … 45 °C during operation |
Permissible humidity (dew point must not be fallen below) |
< 90 % relative humidity as annual average, during storage and transportation |