MAXUM edition II |
|
General |
|
Measuring ranges (depending on application) |
|
Temperature range in oven |
5 ... 225 °C |
Temperature control |
± 0.02 °C |
EMI/RFI design |
|
Calibration |
|
|
Manual or automatic |
|
Automatic baseline correction |
|
Standard sample cylinder |
Design, enclosure |
|
Mounting |
|
Weight |
77 kg |
Degree of protection |
IP54, NEMA 3 |
Danger class |
Standard configurations:
Important note! Use in non-hazardous areas requires purging of the electronics area with air or nitrogen. |
Configuration |
|
Oven options |
|
Detector modules for |
thermal conductivity, flame ionization, flame photometry, helium ionization, photo-ionization and electron capture |
Number of detector modules |
|
Sampling and column valves |
Diaphragm valves, diaphragm piston valves, sliding vane rotary valves or slider valves |
Valveless option |
"Live" switching |
Columns |
Packed, micropacked or capillary columns |
Regulation of gas supply |
Up to 8 electronic pressure regulators and up to 6 mechanical pressure regulators |
Electrical characteristics |
|
Power supply |
|
Gas inlet conditions |
|
Sample flow |
5 ... 100 ml/min (application-dependent) |
Sample filter size |
5 µm |
Minimum sample pressure |
35 kPa, standard |
Maximum sample pressure |
2 070 kPa standard, higher pressure as option |
Maximum sample temperature |
121 °C standard; higher temperature as option |
Materials wetted by sample |
Stainless steel and Teflon; other materials as option |
Liquid injection (valve) |
|
Vaporization temperature |
60 ... 400 °C |
Injection volume |
0.3 ... 9.5 l |
Ambient temperature |
-20 ... +150 °C |
Material of parts wetted by sample |
Stainless steel, mat. no. 1.4571, Hastelloy, Monel or special materials |
Control pressure |
400 ... 600 kPa |
Sample pressure |
Max. 5 000 hPa, recommended 0.5 ... 1 000 kPa |
Connections for pipe |
3 mm outer diameter |
Measuring response |
|
Sensitivity (depending on application) |
± 0.5% of span |
Linearity (depending on application) |
± 2% of span |
Effects of vibrations |
Negligible |
Repeatability in % of full span between |
2 and 100 %: ± 0.5 %; 0.05 and 2 %: ± 1 %; 50 and 500 ppm: ± 2 %; 5 and 50 ppm: ± 3 %; 0.5 and 5 ppm: ± 5 % |
Detection limits |
See detectors |
Influencing variables |
|
Effects of ambient temperature |
None with electronic pressure control Different effects with mechanical pressure control (depending on application) |
Electrical inputs and outputs |
|
Standard input and output |
|
Card slots for optional inputs and outputs |
2 |
Input and output cards |
AO 8: 8 electrically isolated analog output channels D IO: 4 digital inputs and 4 digital outputs A I/O: 2 digital inputs and 2 digital outputs, 2 analog inputs and 2 analog outputs |
Digital inputs |
Optocoupler with internal power supply (12 … 24 V DC); switchable by dry contacts. Alternative: switchable by external power supply 12 … 24 V DC (only isolated relay contacts), external power supply, negative connection linked to ground, for a specific digital input. |
Digital outputs |
Floating changeover contacts, max. contact rating: 1 A with 30 V DC. Diode bypass suppression should be used for inductive loads. |
Analog inputs |
-20 ... +20 mA into 50 or -10 ... +10 V Rin = 1 M , alternate insulation up to 10 V |
Analog outputs |
0/4 ... 20 mA into max. 750 , common negative pole, electrically isolated from ground; freely-connectable to ground |
Termination |
Screw terminal for shielded or solid cable with a maximum area of 16 AWG or 1.5 mm2 |
Climatic conditions |
|
Ambient temperature |
-18 ... 50 °C |
Gas supply |
|
Instrument air |
|
Carrier gas |
|
Combustion gas |
|
Combustion air |
|
Corrosion protection |
|
Communication |
|
Serial output |
RS 232, RS 485 |
Ethernet |
Standard 10BaseT Ethernet with RJ 45 connectors |