Cross-duct sensor CD 6
Cross-duct sensor CD 6

General information

Design

Transmitter and detector units, connected by a sensor cable

Materials

Stainless steel

Installation

Horizontally to the optical axis, perpendicular or parallel to the gas flow

Laser protection class

Class 1, safe to the eye

Explosion protection

Optional, according to ATEX II 1 G Ex ia IIC T4, ATEX II 1 D Ex iaD 20 IP65 T135 °C

A defined leak rate can only be guaranteed when using high-pressure window flanges. Otherwise it may be necessary for the owner to carry out an evaluation in accordance with ATEX (DEMKO 06 ATEX 139648X [17]).

Design, enclosure

Degree of protection

IP65

Dimensions

Diameter: 163, L: 395 mm

Purging gas tube in mm

400 (370 net) x 44 x 40 

800 (770 net) x 44 x 40 

1 200 (1 170 net) x 44 x 40 

Weight

2 x approx. 11 kg

Mounting

DN 65/PN 6 or ANSI 4"/150

Please note:

  • For purging tubes with a length of 800 and 1 200 mm, the wall thickness must not exceed 200 mm with DN 65/PN 6 connections. To carry out measurements with thicker walls, please contact Siemens.
  • The optimum adjustment of the flanges can change with high differences in temperature depending on the type of assembly.

Electrical characteristics

Power supply

24 V DC, supply from central unit via hybrid cable

Power consumption

< 2 W during operation

Climatic conditions

Ambient temperature

-30 ... +70 °C during operation,

-40 ... +70 °C during storage and transportation

Humidity

< 95 % RH, above dew point

Pressure

800 ... 1 100 hPa

Temperature range on the sensor side of the process interface (connection plate)

-20 … +70 °C

Measuring conditions

Measurement path

0.3 ... 12 m (other lengths on request)

Gas temperatures

0 ... 1 200 °C, application-dependent

Gas pressure

General: 1 013 ± 50 hPa

With high-pressure window flanges:
CO/CO2 application KD: 800 ... 1 400 hPa

High-pressure O2 application AP: 950 ... 5 000 hPa

Dust load

The influence of dust is very complex and depends on the path length and particle size. The optical damping increases exponentially at longer path lengths. Smaller particles also have a large influence on the optical damping. With high dust load, long path length and small particle size, the technical support at Siemens should be consulted.


Accessories

Purging

Nitrogen is permissible as the purging gas for the sensor side. Nitrogen, steam, air and gases which are not subject to the pressure equipment directive Cat. 2 are permissible as purging gases for the process side.

Purging with instrument air, N2

  • Pressure at purging inlet

2 000 ... 8 000 hPa

  • Max. overpressure in the sensor

< 500 hPa

  • Quality

  • Instrument air

Free of oil and water

  • Nitrogen

Purity better than 99.7 %. For oxygen measurements, and an O2 content < 0.01% in the purging gas (optical path length  1 m, min. 5% oxygen in the process gas)

  • Maximum flow rate

500 l/min

  • Dew point

Benchmark: < -10 °C, condensation on the optics must be avoided

Blower purging

  • Maximum counter pressure

40 hPa

  • Maximum flow rate

850 l/min

  • Power consumption

370 W

  • Degree of protection (fan)

IP54, cover required to protect against rain

Steam purging

  • Steam conditioning

Overheated

  • Maximum temperature

240 °C

  • Minimum pressure

> 4 000 hPa

  • Maximum pressure

16 000 hPa, refers to a volume flow of approx. 1 100 l/min


Hybrid and sensor cables

General information

Configuration hybrid cable

Two optical fibers and two twisted copper wires in one cable for 24 V DC. Single-mode optical fiber configured at both ends with E2000 angle connectors. Multimode optical fiber configured at both ends with SMA connectors.

Cable sheath

Oil-resistant polyurethane

Dimensions

  • For > 500 m, an external power supply must be additionally ordered
  • For installation in hazardous zones, non-intrinsically-safe cables have to be spatially separated from intrinsically-safe lines
  • Diameter

< 8 mm

  • Length
  • Use in non-hazardous and Ex Zone 2: Up to 700 m
  • Use in Ex Zone 0 and Zone 1: Up to 250 m

Impact resistance

200 N/cm

Maximum tensile strength

500 N

Minimum bending radius

10 cm

Climatic conditions

Ambient temperature

-40 ... +80 °C during operation

Humidity

< 95% rel. humidity, above dew point (in operation and storage)
























































Каталог оборудования
Каталог продуктов Siemens Industry Приводная техника Техника автоматизации Системы автоматизации Системы визуализации SIMATIC HMI Системы идентификации Промышленные коммуникации SIMATIC NET Промышленные аппараты управления SIRIUS Промышленные информационные технологии Управление на базе РС Системы управления процессом Датчики Контрольно-измерительные приборы Анализаторы процесса Непрерывные анализаторы газов Непрерывные анализаторы газов, эктрактивные Непрерывный газовый анализ, in-situ Introduction to LDS 6 and SITRANS SL LDS 6 General information 19" central unit Cross-duct sensor CD 6 Documentation Suggestions for spare parts SITRANS SL Промышленные газовые хроматографы Analytical Application Sets General information Технологии взвешивания Датчики зрения Блоки питания Продукты для специальных требований Автоматизация зданий Низковольтная коммутационная техника Технология безопасности Системные решения и продукты для отраслей Сервис ... и все, что Вам еще необходимо